L3 Security & Detection System Cabinet X-ray system used for security screening of checked airline baggage.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model Number : Examiner 3DX 6000 cabinet x-ray system
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
L3 Security & Detection
Reason for Recall:
L3 has determined in airport environments, there have been several cases where a small air gap has developed where the entrance or exit tunnels meet up with the main x-ray scanner tunnel. This a gap in x-ray shielding can result in x-ray emissions that exceed the limit in the federal performance standard for cabinet x-ray systems (see 21 CFR ¿ 1020.40(c)(1)).
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

L3 Security & Detection System Cabinet X-ray system used for security screening of checked airline baggage.

Product Codes/Lot Numbers:

Model Number : Examiner 3DX 6000 cabinet x-ray system

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0270-2018