Programmers & Remote Monitoring Software Apps: 2090 CareLink Programmer
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: PKK051331R PKK200187R PKK209956R PKK207572R PKK209265R PKK016592R PKK202460R PKK019811R PKK209256R PKK007753R PKK015981R PKK016604R PKK017955R PKK018301R PKK019555R PKK019776R PKK051452R PKK052792R PKK204491R PKK204504R PKK204551R PKK204575R PKK204595R PKK204640R PKK204740R PKK203409R PKK006360R PKK202461R PKK018408R PKK019864R PKK204483R PKK016046R PKK016047R PKK019820R PKK020201R PKK010563R PKK209937R PKK209928R PKK050581R PKK209305R PKK211241R PKK209294R PKK201567R PKK206687R PKK211282R PKK208961R PKK209373R PKK202996R PKK210858R PKK202566R PKK211233R PKK205628R PKK209203R PKK211710R PKK100124R PKK211042R PKK210807R PKK200224R PKK208864R PKK209197R PKK208958R PKK209332R PKK211720R PKK016352R PKK019371R PKK211286R PKK202507R PKK205997R PKK205974R PKK202465R PKK017835R PKK018895R PKK201637R PKK201640R PKK201643R PKK201645R PKK202394R PKK202397R PKK202411R PKK202428R PKK202441R PKK202444R PKK202451R PKK202454R PKK202459R PKK202463R PKK202464R PKK202466R PKK202489R PKK202494R PKK202521R PKK202550R PKK202552R PKK202553R PKK202557R PKK202558R PKK202561R PKK202563R PKK202564R PKK202565R PKK202567R PKK204964R PKK204973R PKK204974R PKK204983R PKK204985R PKK205962R PKK205968R PKK205970R PKK205977R PKK205979R PKK205982R PKK205986R PKK205989R PKK205994R PKK205998R PKK205999R PKK206000R PKK207397R PKK207399R PKK207576R PKK207588R PKK207601R PKK207603R PKK207604R PKK207606R PKK207607R PKK207615R PKK207621R PKK207622R PKK207623R PKK207628R PKK207649R PKK207650R PKK208500R PKK208503R PKK208508R PKK208510R PKK211023R PKK211064R PKK211072R PKK211084R PKK211100R PKK211227R PKK211228R PKK211229R PKK211235R PKK211238R PKK211239R PKK211284R PKK211289R PKK211294R PKK211339R PKK211342R PKK203579R PKK107297R PKK108724R PKK108726R PKK108729R PKK109569R PKK201389R PKK201402R PKK051032R PKK051031R PKK010204R PKK106139R PKK106149R PKK106153R PKK106161R PKK106164R PKK106212R PKK106221R PKK106224R PKK106228R PKK054839R PKK054899R PKK054903R PKK006109R PKK011161R PKK011162R PKK012402R PKK012404R PKK020167R PKK052616R PKK052619R PKK208773R PKK206171R PKK108840R PKK205827R PKK018974R PKK054297R PKK052590R PKK055474R PKK019344R PKK208281R PKK020200R PKK056378R PKK107503R PKK106539R PKK005940R PKK201926R PKK202805R PKK205627R PKK003769R PKK016555R PKK004307R PKK001417R PKK006119R PKK005232R PKK004040R PKK013386R PKK019411R PKK052102R PKK018098R PKK101253R PKK205556R PKK102674R PKK006689R PKK200245R PKK106865R PKK211492R PKK017878R PKK010764R PKK211518R PKK211521R PKK057197R PKK011242R PKK200949R PKK051450R PKK018218R PKK057195R PKK201686R PKK020573R PKK014310R PKK054238R PKK209424R PKK208646R PKK208224R PKK020366R PKK207934R PKK053799R PKK108035R PKK204464R PKK051548R PKK203435R PKK014418R PKK102204R PKK050960R PKK108305R PKK001172R PKK102818R PKK205393R PKK013868R PKK053717R PKK053233R PKK002545R PKK202693R PKK002871R PKK106723R PKK055305R PKK108322R PKK102503R PKK202024R PKK003498R PKK004347R PKK015584R PKK105180R PKK003235R PKK014949R PKK018886R PKK019330R PKK104774R PKK014465R PKK054834R PKK055373R PKK101011R PKK106795R PKK205887R PKK205332R PKK002384R PKK015619R PKK055620R PKK003926R PKK006407R PKK007261R PKK014428R PKK017571R PKK019868R PKK050318R PKK107143R PKK006589R PKK018884R PKK103154R PKK107158R PKK204136R PKK015641R PKK104118R PKK001847R PKK002178R PKK053006R PKK017686R PKK211216R PKK201780R PKK015586R PKK101918R PKK006674R PKK013504R PKK052353R PKK100694R PKK102490R PKK106404R PKK003165R PKK007298R PKK052052R PKK105856R PKK100708R PKK052050R PKK102715R PKK006557R PKK101957R PKK002398R PKK004434R PKK006460R PKK101141R PKK003318R PKK017784R PKK107927R PKK209592R PKK052937R PKK103113R PKK103401R PKK107979R PKK102706R PKK103417R PKK051039R PKK108929R PKK002518R PKK018602R PKK014370R PKK050450R PKK018051R PKK018954R PKK006683R PKK015308R PKK206137R PKK017096R PKK201696R PKK013110R PKK053602R PKK103395R PKK104573R PKK106395R PKK106663R PKK107105R PKK108835R PKK019743R PKK002029R PKK006880R PKK007419R PKK052069R PKK012323R PKK051108R PKK207988R PKK053982R PKK055228R PKK201744R PKK002557R PKK208342R PKK013078R PKK014144R PKK014900R PKK055981R PKK103121R PKK001953R PKK005823R PKK007818R PKK005994R PKK015611R PKK051055R PKK056120R PKK056431R PKK014363R PKK014324R PKK017674R PKK019498R PKK003199R PKK005355R PKK007162R PKK103381R PKK013092R PKK056024R PKK056434R PKK103151R PKK104749R PKK106810R PKK107817R PKK010011R PKK206429R PKK004752R PKK003279R PKK011923R PKK208647R PKK004440R PKK208616R PKK202512R PKK208696R PKK002498R PKK208692R PKK101219R PKK006587R PKK018978R PKK055097R PKK207723R PKK106912R PKK203509R PKK201788R PKK006737R PKK206152R PKK011005R PKK015511R PKK012410R PKK015570R PKK019464R PKK013598R PKK001597R PKK007512R PKK018480R PKK100995R PKK003991R PKK004471R PKK105584R PKK001580R PKK011440R PKK053484R PKK002364R PKK010024R PKK050689R PKK006639R PKK014435R PKK011200R PKK205410R PKK055049R PKK102825R PKK104021R PKK014983R PKK017725R PKK050719R PKK005068R PKK107473R PKK017915R PKK104060R PKK208481R PKK107727R PKK015617R PKK006992R PKK109159R PKK109545R PKK019949R PKK012658R PKK018105R PKK104070R PKK108996R PKK103297R PKK003022R PKK001631R PKK001716R PKK018665R PKK019213R PKK020534R PKK109474R PKK012315R PKK202808R PKK104056R PKK109498R PKK200651R PKK208449R PKK003879R PKK056493R PKK016902R PKK105239R PKK017266R PKK015985R PKK108025R PKK014294R PKK200133R PKK103212R PKK101320R PKK101826R PKK211039R PKK106426R PKK211341R PKK211762R PKK210929R PKK011959R PKK001989R PKK003135R PKK007803R PKK002374R PKK054237R PKK207987R PKK201852R PKK055407R PKK050352R PKK015579R PKK019212R PKK055347R PKK014579R PKK012884R PKK011957R PKK014188R PKK010404R PKK209440R PKK019372R PKK106218R PKK106217R PKK208708R PKK207810R PKK051806R PKK108336R PKK108369R PKK050744R PKK013244R PKK106336R PKK206844R PKK001150R PKK003214R PKK207068R PKK104484R PKK103658R PKK011039R PKK020669R PKK055886R PKK053829R PKK051107R PKK106898R PKK211602R PKK208904R PKK054601R PKK211337R PKK208740R PKK106387R PKK050567R PKK051869R PKK014654R PKK109164R PKK102115R PKK015163R PKK107765R PKK202722R PKK211182R PKK015252R PKK202390R PKK205467R PKK109632R PKK005601R PKK201789R PKK211705R PKK015263R PKK052799R PKK050310R PKK208380R PKK211654R PKK011165R PKK211403R PKK201493R PKK203454R PKK211397R PKK012419R PKK012429R PKK013764R PKK054068R PKK054388R PKK054942R PKK056919R PKK056996R PKK109612R PKK108760R PKK050338R PKK209367R PKK017148R PKK108013R PKK102003R PKK210881R PKK210674R PKK209559R PKK108960R PKK004183R PKK011345R PKK016954R PKK004004R PKK210692R PKK005395R PKK102767R PKK205309R PKK105818R PKK011669R PKK105677R PKK106816R PKK208799R PKK210805R PKK101424R PKK100439R PKK101212R PKK103537R PKK209295R PKK016849R PKK017011R PKK017672R PKK017773R PKK018062R PKK107631R PKK206120R PKK010994R PKK210449R PKK019410R PKK108981R PKK208115R PKK208356R PKK101301R PKK019522R PKK204452R PKK211371R PKK210197R PKK210203R PKK100356R PKK001507R PKK018827R PKK019283R PKK019303R PKK019681R PKK006048R PKK015843R PKK106330R PKK208336R PKK056460R PKK204507R PKK013995R PKK005789R PKK106964R PKK201169R PKK200645R PKK207872R PKK004169R PKK007080R PKK100961R PKK015088R PKK014647R PKK017508R PKK005837R PKK100277R PKK006008R PKK006823R PKK004962R PKK206059R PKK204550R PKK202254R PKK208166R PKK202434R PKK051904R PKK101472R PKK201879R PKK056155R PKK107324R PKK006421R PKK109529R PKK015842R PKK107677R PKK051445R PKK014860R PKK007333R PKK012936R PKK002963R PKK204772R PKK103119R PKK103696R PKK011816R PKK017469R PKK205118R