SideFire Directional Laser Probe, REF: NBP101-01, NBP201-01, NBP301-01, NBP401-01, NBP501-01. These laser probes are part of the NeuroBlate System, a unique collection of MRI-compatible laser devices and accessories that create an MRI guided delivery of precision thermal therapy in the brain. The NeuroBlate System components include Gas-cooled Laser Delivery Probes to deliver controlled energy to a target zone.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    serial numbers: 20368-1-C1112-01, 20368-1-C1112-02, 20368-1-C1112-03, 20368-1-C1112-04, 20368-1-C1112-05, 20368-1-C1112-06, 20368-1-C1112-07, 20368-1-C1112-08, 20368-1-C1112-09, 20368-1-C3173-02, 20368-1-C3173-03, 20368-1-C3173-04, 20368-1-C3173-05, 20368-1-C3173-06, 20368-1-C3173-07, 20368-1-C3173-08, 20368-1-C3173-09, 20368-1-C3173-10, 20368-1-C3173-11, 20368-1-C3173-12, 20368-1-C3173-13, 20368-1-C3173-14, 20368-1-C3173-15, 20368-1-C3173-16, 20368-1-C3173-17, 20368-1-C3173-18, 20368-1-C3173-19, 20368-1-C3173-20, 20368-1-C3258-01, 20368-1-C3258-02, 20368-1-C3258-04, 20368-1-C3258-05, 20368-1-C3258-06, 20368-1-C3258-07, 20368-1-C3258-12, 20368-1-C3258-13, 20368-1-C3258-14, 20368-1-C3462-01, 20368-1-C3462-02, 20368-1-C3462-03, 20368-1-C3462-04, 20368-1-C3462-05, 20368-1-C3462-06, 20368-1-C3462-07, 20368-1-C3462-08, 20368-1-C3462-09, 20368-1-C3462-10, 20368-1-C3462-11, 20368-1-C3462-12, 20368-1-C3462-13, 20368-1-C3462-14, 20368-1-C3462-16, 20368-1-C3462-17, 20368-1-C3462-18, 20368-1-C3462-19, 20368-1-C3462-22, 20368-1-C3462-24, 20368-1-C3462-25, 20368-1-C3756-01, 20368-1-C3756-02, 20368-1-C3756-04, 20368-1-C3756-05, 20368-1-C3756-06, 20368-1-C3756-07, 20368-1-C3756-08, 20368-1-C3756-11, 20368-1-C3756-12, 20368-2-C1276-02, 20368-2-C1276-04, 20368-2-C1276-05, 20368-2-C1276-06, 20368-2-C1276-07, 20368-2-C1276-08, 20368-2-C1276-09, 20368-2-C1276-11, 20368-2-C1276-12, 20368-2-C1276-13, 20368-2-C1276-14, 20368-2-C1276-15, 20368-2-C1276-16, 20368-2-C1276-17, 20368-2-C1276-18, 20368-2-C1276-19, 20368-2-C1276-20, 20368-2-C1276-21, 20368-2-C1276-22, 20368-2-C3287-01, 20368-2-C3287-02, 20368-2-C3287-03, 20368-2-C3287-04, 20368-2-C3287-05, 20368-2-C3287-06, 20368-2-C3287-07, 20368-2-C3287-08, 20368-2-C3287-09, 20368-2-C3287-10, 20368-2-C3287-11, 20368-2-C3287-12, 20368-2-C3287-13, 20368-2-C3287-14, 20368-2-C3287-15, 20368-2-C3287-16, 20368-2-C3287-17, 20368-2-C3287-19, 20368-2-C3287-20, 20368-2-C3287-21, 20368-2-C3287-22, 20368-2-C3472-01, 20368-2-C3472-02, 20368-2-C3472-03, 20368-2-C3472-04, 