OneLIF Torque Limiting Adapter, AO to 1/4 SQ, 2.25Nm Limit, Catalog #015-010-0000, a component of NovApproach Spine OneLIF Interbody System instrument kit, Catalog #OneLIF INS.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Torque adapter - Lot #104713, UDI G07001501000000; Kit IDs - 504, 505, 506, 501, 101, and 105.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Novapproach Spine LLC
Reason for Recall:
The torque limiting adapter may not stay locked inside of the ratcheting handle once fully engaged.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
completed

Product Information

Full Description:

OneLIF Torque Limiting Adapter, AO to 1/4 SQ, 2.25Nm Limit, Catalog #015-010-0000, a component of NovApproach Spine OneLIF Interbody System instrument kit, Catalog #OneLIF INS.

Product Codes/Lot Numbers:

Torque adapter - Lot #104713, UDI G07001501000000; Kit IDs - 504, 505, 506, 501, 101, and 105.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0246-2023