CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems: 21-2101-0100-50, 21-2101-0200-02, 21-2101-0200-03, 21-2101-0200-06, 21-2101-0200-07, 21-2101-0200-12, 21-2101-0200-14, 21-2101-0200-17, 21-2101-0200-238, 21-2101-249, 21-2101-51, 21-2102-0200-02, 21-2102-249, 21-2102-51, 21-2111-0100-00, 21-2111-0100-50, 21-2111-0100-51, 21-2111-0200-02, 21-2111-0200-03, 21-2111-0200-06, 21-2111-0200-07, 21-2111-0200-08, 21-2111-0200-12, 21-2111-0200-14, 21-2111-0200-17, 21-2111-0200-232, 21-2111-0200-50, 21-2111-0300-00, 21-2111-0300-01, 21-2111-0300-02, 21-2111-0300-03, 21-2111-0300-06, 21-2111-0300-07, 21-2111-0300-08, 21-2111-0300-09, 21-2111-0300-12, 21-2111-0300-14, 21-2111-0300-17, 21-2111-0300-231, 21-2111-0300-232, 21-2111-0300-50, 21-2111-0400-01, 21-2111-0400-51, 21-2111-0401-02L, 21-2111-0401-03L, 21-2111-0401-06L, 21-2111-0401-07, 21-2111-0401-07L, 21-2111-0401-08L, 21-2111-0401-12L, 21-2111-0401-14L, 21-2111-0401-17L, 21-2111-0401-249, 21-2111-0401-51, 21-2111-0401-51L, 21-2111-0401-78, 21-2111-0402-02, 21-2111-0402-02L, 21-2111-0402-03L, 21-2111-0402-06L, 21-2111-0402-07L, 21-2111-0402-08, 21-2111-0402-08L, 21-2111-0402-09_LOAN, 21-2111-0402-09-FUK, 21-2111-0402-09JP, 21-2111-0402-09-KYO, 21-2111-0402-09-NAG, 21-2111-0402-09-OKA, 21-2111-0402-09-OSA, 21-2111-0402-09-SAP, 21-2111-0402-09-SEN, 21-2111-0402-09-TOKE, 21-2111-0402-09-TOKW, 21-2111-0402-09-YOK, 21-2111-0402-12L, 21-2111-0402-14L, 21-2111-0402-17L, 21-2111-0402-249, 21-2111-0402-51, 21-2111-0402-51L, 21-2111-0402-78, 21-2112-0100-50, 21-2112-0100-51, 21-2112-0200-02, 21-2112-0200-06, 21-2112-0300-00, 21-2112-0300-01, 21-2112-0300-02, 21-2112-0300-03, 21-2112-0300-06, 21-2112-0300-231, 21-2112-0300-50, 21-2112-0400-01, 21-2112-0400-51, 21-2112-0401-02L, 21-2112-0401-03L, 21-2112-0401-07L, 21-2112-0401-08L, 21-2112-0401-17L, 21-2112-0401-51, 21-2112-0401-51L, 21-2112-0401-78, 21-2112-0402-02, 21-2112-0402-02L, 21-2112-0402-03L, 21-2112-0402-06L, 21-2112-0402-07L, 21-2112-0402-08L, 21-2112-0402-09JP, 21-2112-0402-09-SAI, 21-2112-0402-09-SEN, 21-2112-0402-09-TOKE, 21-2112-0402-17L, 21-2112-0402-249, 21-2112-0402-51, 21-2112-0402-51L, 21-2120-0100-02, 21-2120-0100-03, 21-2120-0100-06, 21-2120-0100-232, 21-2120-0100-233, 21-2120-0100-50, 21-2120-0100-51, 21-2120-0100-95, 21-2120-0102-02, 21-2120-0102-03, 21-2120-0102-06, 21-2120-0102-07, 21-2120-0102-08, 21-2120-0102-12, 21-2120-0102-13, 21-2120-0102-14, 21-2120-0102-15, 21-2120-0102-17, 21-2120-0102-247, 21-2120-0102-249, 21-2120-0102-51, 21-2120-0102-78, 21-2120-0102-92, 21-2120-0102-98, 21-2120-0103-01, 21-2120-0103-02, 21-2120-0103-03, 21-2120-0103-06, 21-2120-0103-07, 21-2120-0103-08, 21-2120-0103-14, 21-2120-0103-15, 21-2120-0103-17, 21-2120-0103-249, 21-2120-0103-51, 21-2120-0103-78, 21-2120-0104-01, 21-2120-0105-00, 21-2120-0105-01, 21-2120-0105-02L, 21-2120-0105-03L, 21-2120-0105-06L, 21-2120-0105-07L, 21-2120-0105-08L, 21-2120-0105-12L, 21-2120-0105-13L, 21-2120-0105-14L, 21-2120-0105-15L, 21-2120-0105-17L, 21-2120-0105-231, 21-2120-0105-50, 21-2120-0105-50L, 21-2125-0104-01, 21-2125-0105-07L, 21-2125-0105-08L, 21-2125-0105-50L, 21-2127-0104-01, 21-2127-0105-01, 21-2127-0105-02L, 21-2127-0105-03L, 21-2127-0105-06L, 21-2127-0105-08L, 21-2127-0105-50L, 22-4095-0203-51, 22-4095-0204-25, 22-4095-0205-25.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All serial numbers.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Smiths Medical ASD Inc.