PKK054711R PKK002299R PKK055374R PKK001606R PKK003774R PKK204420R PKK100105R PKK011153R PKK204288R PKK003453R PKK006077R PKK007003R PKK007301R PKK016731R PKK019004R PKK051301R PKK051736R PKK100943R PKK206186R PKK209436R PKK202707R PKK200492R PKK002691R PKK003626R PKK101391R PKK200232R PKK207985R PKK102179R PKK203110R PKK208455R PKK202703R PKK016635R PKK001489R PKK208765R PKK011302R PKK208258R PKK055211R PKK019927R PKK016787R PKK103057R PKK015775R PKK013051R PKK016884R PKK108005R PKK209562R PKK100246R PKK056007R PKK206175R PKK051893R PKK005793R PKK102822R PKK106590R PKK017528R PKK016873R PKK201371R PKK005927R PKK019505R PKK106976R PKK107003R PKK107159R PKK107281R PKK107348R PKK107452R PKK107462R PKK107466R PKK108721R PKK210252R PKK205357R PKK013222R PKK056089R PKK006178R PKK011529R PKK107762R PKK056303R PKK001357R PKK054097R PKK203694R PKK105007R PKK107761R PKK014477R PKK018965R PKK003560R PKK056028R PKK056100R PKK018469R PKK021398R PKK010864R PKK006633R PKK201086R PKK002785R PKK012829R PKK102887R PKK103270R PKK020284R PKK054670R PKK202224R PKK011069R PKK010718R PKK209931R PKK101274R PKK012267R PKK010743R PKK202822R PKK010492R PKK209929R PKK010214R PKK105192R PKK209922R PKK204686R PKK201863R PKK208817R PKK002937R PKK107359R PKK107377R PKK010138R PKK107540R PKK205040R PKK108523R PKK002457R PKK053254R PKK205837R PKK201069R PKK020725R PKK108522R PKK210035R PKK210040R PKK005472R PKK204663R PKK054448R PKK004455R PKK209604R PKK051143R PKK052827R PKK050261R PKK014350R PKK209823R PKK015900R PKK017605R PKK018263R PKK106769R PKK107402R PKK202417R PKK014327R PKK209718R PKK052633R PKK102908R PKK211431R PKK211456R PKK201254R PKK207559R PKK051609R PKK205687R PKK103561R PKK203956R PKK106963R PKK013924R PKK004774R PKK200573R PKK014739R PKK107819R PKK050308R PKK204091R PKK207016R PKK201384R PKK104930R PKK103890R PKK012505R PKK053073R PKK054901R PKK004912R PKK106433R PKK202219R PKK100312R PKK205699R PKK108610R PKK101590R PKK103167R PKK208786R PKK010752R PKK109488R PKK052953R PKK014727R PKK207675R PKK053032R PKK004624R PKK200830R PKK015097R PKK010213R PKK106549R PKK206590R PKK016658R PKK211123R PKK208373R PKK201226R PKK056691R PKK053336R PKK204600R PKK100789R PKK103907R PKK012007R PKK019302R PKK051250R PKK108326R PKK053004R PKK004839R PKK018578R PKK015951R PKK052377R PKK203772R PKK002519R PKK001865R PKK013324R PKK200221R PKK202329R PKK054668R PKK013087R PKK006355R PKK006631R PKK207359R PKK051470R PKK054866R PKK108533R PKK202227R PKK207241R PKK055357R PKK200206R PKK010422R PKK055724R PKK205742R PKK206002R PKK055268R PKK101105R PKK011093R PKK101577R PKK202503R PKK200815R PKK001364R PKK051321R PKK001376R PKK020271R PKK013984R PKK208243R PKK107932R PKK015754R PKK107940R PKK108952R PKK004064R PKK106190R PKK051987R PKK102663R PKK108970R PKK205078R PKK204330R PKK202082R PKK018740R PKK101128R PKK054029R PKK014466R PKK002363R PKK202069R PKK207917R PKK205478R PKK103971R PKK019026R PKK205106R PKK106843R PKK007838R PKK208408R PKK202997R PKK052609R PKK106945R PKK209647R PKK014786R PKK017583R PKK106282R PKK005904R PKK106821R PKK101492R PKK055343R PKK108778R PKK101071R PKK208378R PKK007813R PKK101067R PKK109005R PKK204331R PKK104929R PKK200747R PKK104386R PKK105798R PKK052338R PKK100161R PKK205733R PKK011812R PKK108868R PKK004206R PKK014899R PKK205131R PKK054687R PKK202750R PKK052470R PKK010576R PKK011817R PKK200247R PKK104452R PKK107155R PKK207706R PKK107913R PKK054898R PKK051070R PKK021137R PKK015047R PKK015770R PKK051511R PKK006787R PKK105792R PKK050580R PKK054776R PKK201766R PKK018422R PKK006733R PKK208103R PKK207199R PKK209244R PKK208544R PKK018813R PKK209736R PKK005647R PKK018928R PKK012484R PKK105032R PKK052648R PKK102660R PKK002477R PKK054772R PKK207858R PKK006237R PKK013653R PKK053347R PKK108891R PKK203497R PKK002388R PKK013955R PKK020601R PKK012369R PKK106566R PKK211734R PKK018159R PKK109556R PKK021429R PKK015565R PKK207335R PKK014115R PKK004547R PKK101910R PKK200676R PKK014558R PKK050472R PKK105646R PKK102753R PKK013553R PKK002094R PKK106791R PKK206195R PKK106730R PKK206405R PKK105071R PKK205644R PKK052886R PKK052497R PKK056005R PKK014394R PKK204449R PKK007628R PKK106206R PKK050858R PKK054267R PKK056391R PKK207958R PKK011369R PKK007388R PKK018019R PKK106088R PKK015125R PKK005086R PKK014202R PKK019839R PKK002522R PKK007270R PKK051960R PKK013740R PKK108613R PKK051120R PKK055800R PKK052997R PKK105300R PKK018680R PKK108601R PKK202334R PKK208989R PKK209800R PKK201598R PKK013558R PKK209361R PKK051199R PKK012650R PKK208919R PKK211070R PKK050573R PKK020538R PKK202554R PKK005627R PKK109313R PKK109548R PKK108649R PKK054715R PKK012094R PKK002144R PKK054252R PKK054409R PKK053256R PKK010558R PKK017042R PKK013976R PKK200969R PKK014705R PKK006328R PKK200910R PKK051242R PKK006202R PKK103305R PKK021465R PKK201822R PKK003174R PKK055598R PKK005283R PKK100340R PKK101288R PKK105117R PKK208874R PKK204897R PKK055802R PKK205134R PKK205967R PKK104528R PKK205528R PKK203756R PKK203224R PKK203615R PKK020793R PKK106665R PKK203866R PKK019862R PKK050817R PKK208612R PKK019853R PKK108156R PKK053482R PKK108641R PKK107356R PKK104988R PKK207082R PKK102383R PKK052432R PKK006128R PKK053120R PKK104803R PKK020678R PKK208640R PKK108995R PKK014092R PKK007342R PKK204106R PKK014110R PKK001738R PKK101024R PKK005666R PKK201840R PKK201111R PKK203964R PKK001339R PKK002352R PKK002508R PKK002535R PKK002836R PKK002923R PKK003668R PKK003892R PKK003958R PKK004610R PKK005375R PKK006270R PKK006445R PKK006548R PKK006724R PKK006853R PKK007253R PKK010847R PKK012019R PKK012325R PKK014252R PKK014258R PKK014303R PKK014515R PKK015487R PKK015499R PKK015650R PKK016273R PKK016307R PKK016312R PKK017035R PKK018900R PKK019459R PKK019765R PKK020281R PKK020407R PKK050166R PKK050173R PKK050241R PKK050298R PKK050372R PKK050478R PKK050574R PKK050660R PKK050762R PKK050767R PKK051015R PKK051131R PKK051266R PKK051613R PKK051677R PKK051741R PKK052082R PKK052283R PKK052659R PKK052994R PKK053008R PKK053170R PKK053719R PKK053747R PKK053771R PKK053792R PKK055006R PKK055030R PKK055250R PKK055267R PKK056379R PKK003118R PKK006089R PKK006770R PKK006900R PKK007819R PKK011009R PKK011285R PKK014717R PKK015683R PKK018707R PKK004021R PKK005090R PKK007303R PKK101086R PKK106134R PKK054727R PKK006325R PKK018046R PKK050906R PKK100590R PKK203438R PKK204017R PKK100397R PKK018302R PKK018307R PKK050781R PKK021301R PKK011441R PKK011036R PKK101206R PKK203938R PKK003057R PKK055070R PKK054140R PKK006523R PKK204713R PKK109595R PKK202531R PKK202605R PKK200222R PKK021323R PKK016083R PKK211566R PKK100672R PKK108637R PKK003533R PKK053615R PKK106875R PKK201821R PKK011045R PKK107412R PKK200497R PKK206133R PKK052026R PKK055304R PKK001516R PKK005442R PKK005392R PKK016410R PKK211664R PKK211706R PKK020787R PKK210510R PKK210512R PKK057223R PKK001963R PKK016206R PKK021117R PKK205734R PKK052910R PKK206020R PKK014204R PKK200708R PKK012367R PKK014595R PKK107686R PKK202201R PKK054220R PKK011851R PKK210657R PKK011633R PKK051638R PKK021304R PKK100132R PKK007448R PKK203377R PKK207326R PKK207015R PKK210884R PKK006765R PKK101340R PKK107248R PKK204215R PKK207996R PKK205614R PKK209677R PKK209826R PKK209684R PKK211686R PKK001684R PKK005493R PKK205116R PKK205019R PKK205114R PKK207264R PKK016789R PKK201704R PKK013848R PKK054305R PKK200627R PKK103306R PKK010668R PKK207487R PKK001589R PKK206042R PKK205959R PKK050866R PKK051785R PKK102313R PKK205560R PKK013845R PKK102284R PKK053072R PKK207743R PKK004794R PKK004467R PKK203505R PKK007011R PKK013043R PKK207742R PKK053259R PKK002752R PKK015594R PKK208593R PKK013822R PKK054927R PKK101354R PKK054730R PKK005428R PKK013906R PKK019760R PKK205420R