20368-2-C3472-05, 20368-2-C3472-06, 20368-2-C3472-07, 20368-2-C3472-08, 20368-2-C3472-09, 20368-2-C3472-10, 20368-2-C3472-11, 20368-3-C1115-01, 20368-3-C1115-02, 20368-3-C1115-03, 20368-3-C1115-04, 20368-3-C1115-05, 20368-3-C1115-06, 20368-3-C1115-07, 20368-3-C1115-08, 20368-3-C1115-09, 20368-3-C1115-10, 20368-3-C1116-01, 20368-3-C1116-02, 20368-3-C1116-04, 20368-3-C1116-05, 20368-3-C1116-06, 20368-3-C3293-01, 20368-3-C3293-02, 20368-3-C3293-03, 20368-3-C3293-04, 20368-3-C3293-05, 20368-3-C3293-07, 20368-3-C3293-10, 20368-3-C3293-11, 20368-3-C3293-13, 20368-3-C3293-14, 20368-3-C3293-15, 20368-3-C3293-16, 20368-3-C3293-17, 20368-3-C3293-19, 20368-3-C3293-20, 20368-3-C3293-21, 20368-3-C3293-23, 20368-3-C3452-04, 20368-3-C3452-06, 20368-3-C3452-07, 20368-3-C3452-09, 20368-3-C3452-10, 20368-3-C3452-11, 20368-3-C3452-13, 20368-3-C3452-14, 20368-3-C3452-15, 20368-3-C3452-17, 20368-3-C3452-18, 20368-3-C3452-20, 20368-3-C3452-21, 20368-3-C3452-22, 20368-3-C3452-24, 20368-3-C3452-25, 20368-3-C3512-01, 20368-3-C3512-02, 20368-3-C3512-03, 20368-3-C3512-04, 20368-3-C3512-05, 20368-3-C3512-06, 20368-3-C3512-08, 20368-4-C1117-01, 20368-4-C1117-02, 20368-4-C1117-03, 20368-4-C1117-05, 20368-4-C1117-06, 20368-4-C1117-08, 20368-4-C1117-09, 20368-4-C1117-10, 20368-4-C3166-01, 20368-4-C3166-02, 20368-4-C3166-03, 20368-4-C3166-04, 20368-4-C3166-05, 20368-4-C3166-06, 20368-4-C3166-07, 20368-4-C3166-08, 20368-4-C3166-09, 20368-4-C3166-10, 20368-4-C3166-12, 20368-4-C3166-13, 20368-4-C3166-14, 20368-4-C3166-15, 20368-4-C3166-16, 20368-4-C3166-17, 20368-4-C3166-18, 20368-4-C3166-19, 20368-4-C3166-20, 20368-4-C3264-01, 20368-4-C3264-02, 20368-4-C3264-03, 20368-4-C3264-04, 20368-4-C3264-06, 20368-4-C3264-08, 20368-4-C3264-09, 20368-4-C3264-10, 20368-4-C3264-12, 20368-4-C3264-14, 20368-4-C3264-15, 20368-4-C3264-16, 20368-4-C3264-17, 20368-4-C3264-18, 20368-4-C3264-19, 20368-4-C3264-21, 20368-4-C3264-22, 20368-4-C3264-24, 20368-4-C3264-25, 20368-5-C1118-03, 20368-5-C1118-04, 20368-5-C1118-05, 20368-5-C3298-03, 20368-5-C3298-04, 20368-5-C3298-05, 20368-5-C3298-06, 20368-5-C3298-07, 20368-5-C3298-08, 20368-5-C3298-09, 20368-5-C3298-10, 20368-5-C3798-01, 20368-5-C3798-02, 20368-5-C3798-21, 20368-5-C3798-22.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Monteris Medical Corp
Reason for Recall:
Monteris Medical received a complaint of a separation of the distal sapphire tip from the laser delivery probe.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

