X-Guide X-Mark Probe Tool -used for registering edentulous patients to the software or performing XMark registration. Catalog Number:P010673

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: (01)00817421020995 (10) 2007010010 Lot number: 2007010010
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
X-NAV Technologies, LLC
Reason for Recall:
Product missing one of two welds connecting the tip to the shaft, if the remaining weld breaks, it is possible that the tip could fall out during X-Mark registration and fall into the patient s mouth. In this case, the patient may aspirate or swallow the probe tip.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

X-Guide X-Mark Probe Tool -used for registering edentulous patients to the software or performing XMark registration. Catalog Number:P010673

Product Codes/Lot Numbers:

UDI-DI: (01)00817421020995 (10) 2007010010 Lot number: 2007010010

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0222-2023

Related Recalls

The X-Guide system utilizes an X-Clip (PN P007665) which contains fiducials to assist with registration of the patient anatomy. The X-Clip is a device worn intraorally by a patient during the CT scan and guided surgery procedures. The X-Clip contains three fiducial markers made from a Stainless Steel material. The X-Nav specification requires these fiducials be fabricated from Stainless Steel 3

Feb 23, 2016 Implants & Prosthetics View Details →