PKK209547R PKK205992R PKK017960R PKK051423R PKK004399R PKK051348R PKK101390R PKK052977R PKK052805R PKK051045R PKK108050R PKK001893R PKK054880R PKK206505R PKK005595R PKK209805R PKK104203R PKK200847R PKK101745R PKK013014R PKK208821R PKK052457R PKK053296R PKK208437R PKK206553R PKK014138R PKK105741R PKK201983R PKK054851R PKK017401R PKK106925R PKK204959R PKK209790R PKK011042R PKK208648R PKK208663R PKK206451R PKK050599R PKK204195R PKK109614R PKK017873R PKK108249R PKK205494R PKK208887R PKK013467R PKK019389R PKK208955R PKK208972R PKK051155R PKK207307R PKK207861R PKK021271R PKK200541R PKK209419R PKK017745R PKK201139R PKK202863R PKK209769R PKK007808R PKK104205R PKK206503R PKK017152R PKK012933R PKK108330R PKK107562R PKK011647R PKK050847R PKK050852R PKK050975R PKK002967R PKK054754R PKK050986R PKK104585R PKK102087R PKK015693R PKK051110R PKK200356R PKK001849R PKK202341R PKK016691R PKK017679R PKK016978R PKK017626R PKK104589R PKK101744R PKK206113R PKK208199R PKK014445R PKK207179R PKK102101R PKK053689R PKK206684R PKK005798R PKK201897R PKK208712R PKK053224R PKK206460R PKK207894R PKK015543R PKK105763R PKK016998R PKK203476R PKK014302R PKK055223R PKK014272R PKK054870R PKK103354R PKK052969R PKK100829R PKK107599R PKK211868R PKK013986R PKK017937R PKK107589R PKK202067R PKK211232R PKK011339R PKK010811R PKK209633R PKK020298R PKK203408R PKK109403R PKK209949R PKK105178R PKK005741R PKK010459R PKK206725R PKK013665R PKK207037R PKK107844R PKK106546R PKK109386R PKK017545R PKK108057R PKK051576R PKK204058R PKK107394R PKK011563R PKK019904R PKK012002R PKK209642R PKK102096R PKK016799R PKK016261R PKK104871R PKK203335R PKK209501R PKK013961R PKK105647R PKK105650R PKK207590R PKK203482R PKK014236R PKK206870R PKK202371R PKK200763R PKK201238R PKK201729R PKK200793R PKK012828R PKK050359R PKK011607R PKK107352R PKK103582R PKK021294R PKK209340R PKK208362R PKK004113R PKK051398R PKK208337R PKK002044R PKK012409R PKK104535R PKK014643R PKK002370R PKK002669R PKK002850R PKK004381R PKK011505R PKK012363R PKK013004R PKK013121R PKK014925R PKK017026R PKK017607R PKK018999R PKK019007R PKK019503R PKK004076R PKK004716R PKK010057R PKK011872R PKK017708R PKK017711R PKK018027R PKK018966R PKK051430R PKK054740R PKK054884R PKK055627R PKK018032R PKK004605R PKK004813R PKK014080R PKK014166R PKK014624R PKK014792R PKK015561R PKK017893R PKK018670R PKK020119R PKK001068R PKK002367R PKK002525R PKK003468R PKK014589R PKK003503R PKK005075R PKK005244R PKK006115R PKK006558R PKK007766R PKK011588R PKK013032R PKK014297R PKK015005R PKK015553R PKK017620R PKK050621R PKK051261R PKK054357R PKK053319R PKK003301R PKK003904R PKK010087R PKK015229R PKK050412R PKK052535R PKK053083R PKK002322R PKK003038R PKK006085R PKK010474R PKK011230R PKK011868R PKK013293R PKK019742R PKK050465R PKK050526R PKK052266R PKK052562R PKK054258R PKK056381R PKK005186R PKK006719R PKK013897R PKK014460R PKK018595R PKK003506R PKK006655R PKK010016R PKK011610R PKK016374R PKK052355R PKK055194R PKK055288R PKK056126R PKK002823R PKK004020R PKK006399R PKK012204R PKK013329R PKK014227R PKK014307R PKK016671R PKK050512R PKK051756R PKK051781R PKK053084R PKK053279R PKK004046R PKK006017R PKK007249R PKK013890R PKK016302R PKK018872R PKK002466R PKK003274R PKK004134R PKK004486R PKK005268R PKK007401R PKK007737R PKK014377R PKK003310R PKK004023R PKK004904R PKK006195R PKK006494R PKK013500R PKK014254R PKK019016R PKK002137R PKK003284R PKK004349R PKK005815R PKK011296R PKK014619R PKK016649R PKK052109R PKK052398R PKK054153R PKK055036R PKK004395R PKK005256R PKK006550R PKK013015R PKK013556R PKK014372R PKK015502R PKK017759R PKK018768R PKK050214R PKK053065R PKK054232R PKK054894R PKK055051R PKK055836R PKK003190R PKK003388R PKK004379R PKK004592R PKK004597R PKK004928R PKK005153R PKK005171R PKK005926R PKK006427R PKK007436R PKK011176R PKK011451R PKK012559R PKK013079R PKK014107R PKK015030R PKK015564R PKK015569R PKK017769R PKK017932R PKK018982R PKK019412R PKK019474R PKK019506R PKK051853R PKK053553R PKK054324R PKK003366R PKK006867R PKK010023R PKK010772R PKK011359R PKK012839R PKK014059R PKK014205R PKK014492R PKK017662R PKK018759R PKK020087R PKK051109R PKK052901R PKK002786R PKK003102R PKK003226R PKK003351R PKK004056R PKK004304R PKK004305R PKK005906R PKK006752R PKK006869R PKK007532R PKK007725R PKK012105R PKK013824R PKK015602R PKK018859R PKK002283R PKK003869R PKK004326R PKK005600R PKK006891R PKK006973R PKK006999R PKK013973R PKK014309R PKK018598R PKK018678R PKK054913R PKK005935R PKK013374R PKK003170R PKK015273R PKK015286R PKK015453R PKK050366R PKK051798R PKK002805R PKK003096R PKK003209R PKK003317R PKK003602R PKK003989R PKK004872R PKK006534R PKK006897R PKK007530R PKK007805R PKK010370R PKK011584R PKK013892R PKK017591R PKK017737R PKK018832R PKK051804R PKK052578R PKK052876R PKK056056R PKK003225R PKK007807R PKK010820R PKK012586R PKK015538R PKK017589R PKK019907R PKK005943R PKK006035R PKK007339R PKK010079R PKK013530R PKK014177R PKK018725R PKK019511R PKK004072R PKK011206R PKK014242R PKK019942R PKK054766R PKK050501R PKK006945R PKK006949R PKK007223R PKK016344R PKK018609R PKK050170R PKK050795R PKK052566R PKK052580R PKK006581R PKK006908R PKK007489R PKK011549R PKK012415R PKK013081R PKK017220R PKK051754R PKK054912R PKK054933R PKK001408R PKK005991R PKK005275R PKK003906R PKK004417R PKK004698R PKK005965R PKK007538R PKK011262R PKK012262R PKK056018R PKK007286R PKK014289R PKK014750R PKK014984R PKK016712R PKK018043R PKK019443R PKK019668R PKK050704R PKK050848R PKK050934R PKK051772R PKK052762R PKK053920R PKK055119R PKK002114R PKK004117R PKK004371R PKK004934R PKK005301R PKK006225R PKK006277R PKK006431R PKK006473R PKK006722R PKK006822R PKK007267R PKK007497R PKK010008R PKK011238R PKK013489R PKK013534R PKK002337R PKK002355R PKK003124R PKK003267R PKK003413R PKK003518R PKK003942R PKK003176R PKK006220R PKK050616R PKK053590R PKK003645R PKK016921R PKK050421R PKK052905R PKK053260R PKK053335R PKK003041R PKK050458R PKK050477R PKK050663R PKK051179R PKK052179R PKK052593R PKK053679R PKK054684R PKK005089R PKK005903R PKK006442R PKK050230R PKK052887R PKK052943R PKK053992R PKK054158R PKK054825R PKK056132R PKK051411R PKK002058R PKK005840R PKK006366R PKK010369R PKK053904R PKK055248R PKK010030R PKK011013R PKK011051R PKK011863R PKK019531R PKK019902R PKK020148R PKK050297R PKK050431R PKK050572R PKK051528R PKK054705R PKK002328R PKK003178R PKK003632R PKK004621R PKK005163R PKK006348R PKK006430R PKK006670R PKK006750R PKK006816R PKK006819R PKK010381R PKK010767R PKK010785R PKK010944R PKK011053R PKK012413R PKK013704R PKK013809R PKK014958R PKK014972R PKK015029R PKK015662R PKK016284R PKK016289R PKK016298R PKK016343R PKK016347R PKK016759R PKK016904R PKK017627R PKK017771R PKK018100R PKK018607R PKK018629R PKK019247R PKK051323R PKK052073R PKK052348R PKK052469R PKK002314R PKK002733R PKK002867R PKK003677R PKK003747R PKK003874R PKK005939R PKK007290R PKK007496R PKK050941R PKK053044R PKK053571R PKK054751R PKK055192R PKK003631R PKK005029R PKK005208R PKK005967R PKK053898R PKK004532R PKK053376R PKK055807R PKK004038R PKK004160R PKK005286R PKK006928R PKK007548R PKK012485R PKK012533R PKK012591R PKK012913R PKK014659R PKK014772R PKK017124R PKK017242R PKK017767R PKK018606R PKK056058R PKK100437R PKK003671R PKK050430R PKK050661R PKK052210R PKK052657R PKK054931R PKK005306R PKK002553R PKK002703R PKK003287R PKK003378R PKK003536R PKK003581R PKK003703R PKK003722R PKK003740R PKK003758R PKK003843R PKK004006R PKK004392R PKK004483R PKK004506R PKK004665R PKK004869R PKK005697R PKK005862R PKK005995R PKK006417R PKK006433R PKK006438R PKK006517R PKK006566R PKK006734R PKK006865R PKK006892R PKK007456R PKK007484R PKK007541R PKK007556R PKK007558R PKK007784R PKK010025R PKK010058R PKK010856R PKK011090R