SideFire Directional Laser Probe, REF: NBP101-01, NBP201-01, NBP301-01, NBP401-01, NBP501-01. These laser probes are part of the NeuroBlate System, a unique collection of MRI-compatible laser devices and accessories that create an MRI guided delivery of precision thermal therapy in the brain. The NeuroBlate System components include Gas-cooled Laser Delivery Probes to deliver controlled energy to a target zone.

Product Codes/Lot Numbers:

serial numbers: 20368-1-C1112-01, 20368-1-C1112-02, 20368-1-C1112-03, 20368-1-C1112-04, 20368-1-C1112-05, 20368-1-C1112-06, 20368-1-C1112-07, 20368-1-C1112-08, 20368-1-C1112-09, 20368-1-C3173-02, 20368-1-C3173-03, 20368-1-C3173-04, 20368-1-C3173-05, 20368-1-C3173-06, 20368-1-C3173-07, 20368-1-C3173-08, 20368-1-C3173-09, 20368-1-C3173-10, 20368-1-C3173-11, 20368-1-C3173-12, 20368-1-C3173-13, 20368-1-C3173-14, 20368-1-C3173-15, 20368-1-C3173-16, 20368-1-C3173-17, 20368-1-C3173-18, 20368-1-C3173-19, 20368-1-C3173-20, 20368-1-C3258-01, 20368-1-C3258-02, 20368-1-C3258-04, 20368-1-C3258-05, 20368-1-C3258-06, 20368-1-C3258-07, 20368-1-C3258-12, 20368-1-C3258-13, 20368-1-C3258-14, 20368-1-C3462-01, 20368-1-C3462-02, 20368-1-C3462-03, 20368-1-C3462-04, 20368-1-C3462-05, 20368-1-C3462-06, 20368-1-C3462-07, 20368-1-C3462-08, 20368-1-C3462-09, 20368-1-C3462-10, 20368-1-C3462-11, 20368-1-C3462-12, 20368-1-C3462-13, 20368-1-C3462-14, 20368-1-C3462-16, 20368-1-C3462-17, 20368-1-C3462-18, 20368-1-C3462-19, 20368-1-C3462-22, 20368-1-C3462-24, 20368-1-C3462-25, 20368-1-C3756-01, 20368-1-C3756-02, 20368-1-C3756-04, 20368-1-C3756-05, 20368-1-C3756-06, 20368-1-C3756-07, 20368-1-C3756-08, 20368-1-C3756-11, 20368-1-C3756-12, 20368-2-C1276-02, 20368-2-C1276-04, 20368-2-C1276-05, 20368-2-C1276-06, 20368-2-C1276-07, 20368-2-C1276-08, 20368-2-C1276-09, 20368-2-C1276-11, 20368-2-C1276-12, 20368-2-C1276-13, 20368-2-C1276-14, 20368-2-C1276-15, 20368-2-C1276-16, 20368-2-C1276-17, 20368-2-C1276-18, 20368-2-C1276-19, 20368-2-C1276-20, 20368-2-C1276-21, 20368-2-C1276-22, 20368-2-C3287-01, 20368-2-C3287-02, 20368-2-C3287-03, 20368-2-C3287-04, 20368-2-C3287-05, 20368-2-C3287-06, 20368-2-C3287-07, 20368-2-C3287-08, 20368-2-C3287-09, 20368-2-C3287-10, 20368-2-C3287-11, 20368-2-C3287-12, 20368-2-C3287-13, 20368-2-C3287-14, 20368-2-C3287-15, 20368-2-C3287-16, 20368-2-C3287-17, 20368-2-C3287-19, 20368-2-C3287-20, 20368-2-C3287-21, 20368-2-C3287-22, 20368-2-C3472-01, 20368-2-C3472-02, 20368-2-C3472-03, 20368-2-C3472-04, 20368-2-C3472-05, 20368-2-C3472-06, 20368-2-C3472-07, 20368-2-C3472-08, 20368-2-C3472-09, 20368-2-C3472-10, 20368-2-C3472-11, 20368-3-C1115-01, 