VirtuoSaph Plus Endoscopic Vessel Harvesting System, Catalog Number: VSP550EX Product Usage: The VittuoSaph¿ Plus Endoscopic Vessel Harvesting System VSP550EX is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surge1y for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requit¿ing blunt dissection of tissue including dissection of blood vessels and dissection of blood vessels of the extremities. Extremity procedures include tissue dissection and/or vessel harvesting along the saphenous vein for coronary artery bypass grafting and peripheral artery bypass grafting or radial artery for use in coronary artery bypass grafting vc
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Numbers: 81K
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Terumo Cardiovascular Systems Corporation
- Reason for Recall:
- The system's IFU contains a list of generators that the have been deemed compatible. Recent reports indicate that one energy platform is not compatible with the system. Therefore, it has been removed from the list of compatible generators within the IFU.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
VirtuoSaph Plus Endoscopic Vessel Harvesting System, Catalog Number: VSP550EX Product Usage: The VittuoSaph¿ Plus Endoscopic Vessel Harvesting System VSP550EX is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surge1y for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requit¿ing blunt dissection of tissue including dissection of blood vessels and dissection of blood vessels of the extremities. Extremity procedures include tissue dissection and/or vessel harvesting along the saphenous vein for coronary artery bypass grafting and peripheral artery bypass grafting or radial artery for use in coronary artery bypass grafting vc
Product Codes/Lot Numbers:
Lot Numbers: 81K
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0227-2019
Related Recalls
CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial Kit
Terumo Cardiovascular Systems
Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.
Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754
Terumo Cardiovascular Systems
During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be triggered to the user, due to the dashed-out value.
Pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmHg versus 400mmHg, potential myocardial tissue damage potentially requiring surgical or medical/pharmacological intervention