PKK011616R PKK012346R PKK013296R PKK013804R PKK013829R PKK013870R PKK013902R PKK014213R PKK014359R PKK014518R PKK014520R PKK014630R PKK014696R PKK015111R PKK016702R PKK016733R PKK016765R PKK017034R PKK017314R PKK017574R PKK018547R PKK018890R PKK018977R PKK019333R PKK019518R PKK019870R PKK004704R PKK004845R PKK005099R PKK007458R PKK007560R PKK011365R PKK012399R PKK012918R PKK013962R PKK014442R PKK016345R PKK016876R PKK018811R PKK051089R PKK055422R PKK002061R PKK002369R PKK003356R PKK003505R PKK003528R PKK003535R PKK003543R PKK003734R PKK003746R PKK003877R PKK003922R PKK003943R PKK003980R PKK003992R PKK003995R PKK004037R PKK004052R PKK004664R PKK004718R PKK004942R PKK005284R PKK005461R PKK005464R PKK005849R PKK006489R PKK006505R PKK006706R PKK006728R PKK006782R PKK006899R PKK006901R PKK007045R PKK007058R PKK007351R PKK007352R PKK007578R PKK007667R PKK007791R PKK010032R PKK010679R PKK010987R PKK011014R PKK011546R PKK012513R PKK013866R PKK013932R PKK014364R PKK014369R PKK014387R PKK015259R PKK015460R PKK015490R PKK015776R PKK016729R PKK017667R PKK017993R PKK018622R PKK018647R PKK018738R PKK019923R PKK020089R PKK050382R PKK050497R PKK050614R PKK050823R PKK051773R PKK052732R PKK055396R PKK006529R PKK011040R PKK019751R PKK020393R PKK051316R PKK100456R PKK205949R PKK208244R PKK208471R PKK101907R PKK107087R PKK055477R PKK001165R PKK050322R PKK052491R PKK054157R PKK102999R PKK102358R PKK200620R PKK007775R PKK203800R PKK006308R PKK200474R PKK100997R PKK105480R PKK051895R PKK004124R PKK106417R PKK100663R PKK202324R PKK203422R PKK052463R PKK201680R PKK201955R PKK014865R PKK200888R PKK211318R PKK211417R PKK109619R PKK056502R PKK207497R PKK105227R PKK107142R PKK107236R PKK204189R PKK103541R PKK011336R PKK011099R PKK208280R PKK206545R PKK205895R PKK006510R PKK006395R PKK003684R PKK208679R PKK208582R PKK004867R PKK005515R PKK054702R PKK051184R PKK108513R PKK108017R PKK101652R PKK107327R PKK202516R PKK105694R PKK016836R PKK106775R PKK003423R PKK105612R PKK205645R PKK050834R PKK055948R PKK054765R PKK011472R PKK204489R PKK105667R PKK055874R PKK004194R PKK050357R PKK103105R PKK003280R PKK005077R PKK202783R PKK202781R PKK106880R PKK006708R PKK055998R PKK201868R PKK105878R PKK208160R PKK020107R PKK016862R PKK201330R PKK016325R PKK014139R PKK201732R PKK055336R PKK208087R PKK051727R PKK210096R PKK210100R PKK055315R PKK202325R PKK055523R PKK015302R PKK055415R PKK002354R PKK103142R PKK010535R PKK002709R PKK010771R PKK052452R PKK208051R PKK004237R PKK020122R PKK203018R PKK105453R PKK050498R PKK201698R PKK109477R PKK106295R PKK106293R PKK013261R PKK204344R PKK201739R PKK010703R PKK100232R PKK107381R PKK107558R PKK108430R PKK206618R PKK001964R PKK100777R PKK101815R PKK108686R PKK209109R PKK002135R PKK002262R PKK010597R PKK010677R PKK011335R PKK011516R PKK011519R PKK011533R PKK011791R PKK013357R PKK021355R PKK104455R PKK104522R PKK104709R PKK200171R PKK200755R PKK201557R PKK202332R PKK203282R PKK203352R PKK203882R PKK204108R PKK205688R PKK209381R PKK210748R PKK211701R PKK211760R PKK005646R PKK013275R PKK205739R PKK207237R PKK001494R PKK016300R PKK016934R PKK053471R PKK101794R PKK107306R PKK202473R PKK202650R PKK205550R PKK207320R PKK007015R PKK011777R PKK017939R PKK054646R PKK106443R PKK209147R PKK107833R PKK211374R PKK001873R PKK003578R PKK010094R PKK013347R PKK053706R PKK054051R PKK054361R PKK101972R PKK105561R PKK108169R PKK109058R PKK200742R PKK200788R PKK204414R PKK205537R PKK209006R PKK013228R PKK206509R PKK209879R PKK010759R PKK103592R PKK109784R PKK206542R PKK210780R PKK021001R PKK105086R PKK206697R PKK207479R PKK102302R PKK108547R PKK109127R PKK010133R PKK013657R PKK202544R PKK203995R PKK010605R PKK013201R PKK101521R PKK001741R PKK007604R PKK105076R PKK203663R PKK007584R PKK207306R PKK014090R PKK016262R PKK103200R PKK103211R PKK104482R PKK109558R PKK202670R PKK005739R PKK011327R PKK052530R PKK004987R PKK051332R PKK051919R PKK005576R PKK105438R PKK014590R PKK207249R PKK017173R PKK104000R PKK002900R PKK002016R PKK012601R PKK208938R PKK209177R PKK105259R PKK004581R PKK200351R PKK203265R PKK107622R PKK004162R PKK103556R PKK011714R PKK054877R PKK101499R PKK105997R PKK205564R PKK004285R PKK103014R PKK005323R PKK005885R PKK011906R PKK016936R PKK105095R PKK201713R PKK013484R PKK016426R PKK210442R PKK002507R PKK200379R PKK013472R PKK002231R PKK007451R PKK102968R PKK006344R PKK202582R PKK206645R PKK053951R PKK211630R PKK105275R PKK005008R PKK005735R PKK104136R PKK055967R PKK200764R PKK053903R PKK103301R PKK201313R PKK014002R PKK056453R PKK200337R PKK002164R PKK100097R PKK102073R PKK108913R PKK001920R PKK012672R PKK101294R PKK016077R PKK054823R PKK105751R PKK016359R PKK109642R PKK200782R PKK205691R PKK205720R PKK209509R PKK211572R PKK103331R PKK050178R PKK106000R PKK203001R PKK202060R PKK055517R PKK104397R PKK106155R PKK011986R PKK211610R PKK051258R PKK012657R PKK003060R PKK017979R PKK201977R PKK203317R PKK021423R PKK209431R PKK002859R PKK206927R PKK102472R PKK002075R PKK102648R PKK011970R PKK100235R PKK017493R PKK018308R PKK012964R PKK015152R PKK103267R PKK003437R PKK004331R PKK007139R PKK007522R PKK011061R PKK011297R PKK011659R PKK106860R PKK200205R PKK102360R PKK202307R PKK051217R PKK053389R PKK004139R PKK004454R PKK101612R PKK203261R PKK200257R PKK101328R PKK108502R PKK206874R PKK012146R PKK107580R PKK202613R PKK006574R PKK109342R PKK105530R PKK005703R PKK017496R PKK021078R PKK011726R PKK100123R PKK100193R PKK101614R PKK102959R PKK109435R PKK109767R PKK109769R PKK200776R PKK201710R PKK210740R PKK105520R PKK018418R PKK016175R PKK013377R PKK013947R PKK005659R PKK013686R PKK107417R PKK010977R PKK107774R PKK012397R PKK014850R PKK002080R PKK002090R PKK012245R PKK005327R PKK205700R PKK011998R PKK002233R PKK002808R PKK004441R PKK004462R PKK004922R PKK007277R PKK010448R PKK010947R PKK012158R PKK054627R PKK001956R PKK002945R PKK005173R PKK005775R PKK005784R PKK005830R PKK006215R PKK007575R PKK007647R PKK010426R PKK011477R PKK011514R PKK012566R PKK013283R PKK013407R PKK014835R PKK015634R PKK001948R PKK003618R PKK005036R PKK005568R PKK011406R PKK011425R PKK011858R PKK013312R PKK013459R PKK013645R PKK051337R PKK201072R PKK003924R PKK006001R PKK013922R PKK053632R PKK005509R PKK013421R PKK013708R PKK005370R PKK015766R PKK011720R PKK004218R PKK055522R PKK005535R PKK014045R PKK016442R PKK017184R PKK020227R PKK103650R PKK103651R PKK021084R PKK010735R PKK001742R PKK002127R PKK002243R PKK004697R PKK203395R PKK207960R PKK205029R PKK205032R PKK205033R PKK053140R PKK051000R PKK202569R PKK211773R PKK051456R PKK053009R PKK001361R PKK203365R PKK208812R PKK053608R PKK052900R PKK053135R PKK109318R PKK007422R PKK018645R PKK003485R PKK107107R PKK005007R PKK053142R PKK053425R PKK053429R PKK053683R PKK053715R PKK108673R PKK021300R PKK204905R PKK005287R PKK206926R PKK002198R PKK020966R PKK201738R PKK017293R PKK106626R PKK006447R PKK105446R PKK017594R PKK007771R PKK103178R PKK002968R PKK101437R PKK021108R PKK055346R PKK104969R PKK020657R PKK007090R PKK055939R PKK010775R PKK006004R PKK006034R PKK104158R PKK104285R PKK052141R PKK005616R PKK208476R PKK100550R PKK020371R PKK013893R PKK017384R PKK109265R PKK018406R PKK100384R PKK015140R PKK013591R PKK205646R PKK019361R PKK016051R PKK019600R PKK019735R PKK002821R PKK006471R PKK015211R PKK005346R PKK018493R PKK020398R PKK053658R PK
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Reason for Recall:
- There is a potential for Medtronic programmer and remote monitoring software applications to display an inaccurate remaining longevity estimate for a subset of implanted cardiac device models. This issue does not impact device functionality.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Programmers & Remote Monitoring Software Apps: 2090 CareLink Programmer
