Automated External Defibrillator - HeartStart MRx Monitor/Defibrillator
Class I - DangerousWhat Should You Do?
- Check if you have this product: 20 units - US00597060, US00597964, US00598048, US00598210, US00598784, US00599490, US00599495, US00599508, US00599572, US00599613, US00599615, US00599623, US00599726, US00600362, US00600363, US00600370, US00600422, US00600423, US00600438, US00600483
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips Electronics North America Corporation
- Reason for Recall:
- Philips MRx devices have an EtCO2 module that may contain contamination; this contamination (will/may) cause the device to be unable to display the EtCO2 or CO2 waveform (capnogram).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Automated External Defibrillator - HeartStart MRx Monitor/Defibrillator
Product Codes/Lot Numbers:
20 units - US00597060, US00597964, US00598048, US00598210, US00598784, US00599490, US00599495, US00599508, US00599572, US00599613, US00599615, US00599623, US00599726, US00600362, US00600363, US00600370, US00600422, US00600423, US00600438, US00600483
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0278-2018
Related Recalls
Xper Flex Cardio Physiomonitoring System, Software versions 1.5.16.0322 and 1.6.1.0390, Model Numbers 453564241901, 453564241911, 453564483321, 453564483331, 453564621791, 453564621801, 453564669081
Philips Electronics North America
There are 3 issues: 1) Under certain conditions, real time waveforms may not be accurately plotted or displayed. 2) The oxygen saturation (SpO2) numeric value displayed on the device may freeze. 3) In auto cycle mode, if the NIBP communication is lost, the pump would not cycle and the non-invasive blood pressure (NIBP) numeric value displayed on the device would not update.
A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above 4,500 feet when operating in Monitor Mode. These devices may intermittently exhibit Speaker Malfunct INOP messages after the Power On Self-Test. The issue is only apparent when the device is being used in Monitor Mode at an altitude of greater than 4,500 feet.
Automated external defibrillators may not fully meet IPx5 water ingress specification. The device may fail to function should water intrusion occur.