20368-3-C1115-02, 20368-3-C1115-03, 20368-3-C1115-04, 20368-3-C1115-05, 20368-3-C1115-06, 20368-3-C1115-07, 20368-3-C1115-08, 20368-3-C1115-09, 20368-3-C1115-10, 20368-3-C1116-01, 20368-3-C1116-02, 20368-3-C1116-04, 20368-3-C1116-05, 20368-3-C1116-06, 20368-3-C3293-01, 20368-3-C3293-02, 20368-3-C3293-03, 20368-3-C3293-04, 20368-3-C3293-05, 20368-3-C3293-07, 20368-3-C3293-10, 20368-3-C3293-11, 20368-3-C3293-13, 20368-3-C3293-14, 20368-3-C3293-15, 20368-3-C3293-16, 20368-3-C3293-17, 20368-3-C3293-19, 20368-3-C3293-20, 20368-3-C3293-21, 20368-3-C3293-23, 20368-3-C3452-04, 20368-3-C3452-06, 20368-3-C3452-07, 20368-3-C3452-09, 20368-3-C3452-10, 20368-3-C3452-11, 20368-3-C3452-13, 20368-3-C3452-14, 20368-3-C3452-15, 20368-3-C3452-17, 20368-3-C3452-18, 20368-3-C3452-20, 20368-3-C3452-21, 20368-3-C3452-22, 20368-3-C3452-24, 20368-3-C3452-25, 20368-3-C3512-01, 20368-3-C3512-02, 20368-3-C3512-03, 20368-3-C3512-04, 20368-3-C3512-05, 20368-3-C3512-06, 20368-3-C3512-08, 20368-4-C1117-01, 20368-4-C1117-02, 20368-4-C1117-03, 20368-4-C1117-05, 20368-4-C1117-06, 20368-4-C1117-08, 20368-4-C1117-09, 20368-4-C1117-10, 20368-4-C3166-01, 20368-4-C3166-02, 20368-4-C3166-03, 20368-4-C3166-04, 20368-4-C3166-05, 20368-4-C3166-06, 20368-4-C3166-07, 20368-4-C3166-08, 20368-4-C3166-09, 20368-4-C3166-10, 20368-4-C3166-12, 20368-4-C3166-13, 20368-4-C3166-14, 20368-4-C3166-15, 20368-4-C3166-16, 20368-4-C3166-17, 20368-4-C3166-18, 20368-4-C3166-19, 20368-4-C3166-20, 20368-4-C3264-01, 20368-4-C3264-02, 20368-4-C3264-03, 20368-4-C3264-04, 20368-4-C3264-06, 20368-4-C3264-08, 20368-4-C3264-09, 20368-4-C3264-10, 20368-4-C3264-12, 20368-4-C3264-14, 20368-4-C3264-15, 20368-4-C3264-16, 20368-4-C3264-17, 20368-4-C3264-18, 20368-4-C3264-19, 20368-4-C3264-21, 20368-4-C3264-22, 20368-4-C3264-24, 20368-4-C3264-25, 20368-5-C1118-03, 20368-5-C1118-04, 20368-5-C1118-05, 20368-5-C3298-03, 20368-5-C3298-04, 20368-5-C3298-05, 20368-5-C3298-06, 20368-5-C3298-07, 20368-5-C3298-08, 20368-5-C3298-09, 20368-5-C3298-10, 20368-5-C3798-01, 20368-5-C3798-02, 20368-5-C3798-21, 20368-5-C3798-22.

Distribution:

Distributed in: US, CA, CT, FL, GA, KS, MN, MO, NC, OH, PA, TX, VA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0227-2015

Related Recalls

Monteris received a report that the sapphire lens on a 2.2SF probe fractured during a LITT procedure, with resulting introduction of C02 into the patient's brain.

Mar 25, 2016 Diagnostic Equipment Nationwide View Details →

Monteris Medical received a complaint of a separation of the distal sapphire tip from the laser delivery probe.

Oct 3, 2014 Infusion Pumps View Details →