Product Codes/Lot Numbers:
Serial Numbers: PKK051331R PKK200187R PKK209956R PKK207572R PKK209265R PKK016592R PKK202460R PKK019811R PKK209256R PKK007753R PKK015981R PKK016604R PKK017955R PKK018301R PKK019555R PKK019776R PKK051452R PKK052792R PKK204491R PKK204504R PKK204551R PKK204575R PKK204595R PKK204640R PKK204740R PKK203409R PKK006360R PKK202461R PKK018408R PKK019864R PKK204483R PKK016046R PKK016047R PKK019820R PKK020201R PKK010563R PKK209937R PKK209928R PKK050581R PKK209305R PKK211241R PKK209294R PKK201567R PKK206687R PKK211282R PKK208961R PKK209373R PKK202996R PKK210858R PKK202566R PKK211233R PKK205628R PKK209203R PKK211710R PKK100124R PKK211042R PKK210807R PKK200224R PKK208864R PKK209197R PKK208958R PKK209332R PKK211720R PKK016352R PKK019371R PKK211286R PKK202507R PKK205997R PKK205974R PKK202465R PKK017835R PKK018895R PKK201637R PKK201640R PKK201643R PKK201645R PKK202394R PKK202397R PKK202411R PKK202428R PKK202441R PKK202444R PKK202451R PKK202454R PKK202459R PKK202463R PKK202464R PKK202466R PKK202489R PKK202494R PKK202521R PKK202550R PKK202552R PKK202553R PKK202557R PKK202558R PKK202561R PKK202563R PKK202564R PKK202565R PKK202567R PKK204964R PKK204973R PKK204974R PKK204983R PKK204985R PKK205962R PKK205968R PKK205970R PKK205977R PKK205979R PKK205982R PKK205986R PKK205989R PKK205994R PKK205998R PKK205999R PKK206000R PKK207397R PKK207399R PKK207576R PKK207588R PKK207601R PKK207603R PKK207604R PKK207606R PKK207607R PKK207615R PKK207621R PKK207622R PKK207623R PKK207628R PKK207649R PKK207650R PKK208500R PKK208503R PKK208508R PKK208510R PKK211023R PKK211064R PKK211072R PKK211084R PKK211100R PKK211227R PKK211228R PKK211229R PKK211235R PKK211238R PKK211239R PKK211284R PKK211289R PKK211294R PKK211339R PKK211342R PKK203579R PKK107297R PKK108724R PKK108726R PKK108729R PKK109569R PKK201389R PKK201402R PKK051032R PKK051031R PKK010204R PKK106139R PKK106149R PKK106153R PKK106161R PKK106164R PKK106212R PKK106221R PKK106224R PKK106228R PKK054839R PKK054899R PKK054903R PKK006109R PKK011161R PKK011162R PKK012402R PKK012404R PKK020167R PKK052616R PKK052619R PKK208773R PKK206171R PKK108840R PKK205827R PKK018974R PKK054297R PKK052590R PKK055474R PKK019344R PKK208281R PKK020200R PKK056378R PKK107503R PKK106539R PKK005940R PKK201926R PKK202805R PKK205627R PKK003769R PKK016555R PKK004307R PKK001417R PKK006119R PKK005232R PKK004040R PKK013386R PKK019411R PKK052102R PKK018098R PKK101253R PKK205556R PKK102674R PKK006689R PKK200245R PKK106865R PKK211492R PKK017878R PKK010764R PKK211518R PKK211521R PKK057197R PKK011242R PKK200949R PKK051450R PKK018218R PKK057195R PKK201686R PKK020573R PKK014310R PKK054238R PKK209424R PKK208646R PKK208224R PKK020366R PKK207934R PKK053799R PKK108035R PKK204464R PKK051548R PKK203435R PKK014418R PKK102204R PKK050960R PKK108305R PKK001172R PKK102818R PKK205393R PKK013868R PKK053717R PKK053233R PKK002545R PKK202693R PKK002871R PKK106723R PKK055305R PKK108322R PKK102503R PKK202024R PKK003498R PKK004347R PKK015584R PKK105180R PKK003235R PKK014949R PKK018886R PKK019330R PKK104774R PKK014465R PKK054834R PKK055373R PKK101011R PKK106795R PKK205887R PKK205332R PKK002384R PKK015619R PKK055620R PKK003926R PKK006407R PKK007261R PKK014428R PKK017571R PKK019868R PKK050318R PKK107143R PKK006589R PKK018884R PKK103154R PKK107158R PKK204136R PKK015641R PKK104118R PKK001847R PKK002178R PKK053006R PKK017686R PKK211216R PKK201780R PKK015586R PKK101918R PKK006674R PKK013504R PKK052353R PKK100694R PKK102490R PKK106404R PKK003165R PKK007298R PKK052052R PKK105856R PKK100708R PKK052050R PKK102715R PKK006557R PKK101957R PKK002398R PKK004434R PKK006460R PKK101141R PKK003318R PKK017784R PKK107927R PKK209592R PKK052937R PKK103113R PKK103401R PKK107979R PKK102706R PKK103417R PKK051039R PKK108929R PKK002518R PKK018602R PKK014370R PKK050450R PKK018051R PKK018954R PKK006683R PKK015308R PKK206137R PKK017096R PKK201696R PKK013110R PKK053602R PKK103395R PKK104573R PKK106395R PKK106663R PKK107105R PKK108835R PKK019743R PKK002029R PKK006880R PKK007419R PKK052069R PKK012323R PKK051108R PKK207988R PKK053982R PKK055228R PKK201744R PKK002557R PKK208342R PKK013078R PKK014144R PKK014900R PKK055981R PKK103121R PKK001953R PKK005823R PKK007818R PKK005994R PKK015611R PKK051055R PKK056120R PKK056431R PKK014363R PKK014324R PKK017674R PKK019498R PKK003199R PKK005355R PKK007162R PKK103381R PKK013092R PKK056024R PKK056434R PKK103151R PKK104749R PKK106810R PKK107817R PKK010011R PKK206429R PKK004752R PKK003279R PKK011923R PKK208647R PKK004440R PKK208616R PKK202512R PKK208696R PKK002498R PKK208692R PKK101219R PKK006587R PKK018978R PKK055097R PKK207723R PKK106912R PKK203509R PKK201788R PKK006737R PKK206152R PKK011005R PKK015511R PKK012410R PKK015570R PKK019464R PKK013598R PKK001597R PKK007512R PKK018480R PKK100995R PKK003991R PKK004471R PKK105584R PKK001580R PKK011440R PKK053484R PKK002364R PKK010024R PKK050689R PKK006639R PKK014435R PKK011200R PKK205410R PKK055049R PKK102825R PKK104021R PKK014983R PKK017725R PKK050719R PKK005068R PKK107473R PKK017915R PKK104060R PKK208481R PKK107727R PKK015617R PKK006992R PKK109159R PKK109545R PKK019949R PKK012658R PKK018105R PKK104070R PKK108996R PKK103297R PKK003022R PKK001631R PKK001716R PKK018665R PKK019213R PKK020534R PKK109474R PKK012315R PKK202808R PKK104056R PKK109498R PKK200651R PKK208449R PKK003879R PKK056493R PKK016902R PKK105239R PKK017266R PKK015985R PKK108025R PKK014294R PKK200133R PKK103212R PKK101320R PKK101826R PKK211039R PKK106426R PKK211341R PKK211762R PKK210929R PKK011959R PKK001989R PKK003135R PKK007803R PKK002374R PKK054237R PKK207987R PKK201852R PKK055407R PKK050352R PKK015579R PKK019212R PKK055347R PKK014579R PKK012884R PKK011957R PKK014188R PKK010404R PKK209440R PKK019372R PKK106218R PKK106217R PKK208708R PKK207810R PKK051806R PKK108336R PKK108369R PKK050744R PKK013244R PKK106336R PKK206844R PKK001150R PKK003214R PKK207068R PKK104484R PKK103658R PKK011039R PKK020669R PKK055886R PKK053829R PKK051107R PKK106898R PKK211602R PKK208904R PKK054601R PKK211337R PKK208740R PKK106387R PKK050567R PKK051869R PKK014654R PKK109164R PKK102115R PKK015163R PKK107765R PKK202722R PKK211182R PKK015252R PKK202390R PKK205467R PKK109632R PKK005601R PKK201789R PKK211705R PKK015263R PKK052799R PKK050310R PKK208380R PKK211654R PKK011165R PKK211403R PKK201493R PKK203454R PKK211397R PKK012419R PKK012429R PKK013764R PKK054068R PKK054388R PKK054942R PKK056919R PKK056996R PKK109612R PKK108760R PKK050338R PKK209367R PKK017148R PKK108013R PKK102003R PKK210881R PKK210674R PKK209559R PKK108960R PKK004183R PKK011345R PKK016954R PKK004004R PKK210692R PKK005395R PKK102767R PKK205309R PKK105818R PKK011669R PKK105677R PKK106816R PKK208799R PKK210805R PKK101424R PKK100439R PKK101212R PKK103537R PKK209295R PKK016849R PKK017011R PKK017672R PKK017773R PKK018062R PKK107631R PKK206120R PKK010994R PKK210449R PKK019410R PKK108981R PKK208115R PKK208356R PKK101301R PKK019522R PKK204452R PKK211371R PKK210197R PKK210203R PKK100356R PKK001507R PKK018827R PKK019283R PKK019303R PKK019681R PKK006048R PKK015843R PKK106330R PKK208336R PKK056460R PKK204507R PKK013995R PKK005789R PKK106964R PKK201169R PKK200645R PKK207872R PKK004169R PKK007080R PKK100961R PKK015088R PKK014647R PKK017508R PKK005837R PKK100277R PKK006008R PKK006823R PKK004962R PKK206059R PKK204550R PKK202254R PKK208166R PKK202434R PKK051904R PKK101472R PKK201879R PKK056155R PKK107324R PKK006421R PKK109529R PKK015842R PKK107677R PKK051445R PKK014860R PKK007333R PKK012936R PKK002963R PKK204772R PKK103119R PKK103696R