EnVision FLEX/HRP visualization reagent found in the following kits: EnVision FLEX, High pH (Link) EnVision FLEX+, Mouse, High pH (Link) EnVision FLEX Mini Kit, High pH (Link) EnVision FLEX, High pH (Dako Omnis) Multi-component, sealed package. For in vitro diagnostic use.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Item No.: K800021-2, K800221-2, K802321-2 (Link) K801021-2, K801221-2, K802421-2 (AS/AS+) GV80011-2, GV82311-2 (Dako Omnis) Catalog/Model No.: K8000 K8002 K8023 GV800 SM802 EnVision FLEX / HRP Lot 20019374 used in: Envision FLEX, High pH (link) Lot No.: 20019097, 20020777, 20021381 Envision FLEX+, Mouse, High pH (link) Lot No.: 20019103, 20019113, 20020772 EnVision FLEX Mini Kit, High pH (Link) Lot No.: 20021362 DM 842 EnVision FLEX / HRP (Dako Omnis) Lot 20019013 used in: EnVision FLEX, High pH (Dako Omnis) Lot No.: 20019102
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Dako North America Inc.
Reason for Recall:
There is a defect in one lot of a buffer, non-specific background staining that has, in some instances, interfered with interpretation of the staining results.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

EnVision FLEX/HRP visualization reagent found in the following kits: EnVision FLEX, High pH (Link) EnVision FLEX+, Mouse, High pH (Link) EnVision FLEX Mini Kit, High pH (Link) EnVision FLEX, High pH (Dako Omnis) Multi-component, sealed package. For in vitro diagnostic use.

Product Codes/Lot Numbers:

Item No.: K800021-2, K800221-2, K802321-2 (Link) K801021-2, K801221-2, K802421-2 (AS/AS+) GV80011-2, GV82311-2 (Dako Omnis) Catalog/Model No.: K8000 K8002 K8023 GV800 SM802 EnVision FLEX / HRP Lot 20019374 used in: Envision FLEX, High pH (link) Lot No.: 20019097, 20020777, 20021381 Envision FLEX+, Mouse, High pH (link) Lot No.: 20019103, 20019113, 20020772 EnVision FLEX Mini Kit, High pH (Link) Lot No.: 20021362 DM 842 EnVision FLEX / HRP (Dako Omnis) Lot 20019013 used in: EnVision FLEX, High pH (Dako Omnis) Lot No.: 20019102

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0220-2016

Related Recalls

An incorrect HER2 probe concentration was used when producing the probe component (Vial 3) in this kit lot. This has resulted in the HER2 probe concentration being too low.

Jun 10, 2016 Diagnostic Equipment Nationwide View Details →

If a user requests slides from the LIS or TPID, then updates a request by changing the test, the TRD software will reject the update. The LIS and printed slide label will display the updated test with the requested change, but the Omnis system will execute the initial test, not reflecting the change. The system will not warn the user that the test request was rejected.

Apr 1, 2015 Diagnostic Equipment Nationwide View Details →