Nidek MC-500 Multicolor Laser Photocoagulator using software version 2.20. Opthalmic Device. The Nidek Multicolor Laser photocoagulator Model MC-500 is indicated for use in retinal photocoagulation for treatment of ocular fundus diseases.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model MC-500 Serial numbers: 50277 50310 50330 50331 50357 50358 50383 50408 50417 50418 50425 50426 50439 50440 50445 50446 50471 50514 50311 50359 50382,
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Nidek Inc
Reason for Recall:
When the MC-500 is turned on and coagulation is performed without changing the laser color, the delivery unit occasionally fails to emit the laser, or the power output is lower than the set program.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Nidek MC-500 Multicolor Laser Photocoagulator using software version 2.20. Opthalmic Device. The Nidek Multicolor Laser photocoagulator Model MC-500 is indicated for use in retinal photocoagulation for treatment of ocular fundus diseases.

Product Codes/Lot Numbers:

Model MC-500 Serial numbers: 50277 50310 50330 50331 50357 50358 50383 50408 50417 50418 50425 50426 50439 50440 50445 50446 50471 50514 50311 50359 50382,

Distribution:

Distributed in: US, NJ, SC, IL, MI, NY, CA, TX, OH, IN, UT, MN

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0224-2015

Related Recalls

The electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2)

Feb 25, 2019 Other Medical Devices Nationwide View Details →
Class I - Dangerous

There is possibly a difference between the measured dioptric value of the subjective measurement mode and the objective measurement mode on ARK-1s.

Dec 20, 2017 Diagnostic Equipment Nationwide View Details →