PKK011816R PKK017469R PKK205118R PKK054711R PKK002299R PKK055374R PKK001606R PKK003774R PKK204420R PKK100105R PKK011153R PKK204288R PKK003453R PKK006077R PKK007003R PKK007301R PKK016731R PKK019004R PKK051301R PKK051736R PKK100943R PKK206186R PKK209436R PKK202707R PKK200492R PKK002691R PKK003626R PKK101391R PKK200232R PKK207985R PKK102179R PKK203110R PKK208455R PKK202703R PKK016635R PKK001489R PKK208765R PKK011302R PKK208258R PKK055211R PKK019927R PKK016787R PKK103057R PKK015775R PKK013051R PKK016884R PKK108005R PKK209562R PKK100246R PKK056007R PKK206175R PKK051893R PKK005793R PKK102822R PKK106590R PKK017528R PKK016873R PKK201371R PKK005927R PKK019505R PKK106976R PKK107003R PKK107159R PKK107281R PKK107348R PKK107452R PKK107462R PKK107466R PKK108721R PKK210252R PKK205357R PKK013222R PKK056089R PKK006178R PKK011529R PKK107762R PKK056303R PKK001357R PKK054097R PKK203694R PKK105007R PKK107761R PKK014477R PKK018965R PKK003560R PKK056028R PKK056100R PKK018469R PKK021398R PKK010864R PKK006633R PKK201086R PKK002785R PKK012829R PKK102887R PKK103270R PKK020284R PKK054670R PKK202224R PKK011069R PKK010718R PKK209931R PKK101274R PKK012267R PKK010743R PKK202822R PKK010492R PKK209929R PKK010214R PKK105192R PKK209922R PKK204686R PKK201863R PKK208817R PKK002937R PKK107359R PKK107377R PKK010138R PKK107540R PKK205040R PKK108523R PKK002457R PKK053254R PKK205837R PKK201069R PKK020725R PKK108522R PKK210035R PKK210040R PKK005472R PKK204663R PKK054448R PKK004455R PKK209604R PKK051143R PKK052827R PKK050261R PKK014350R PKK209823R PKK015900R PKK017605R PKK018263R PKK106769R PKK107402R PKK202417R PKK014327R PKK209718R PKK052633R PKK102908R PKK211431R PKK211456R PKK201254R PKK207559R PKK051609R PKK205687R PKK103561R PKK203956R PKK106963R PKK013924R PKK004774R PKK200573R PKK014739R PKK107819R PKK050308R PKK204091R PKK207016R PKK201384R PKK104930R PKK103890R PKK012505R PKK053073R PKK054901R PKK004912R PKK106433R PKK202219R PKK100312R PKK205699R PKK108610R PKK101590R PKK103167R PKK208786R PKK010752R PKK109488R PKK052953R PKK014727R PKK207675R PKK053032R PKK004624R PKK200830R PKK015097R PKK010213R PKK106549R PKK206590R PKK016658R PKK211123R PKK208373R PKK201226R PKK056691R PKK053336R PKK204600R PKK100789R PKK103907R PKK012007R PKK019302R PKK051250R PKK108326R PKK053004R PKK004839R PKK018578R PKK015951R PKK052377R PKK203772R PKK002519R PKK001865R PKK013324R PKK200221R PKK202329R PKK054668R PKK013087R PKK006355R PKK006631R PKK207359R PKK051470R PKK054866R PKK108533R PKK202227R PKK207241R PKK055357R PKK200206R PKK010422R PKK055724R PKK205742R PKK206002R PKK055268R PKK101105R PKK011093R PKK101577R PKK202503R PKK200815R PKK001364R PKK051321R PKK001376R PKK020271R PKK013984R PKK208243R PKK107932R PKK015754R PKK107940R PKK108952R PKK004064R PKK106190R PKK051987R PKK102663R PKK108970R PKK205078R PKK204330R PKK202082R PKK018740R PKK101128R PKK054029R PKK014466R PKK002363R PKK202069R PKK207917R PKK205478R PKK103971R PKK019026R PKK205106R PKK106843R PKK007838R PKK208408R PKK202997R PKK052609R PKK106945R PKK209647R PKK014786R PKK017583R PKK106282R PKK005904R PKK106821R PKK101492R PKK055343R PKK108778R PKK101071R PKK208378R PKK007813R PKK101067R PKK109005R PKK204331R PKK104929R PKK200747R PKK104386R PKK105798R PKK052338R PKK100161R PKK205733R PKK011812R PKK108868R PKK004206R PKK014899R PKK205131R PKK054687R PKK202750R PKK052470R PKK010576R PKK011817R PKK200247R PKK104452R PKK107155R PKK207706R PKK107913R PKK054898R PKK051070R PKK021137R PKK015047R PKK015770R PKK051511R PKK006787R PKK105792R PKK050580R PKK054776R PKK201766R PKK018422R PKK006733R PKK208103R PKK207199R PKK209244R PKK208544R PKK018813R PKK209736R PKK005647R PKK018928R PKK012484R PKK105032R PKK052648R PKK102660R PKK002477R PKK054772R PKK207858R PKK006237R PKK013653R PKK053347R PKK108891R PKK203497R PKK002388R PKK013955R PKK020601R PKK012369R PKK106566R PKK211734R PKK018159R PKK109556R PKK021429R PKK015565R PKK207335R PKK014115R PKK004547R PKK101910R PKK200676R PKK014558R PKK050472R PKK105646R PKK102753R PKK013553R PKK002094R PKK106791R PKK206195R PKK106730R PKK206405R PKK105071R PKK205644R PKK052886R PKK052497R PKK056005R PKK014394R PKK204449R PKK007628R PKK106206R PKK050858R PKK054267R PKK056391R PKK207958R PKK011369R PKK007388R PKK018019R PKK106088R PKK015125R PKK005086R PKK014202R PKK019839R PKK002522R PKK007270R PKK051960R PKK013740R PKK108613R PKK051120R PKK055800R PKK052997R PKK105300R PKK018680R PKK108601R PKK202334R PKK208989R PKK209800R PKK201598R PKK013558R PKK209361R PKK051199R PKK012650R PKK208919R PKK211070R PKK050573R PKK020538R PKK202554R PKK005627R PKK109313R PKK109548R PKK108649R PKK054715R PKK012094R PKK002144R PKK054252R PKK054409R PKK053256R PKK010558R PKK017042R PKK013976R PKK200969R PKK014705R PKK006328R PKK200910R PKK051242R PKK006202R PKK103305R PKK021465R PKK201822R PKK003174R PKK055598R PKK005283R PKK100340R PKK101288R PKK105117R PKK208874R PKK204897R PKK055802R PKK205134R PKK205967R PKK104528R PKK205528R PKK203756R PKK203224R PKK203615R PKK020793R PKK106665R PKK203866R PKK019862R PKK050817R PKK208612R PKK019853R PKK108156R PKK053482R PKK108641R PKK107356R PKK104988R PKK207082R PKK102383R PKK052432R PKK006128R PKK053120R PKK104803R PKK020678R PKK208640R PKK108995R PKK014092R PKK007342R PKK204106R PKK014110R PKK001738R PKK101024R PKK005666R PKK201840R PKK201111R PKK203964R PKK001339R PKK002352R PKK002508R PKK002535R PKK002836R PKK002923R PKK003668R PKK003892R PKK003958R PKK004610R PKK005375R PKK006270R PKK006445R PKK006548R PKK006724R PKK006853R PKK007253R PKK010847R PKK012019R PKK012325R PKK014252R PKK014258R PKK014303R PKK014515R PKK015487R PKK015499R PKK015650R PKK016273R PKK016307R PKK016312R PKK017035R PKK018900R PKK019459R PKK019765R PKK020281R PKK020407R PKK050166R PKK050173R PKK050241R PKK050298R PKK050372R PKK050478R PKK050574R PKK050660R PKK050762R PKK050767R PKK051015R PKK051131R PKK051266R PKK051613R PKK051677R PKK051741R PKK052082R PKK052283R PKK052659R PKK052994R PKK053008R PKK053170R PKK053719R PKK053747R PKK053771R PKK053792R PKK055006R PKK055030R PKK055250R PKK055267R PKK056379R PKK003118R PKK006089R PKK006770R PKK006900R PKK007819R PKK011009R PKK011285R PKK014717R PKK015683R PKK018707R PKK004021R PKK005090R PKK007303R PKK101086R PKK106134R PKK054727R PKK006325R PKK018046R PKK050906R PKK100590R PKK203438R PKK204017R PKK100397R PKK018302R PKK018307R PKK050781R PKK021301R PKK011441R PKK011036R PKK101206R PKK203938R PKK003057R PKK055070R PKK054140R PKK006523R PKK204713R PKK109595R PKK202531R PKK202605R PKK200222R PKK021323R PKK016083R PKK211566R PKK100672R PKK108637R PKK003533R PKK053615R PKK106875R PKK201821R PKK011045R PKK107412R PKK200497R PKK206133R PKK052026R PKK055304R PKK001516R PKK005442R PKK005392R PKK016410R PKK211664R PKK211706R PKK020787R PKK210510R PKK210512R PKK057223R PKK001963R PKK016206R PKK021117R PKK205734R PKK052910R PKK206020R PKK014204R PKK200708R PKK012367R PKK014595R PKK107686R PKK202201R PKK054220R PKK011851R PKK210657R PKK011633R PKK051638R PKK021304R PKK100132R PKK007448R PKK203377R PKK207326R PKK207015R PKK210884R PKK006765R PKK101340R PKK107248R PKK204215R PKK207996R PKK205614R PKK209677R PKK209826R PKK209684R PKK211686R PKK001684R PKK005493R PKK205116R PKK205019R PKK205114R PKK207264R PKK016789R PKK201704R PKK013848R PKK054305R PKK200627R PKK103306R PKK010668R PKK207487R PKK001589R PKK206042R PKK205959R PKK050866R PKK051785R PKK102313R PKK205560R PKK013845R PKK102284R PKK053072R PKK207743R PKK004794R PKK004467R PKK203505R PKK007011R PKK013043R PKK207742R PKK053259R PKK002752R PKK015594R PKK208593R PKK013822R PKK054927R PKK101354R PKK054730R PKK005428R