Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10) REF LK018.10S

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model Number: LK018.S UDI Code: 05051700006477 Lot Number: 472776 Model Number:LK018.10S UDI Code: 05051700012201 Lot Number: 472664
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
The Binding Site Group, Ltd.
Reason for Recall:
Due to calibration curves producing a higher activity than routinely expected and that there may be a decrease in separation between calibrator point 5 and calibrator point 6, which can be exacerbated with calibrator 6 activity reaches the optical ceiling of the SPA+ analyzer.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10) REF LK018.10S

Product Codes/Lot Numbers:

Model Number: LK018.S UDI Code: 05051700006477 Lot Number: 472776 Model Number:LK018.10S UDI Code: 05051700012201 Lot Number: 472664

Distribution:

Distributed in: CA, CO, FL, IL, IN, KY, LA, MI, NY, OK, SD, TX, VA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0274-2022

Related Recalls

Assigned Rheumatoid Factor (RF) calibrator value for calibrator component included in kit used to aid in the diagnosis of rheumatoid arthritis over recovers by 7% in comparison to the reference material RF serum. This could result in reporting abnormal (false positive) results, which may lead to unnecessary additional serological testing and a delay in treatment.

Oct 7, 2019 Infusion Pumps Nationwide View Details →