- Reason for Recall:
- Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pack case. If this issue occurs, the battery pack charging circuit may become inoperable.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems: 21-2101-0100-50, 21-2101-0200-02, 21-2101-0200-03, 21-2101-0200-06, 21-2101-0200-07, 21-2101-0200-12, 21-2101-0200-14, 21-2101-0200-17, 21-2101-0200-238, 21-2101-249, 21-2101-51, 21-2102-0200-02, 21-2102-249, 21-2102-51, 21-2111-0100-00, 21-2111-0100-50, 21-2111-0100-51, 21-2111-0200-02, 21-2111-0200-03, 21-2111-0200-06, 21-2111-0200-07, 21-2111-0200-08, 21-2111-0200-12, 21-2111-0200-14, 21-2111-0200-17, 21-2111-0200-232, 21-2111-0200-50, 21-2111-0300-00, 21-2111-0300-01, 21-2111-0300-02, 21-2111-0300-03, 21-2111-0300-06, 21-2111-0300-07, 21-2111-0300-08, 21-2111-0300-09, 21-2111-0300-12, 21-2111-0300-14, 21-2111-0300-17, 21-2111-0300-231, 21-2111-0300-232, 21-2111-0300-50, 21-2111-0400-01, 21-2111-0400-51, 21-2111-0401-02L, 21-2111-0401-03L, 21-2111-0401-06L, 21-2111-0401-07, 21-2111-0401-07L, 21-2111-0401-08L, 21-2111-0401-12L, 21-2111-0401-14L, 21-2111-0401-17L, 21-2111-0401-249, 21-2111-0401-51, 21-2111-0401-51L, 21-2111-0401-78, 21-2111-0402-02, 21-2111-0402-02L, 21-2111-0402-03L, 21-2111-0402-06L, 21-2111-0402-07L, 21-2111-0402-08, 21-2111-0402-08L, 21-2111-0402-09_LOAN, 21-2111-0402-09-FUK, 21-2111-0402-09JP, 21-2111-0402-09-KYO, 21-2111-0402-09-NAG, 21-2111-0402-09-OKA, 21-2111-0402-09-OSA, 21-2111-0402-09-SAP, 21-2111-0402-09-SEN, 21-2111-0402-09-TOKE, 21-2111-0402-09-TOKW, 21-2111-0402-09-YOK, 21-2111-0402-12L, 21-2111-0402-14L, 21-2111-0402-17L, 21-2111-0402-249, 21-2111-0402-51, 21-2111-0402-51L, 21-2111-0402-78, 21-2112-0100-50, 21-2112-0100-51, 21-2112-0200-02, 21-2112-0200-06, 21-2112-0300-00, 21-2112-0300-01, 21-2112-0300-02, 21-2112-0300-03, 21-2112-0300-06, 21-2112-0300-231, 21-2112-0300-50, 21-2112-0400-01, 21-2112-0400-51, 21-2112-0401-02L, 21-2112-0401-03L, 21-2112-0401-07L, 21-2112-0401-08L, 21-2112-0401-17L, 21-2112-0401-51, 21-2112-0401-51L, 21-2112-0401-78, 21-2112-0402-02, 21-2112-0402-02L, 21-2112-0402-03L, 21-2112-0402-06L, 21-2112-0402-07L, 21-2112-0402-08L, 21-2112-0402-09JP, 21-2112-0402-09-SAI, 21-2112-0402-09-SEN, 21-2112-0402-09-TOKE, 21-2112-0402-17L, 21-2112-0402-249, 21-2112-0402-51, 21-2112-0402-51L, 21-2120-0100-02, 21-2120-0100-03, 21-2120-0100-06, 21-2120-0100-232, 21-2120-0100-233, 21-2120-0100-50, 21-2120-0100-51, 21-2120-0100-95, 21-2120-0102-02, 21-2120-0102-03, 21-2120-0102-06, 21-2120-0102-07, 21-2120-0102-08, 21-2120-0102-12, 21-2120-0102-13, 21-2120-0102-14, 21-2120-0102-15, 21-2120-0102-17, 21-2120-0102-247, 21-2120-0102-249, 21-2120-0102-51, 21-2120-0102-78, 21-2120-0102-92, 21-2120-0102-98, 21-2120-0103-01, 21-2120-0103-02, 21-2120-0103-03, 21-2120-0103-06, 21-2120-0103-07, 21-2120-0103-08, 21-2120-0103-14, 21-2120-0103-15, 21-2120-0103-17, 21-2120-0103-249, 21-2120-0103-51, 21-2120-0103-78, 21-2120-0104-01, 21-2120-0105-00, 21-2120-0105-01, 21-2120-0105-02L, 21-2120-0105-03L, 21-2120-0105-06L, 21-2120-0105-07L, 21-2120-0105-08L, 21-2120-0105-12L, 21-2120-0105-13L, 21-2120-0105-14L, 21-2120-0105-15L, 21-2120-0105-17L, 21-2120-0105-231, 21-2120-0105-50, 21-2120-0105-50L, 21-2125-0104-01, 21-2125-0105-07L, 21-2125-0105-08L, 21-2125-0105-50L, 21-2127-0104-01, 21-2127-0105-01, 21-2127-0105-02L, 21-2127-0105-03L, 21-2127-0105-06L, 21-2127-0105-08L, 21-2127-0105-50L, 22-4095-0203-51, 22-4095-0204-25, 22-4095-0205-25.
Product Codes/Lot Numbers:
All serial numbers.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0222-2025
Related Recalls
Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.