PKK013906R PKK019760R PKK205420R PKK209547R PKK205992R PKK017960R PKK051423R PKK004399R PKK051348R PKK101390R PKK052977R PKK052805R PKK051045R PKK108050R PKK001893R PKK054880R PKK206505R PKK005595R PKK209805R PKK104203R PKK200847R PKK101745R PKK013014R PKK208821R PKK052457R PKK053296R PKK208437R PKK206553R PKK014138R PKK105741R PKK201983R PKK054851R PKK017401R PKK106925R PKK204959R PKK209790R PKK011042R PKK208648R PKK208663R PKK206451R PKK050599R PKK204195R PKK109614R PKK017873R PKK108249R PKK205494R PKK208887R PKK013467R PKK019389R PKK208955R PKK208972R PKK051155R PKK207307R PKK207861R PKK021271R PKK200541R PKK209419R PKK017745R PKK201139R PKK202863R PKK209769R PKK007808R PKK104205R PKK206503R PKK017152R PKK012933R PKK108330R PKK107562R PKK011647R PKK050847R PKK050852R PKK050975R PKK002967R PKK054754R PKK050986R PKK104585R PKK102087R PKK015693R PKK051110R PKK200356R PKK001849R PKK202341R PKK016691R PKK017679R PKK016978R PKK017626R PKK104589R PKK101744R PKK206113R PKK208199R PKK014445R PKK207179R PKK102101R PKK053689R PKK206684R PKK005798R PKK201897R PKK208712R PKK053224R PKK206460R PKK207894R PKK015543R PKK105763R PKK016998R PKK203476R PKK014302R PKK055223R PKK014272R PKK054870R PKK103354R PKK052969R PKK100829R PKK107599R PKK211868R PKK013986R PKK017937R PKK107589R PKK202067R PKK211232R PKK011339R PKK010811R PKK209633R PKK020298R PKK203408R PKK109403R PKK209949R PKK105178R PKK005741R PKK010459R PKK206725R PKK013665R PKK207037R PKK107844R PKK106546R PKK109386R PKK017545R PKK108057R PKK051576R PKK204058R PKK107394R PKK011563R PKK019904R PKK012002R PKK209642R PKK102096R PKK016799R PKK016261R PKK104871R PKK203335R PKK209501R PKK013961R PKK105647R PKK105650R PKK207590R PKK203482R PKK014236R PKK206870R PKK202371R PKK200763R PKK201238R PKK201729R PKK200793R PKK012828R PKK050359R PKK011607R PKK107352R PKK103582R PKK021294R PKK209340R PKK208362R PKK004113R PKK051398R PKK208337R PKK002044R PKK012409R PKK104535R PKK014643R PKK002370R PKK002669R PKK002850R PKK004381R PKK011505R PKK012363R PKK013004R PKK013121R PKK014925R PKK017026R PKK017607R PKK018999R PKK019007R PKK019503R PKK004076R PKK004716R PKK010057R PKK011872R PKK017708R PKK017711R PKK018027R PKK018966R PKK051430R PKK054740R PKK054884R PKK055627R PKK018032R PKK004605R PKK004813R PKK014080R PKK014166R PKK014624R PKK014792R PKK015561R PKK017893R PKK018670R PKK020119R PKK001068R PKK002367R PKK002525R PKK003468R PKK014589R PKK003503R PKK005075R PKK005244R PKK006115R PKK006558R PKK007766R PKK011588R PKK013032R PKK014297R PKK015005R PKK015553R PKK017620R PKK050621R PKK051261R PKK054357R PKK053319R PKK003301R PKK003904R PKK010087R PKK015229R PKK050412R PKK052535R PKK053083R PKK002322R PKK003038R PKK006085R PKK010474R PKK011230R PKK011868R PKK013293R PKK019742R PKK050465R PKK050526R PKK052266R PKK052562R PKK054258R PKK056381R PKK005186R PKK006719R PKK013897R PKK014460R PKK018595R PKK003506R PKK006655R PKK010016R PKK011610R PKK016374R PKK052355R PKK055194R PKK055288R PKK056126R PKK002823R PKK004020R PKK006399R PKK012204R PKK013329R PKK014227R PKK014307R PKK016671R PKK050512R PKK051756R PKK051781R PKK053084R PKK053279R PKK004046R PKK006017R PKK007249R PKK013890R PKK016302R PKK018872R PKK002466R PKK003274R PKK004134R PKK004486R PKK005268R PKK007401R PKK007737R PKK014377R PKK003310R PKK004023R PKK004904R PKK006195R PKK006494R PKK013500R PKK014254R PKK019016R PKK002137R PKK003284R PKK004349R PKK005815R PKK011296R PKK014619R PKK016649R PKK052109R PKK052398R PKK054153R PKK055036R PKK004395R PKK005256R PKK006550R PKK013015R PKK013556R PKK014372R PKK015502R PKK017759R PKK018768R PKK050214R PKK053065R PKK054232R PKK054894R PKK055051R PKK055836R PKK003190R PKK003388R PKK004379R PKK004592R PKK004597R PKK004928R PKK005153R PKK005171R PKK005926R PKK006427R PKK007436R PKK011176R PKK011451R PKK012559R PKK013079R PKK014107R PKK015030R PKK015564R PKK015569R PKK017769R PKK017932R PKK018982R PKK019412R PKK019474R PKK019506R PKK051853R PKK053553R PKK054324R PKK003366R PKK006867R PKK010023R PKK010772R PKK011359R PKK012839R PKK014059R PKK014205R PKK014492R PKK017662R PKK018759R PKK020087R PKK051109R PKK052901R PKK002786R PKK003102R PKK003226R PKK003351R PKK004056R PKK004304R PKK004305R PKK005906R PKK006752R PKK006869R PKK007532R PKK007725R PKK012105R PKK013824R PKK015602R PKK018859R PKK002283R PKK003869R PKK004326R PKK005600R PKK006891R PKK006973R PKK006999R PKK013973R PKK014309R PKK018598R PKK018678R PKK054913R PKK005935R PKK013374R PKK003170R PKK015273R PKK015286R PKK015453R PKK050366R PKK051798R PKK002805R PKK003096R PKK003209R PKK003317R PKK003602R PKK003989R PKK004872R PKK006534R PKK006897R PKK007530R PKK007805R PKK010370R PKK011584R PKK013892R PKK017591R PKK017737R PKK018832R PKK051804R PKK052578R PKK052876R PKK056056R PKK003225R PKK007807R PKK010820R PKK012586R PKK015538R PKK017589R PKK019907R PKK005943R PKK006035R PKK007339R PKK010079R PKK013530R PKK014177R PKK018725R PKK019511R PKK004072R PKK011206R PKK014242R PKK019942R PKK054766R PKK050501R PKK006945R PKK006949R PKK007223R PKK016344R PKK018609R PKK050170R PKK050795R PKK052566R PKK052580R PKK006581R PKK006908R PKK007489R PKK011549R PKK012415R PKK013081R PKK017220R PKK051754R PKK054912R PKK054933R PKK001408R PKK005991R PKK005275R PKK003906R PKK004417R PKK004698R PKK005965R PKK007538R PKK011262R PKK012262R PKK056018R PKK007286R PKK014289R PKK014750R PKK014984R PKK016712R PKK018043R PKK019443R PKK019668R PKK050704R PKK050848R PKK050934R PKK051772R PKK052762R PKK053920R PKK055119R PKK002114R PKK004117R PKK004371R PKK004934R PKK005301R PKK006225R PKK006277R PKK006431R PKK006473R PKK006722R PKK006822R PKK007267R PKK007497R PKK010008R PKK011238R PKK013489R PKK013534R PKK002337R PKK002355R PKK003124R PKK003267R PKK003413R PKK003518R PKK003942R PKK003176R PKK006220R PKK050616R PKK053590R PKK003645R PKK016921R PKK050421R PKK052905R PKK053260R PKK053335R PKK003041R PKK050458R PKK050477R PKK050663R PKK051179R PKK052179R PKK052593R PKK053679R PKK054684R PKK005089R PKK005903R PKK006442R PKK050230R PKK052887R PKK052943R PKK053992R PKK054158R PKK054825R PKK056132R PKK051411R PKK002058R PKK005840R PKK006366R PKK010369R PKK053904R PKK055248R PKK010030R PKK011013R PKK011051R PKK011863R PKK019531R PKK019902R PKK020148R PKK050297R PKK050431R PKK050572R PKK051528R PKK054705R PKK002328R PKK003178R PKK003632R PKK004621R PKK005163R PKK006348R PKK006430R PKK006670R PKK006750R PKK006816R PKK006819R PKK010381R PKK010767R PKK010785R PKK010944R PKK011053R PKK012413R PKK013704R PKK013809R PKK014958R PKK014972R PKK015029R PKK015662R PKK016284R PKK016289R PKK016298R PKK016343R PKK016347R PKK016759R PKK016904R PKK017627R PKK017771R PKK018100R PKK018607R PKK018629R PKK019247R PKK051323R PKK052073R PKK052348R PKK052469R PKK002314R PKK002733R PKK002867R PKK003677R PKK003747R PKK003874R PKK005939R PKK007290R PKK007496R PKK050941R PKK053044R PKK053571R PKK054751R PKK055192R PKK003631R PKK005029R PKK005208R PKK005967R PKK053898R PKK004532R PKK053376R PKK055807R PKK004038R PKK004160R PKK005286R PKK006928R PKK007548R PKK012485R PKK012533R PKK012591R PKK012913R PKK014659R PKK014772R PKK017124R PKK017242R PKK017767R PKK018606R PKK056058R PKK100437R PKK003671R PKK050430R PKK050661R PKK052210R PKK052657R PKK054931R PKK005306R PKK002553R PKK002703R PKK003287R PKK003378R PKK003536R PKK003581R PKK003703R PKK003722R PKK003740R PKK003758R PKK003843R PKK004006R PKK004392R PKK004483R PKK004506R PKK004665R PKK004869R PKK005697R PKK005862R PKK005995R PKK006417R PKK006433R PKK006438R PKK006517R PKK006566R PKK006734R PKK006865R PKK006892R PKK007456R PKK007484R PKK007541R PKK007556R PKK007558R PKK007784R PKK010025R PKK010058R PKK010856R PKK011090R PKK011616R PKK012346R PKK013296R PKK013804R PKK013829R PKK013870R PKK013902R PKK014213R PKK014359R PKK014518R PKK014520R PKK014630R PKK014696R PKK015111R PKK016702R PKK016733R PKK016765R PKK017034R PKK017314R PKK017574R PKK018547R PKK018890R PKK018977R PKK019333R PKK019518R PKK019870R PKK004704R PKK004845R PKK005099R PKK007458R PKK007560R PKK011365R PKK012399R PKK012918R PKK013962R PKK014442R PKK016345R PKK016876R PKK018811R PKK051089R PKK055422R PKK002061R PKK002369R PKK003356R PKK003505R PKK003528R PKK003535R PKK003543R PKK003734R PKK003746R PKK003877R PKK003922R PKK003943R PKK003980R PKK003992R PKK003995R PKK004037R PKK004052R PKK004664R PKK004718R PKK004942R PKK005284R PKK005461R PKK005464R PKK005849R PKK006489R PKK006505R PKK006706R PKK006728R PKK006782R PKK006899R PKK006901R PKK007045R PKK007058R PKK007351R PKK007352R PKK007578R PKK007667R PKK007791R PKK010032R PKK010679R PKK010987R PKK011014R PKK011546R PKK012513R PKK013866R PKK013932R PKK014364R PKK014369R PKK014387R PKK015259R PKK015460R PKK015490R PKK015776R PKK016729R PKK017667R PKK017993R PKK018622R PKK018647R PKK018738R PKK019923R PKK020089R PKK050382R PKK050497R PKK050614R PKK050823R PKK051773R PKK052732R PKK055396R PKK006529R PKK011040R PKK019751R PKK020393R PKK051316R PKK100456R PKK205949R PKK208244R PKK208471R PKK101907R PKK107087R PKK055477R PKK001165R PKK050322R PKK052491R PKK054157R PKK102999R PKK102358R PKK200620R PKK007775R PKK203800R PKK006308R PKK200474R PKK100997R PKK105480R PKK051895R PKK004124R PKK106417R PKK100663R PKK202324R PKK203422R PKK052463R PKK201680R PKK201955R PKK014865R PKK200888R PKK211318R PKK211417R PKK109619R PKK056502R PKK207497R PKK105227R PKK107142R PKK107236R PKK204189R PKK103541R PKK011336R PKK011099R PKK208280R PKK206545R PKK205895R PKK006510R PKK006395R PKK003684R PKK208679R PKK208582R PKK004867R PKK005515R PKK054702R PKK051184R PKK108513R PKK108017R PKK101652R PKK107327R PKK202516R PKK105694R PKK016836R PKK106775R PKK003423R PKK105612R PKK205645R PKK050834R PKK055948R PKK054765R PKK011472R PKK204489R PKK105667R PKK055874R PKK004194R PKK050357R PKK103105R PKK003280R PKK005077R PKK202783R PKK202781R PKK106880R PKK006708R PKK055998R PKK201868R PKK105878R PKK208160R PKK020107R PKK016862R PKK201330R PKK016325R PKK014139R PKK201732R PKK055336R PKK208087R PKK051727R PKK210096R PKK210100R PKK055315R PKK202325R PKK055523R PKK015302R PKK055415R PKK002354R PKK103142R PKK010535R PKK002709R PKK010771R PKK052452R PKK208051R PKK004237R PKK020122R PKK203018R PKK105453R PKK050498R PKK201698R PKK109477R PKK106295R PKK106293R PKK013261R PKK204344R PKK201739R PKK010703R PKK100232R PKK107381R PKK107558R PKK108430R PKK206618R PKK001964R PKK100777R PKK101815R PKK108686R PKK209109R PKK002135R PKK002262R PKK010597R PKK010677R PKK011335R PKK011516R PKK011519R PKK011533R PKK011791R PKK013357R PKK021355R PKK104455R PKK104522R PKK104709R PKK200171R PKK200755R PKK201557R PKK202332R PKK203282R PKK203352R PKK203882R PKK204108R PKK205688R PKK209381R PKK210748R PKK211701R PKK211760R PKK005646R PKK013275R PKK205739R PKK207237R PKK001494R PKK016300R PKK016934R PKK053471R PKK101794R PKK107306R PKK202473R PKK202650R PKK205550R PKK207320R PKK007015R PKK011777R PKK017939R PKK054646R PKK106443R PKK209147R PKK107833R PKK211374R PKK001873R PKK003578R PKK010094R PKK013347R PKK053706R PKK054051R PKK054361R PKK101972R PKK105561R PKK108169R PKK109058R PKK200742R PKK200788R PKK204414R PKK205537R PKK209006R PKK013228R PKK206509R PKK209879R PKK010759R PKK103592R PKK109784R PKK206542R PKK210780R PKK021001R PKK105086R PKK206697R PKK207479R PKK102302R PKK108547R PKK109127R PKK010133R PKK013657R PKK202544R PKK203995R PKK010605R PKK013201R PKK101521R PKK001741R PKK007604R PKK105076R PKK203663R PKK007584R PKK207306R PKK014090R PKK016262R PKK103200R PKK103211R PKK104482R PKK109558R PKK202670R PKK005739R PKK011327R PKK052530R PKK004987R PKK051332R PKK051919R PKK005576R PKK105438R PKK014590R PKK207249R PKK017173R PKK104000R PKK002900R PKK002016R PKK012601R PKK208938R PKK209177R PKK105259R PKK004581R PKK200351R PKK203265R PKK107622R PKK004162R PKK103556R PKK011714R PKK054877R PKK101499R PKK105997R PKK205564R PKK004285R PKK103014R PKK005323R PKK005885R PKK011906R PKK016936R PKK105095R PKK201713R PKK013484R PKK016426R PKK210442R PKK002507R PKK200379R PKK013472R PKK002231R PKK007451R PKK102968R PKK006344R PKK202582R PKK206645R PKK053951R PKK211630R PKK105275R PKK005008R PKK005735R PKK104136R PKK055967R PKK200764R PKK053903R PKK103301R PKK201313R PKK014002R PKK056453R PKK200337R PKK002164R PKK100097R PKK102073R PKK108913R PKK001920R PKK012672R PKK101294R PKK016077R PKK054823R PKK105751R PKK016359R PKK109642R PKK200782R PKK205691R PKK205720R PKK209509R PKK211572R PKK103331R PKK050178R PKK106000R PKK203001R PKK202060R PKK055517R PKK104397R PKK106155R PKK011986R PKK211610R PKK051258R PKK012657R PKK003060R PKK017979R PKK201977R PKK203317R PKK021423R PKK209431R PKK002859R PKK206927R PKK102472R PKK002075R PKK102648R PKK011970R PKK100235R PKK017493R PKK018308R PKK012964R PKK015152R PKK103267R PKK003437R PKK004331R PKK007139R PKK007522R PKK011061R PKK011297R PKK011659R PKK106860R PKK200205R PKK102360R PKK202307R PKK051217R PKK053389R PKK004139R PKK004454R PKK101612R PKK203261R PKK200257R PKK101328R PKK108502R PKK206874R PKK012146R PKK107580R PKK202613R PKK006574R PKK109342R PKK105530R PKK005703R PKK017496R PKK021078R PKK011726R PKK100123R PKK100193R PKK101614R PKK102959R PKK109435R PKK109767R PKK109769R PKK200776R PKK201710R PKK210740R PKK105520R PKK018418R PKK016175R PKK013377R PKK013947R PKK005659R PKK013686R PKK107417R PKK010977R PKK107774R PKK012397R PKK014850R PKK002080R PKK002090R PKK012245R PKK005327R PKK205700R PKK011998R PKK002233R PKK002808R PKK004441R PKK004462R PKK004922R PKK007277R PKK010448R PKK010947R PKK012158R PKK054627R PKK001956R PKK002945R PKK005173R PKK005775R PKK005784R PKK005830R PKK006215R PKK007575R PKK007647R PKK010426R PKK011477R PKK011514R PKK012566R PKK013283R PKK013407R PKK014835R PKK015634R PKK001948R PKK003618R PKK005036R PKK005568R PKK011406R PKK011425R PKK011858R PKK013312R PKK013459R PKK013645R PKK051337R PKK201072R PKK003924R PKK006001R PKK013922R PKK053632R PKK005509R PKK013421R PKK013708R PKK005370R PKK015766R PKK011720R PKK004218R PKK055522R PKK005535R PKK014045R PKK016442R PKK017184R PKK020227R PKK103650R PKK103651R PKK021084R PKK010735R PKK001742R PKK002127R PKK002243R PKK004697R PKK203395R PKK207960R PKK205029R PKK205032R PKK205033R PKK053140R PKK051000R PKK202569R PKK211773R PKK051456R PKK053009R PKK001361R PKK203365R PKK208812R PKK053608R PKK052900R PKK053135R PKK109318R PKK007422R PKK018645R PKK003485R PKK107107R PKK005007R PKK053142R PKK053425R PKK053429R PKK053683R PKK053715R PKK108673R PKK021300R PKK204905R PKK005287R PKK206926R PKK002198R PKK020966R PKK201738R PKK017293R PKK106626R PKK006447R PKK105446R PKK017594R PKK007771R PKK103178R PKK002968R PKK101437R PKK021108R PKK055346R PKK104969R PKK020657R PKK007090R PKK055939R PKK010775R PKK006004R PKK006034R PKK104158R PKK104285R PKK052141R PKK005616R PKK208476R PKK100550R PKK020371R PKK013893R PKK017384R PKK109265R PKK018406R PKK100384R PKK015140R PKK013591R PKK205646R PKK019361R PKK016051R PKK019600R PKK019735R PKK002821R PKK006471R PKK015211R PKK005346R PKK018493R PKK020398R PKK053658R PK
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0239-2020
Related Recalls
ICD-DR DDBB2D1 EVERA XT IS1/DF1 INTL, Model Number DDBB2D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD DDPA2D1G COBALT XT DR MRI DF1 GOLD, Model Number DDPA2D1G; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.