e.cam Gamma Camera, an emission computed tomography system. It is intended to detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    E.cam Model numbers affected : 4380213, 4380221, 5242826, 5077066, 5077074, 5984005, 5989079, 5989087, 5989095, 5991109, 5991117, 5992099, 7324119, 7324135, 7324143, 7324150, 7760809, 7760932, 7760161, 7761161, 7823920, 7823938, 7823946, 7823953, 7823961, 7823979, 10151531, 10151532, 10520745. This recall applies only to ecam systems that were installed before December 31st 2004. Serial Numbers 7438, 7440, 81010, 1011, 8509, 8510, 1, 1007, 1008, 1010, 1019, 1020, 2001, 2004, 2006, 2007, 2009, 2010, 2014, 2017, 2019, 2022, 2023, 2025, 3002, 3006, 3008, 3010, 3011, 3012, 3013, 3014, 3017, 3021, 3022, 3023, 3024, 3029, 3031, 3034, 4001, 4002, 4004, 4008, 4009, 4010, 4011, 4013, 4014, 4016, 1006, 1009, 1015, 1016, 1020, 1021, 1024, 1025, 1029, 1030, 1031, 1032, 1033, 1041, 1044, 1046, 1051, 1056, 1057, 1063, 1068, 1069, 1070, 1071, 10711, 1074, 1076, 1077, 1079, 1084, 1092, 1095, 1096, 1098, 1100, 1102, 1105, 1106, 1111, 1116, 1117, 1118, 1122, 1123, 1124, 1127, 1130, 1132, 1133, 1134, 1135, 1136, 1137, 1142, 1143, 1146, 1148, 1150, 1151, 1153, 1155, 1157, 1158, 1160, 1161, 1162, 1163, 1165, 1166, 1169, 1172, 1174, 1176, 1180, 1181, 1182, 1184, 1186, 1187, 1188, 1189, 1190, 1191, 1192, 1196, 1197, 1199, 1200, 12009, 1201, 1202, 1205, 1206, 1207, 1208, 121010, 1211, 1234, 126110, 21909, 26011, 31609, 31711, 31827, 41207, 62111, 62310, 7030, 71410, 71411, 7440, 7502, 7582, 8149, 82409, 9, 918073, 1002, 1006, 1009, 1013, 1014, 1015, 1016, 1019, 1020, 1021, 1024, 1029, 1034, 1036, 1037, 1038, 1039, 1046, 1049, 1052, 1056, 1058, 1059, 1062, 1063, 1065, 1067, 1071, 1077, 1080, 1081, 1082, 1083, 1088, 1089, 1094, 1095, 1099, 1101, 1103, 1104, 1105, 1107, 1109, 1110, 1113, 1114, 1116, 1118, 1121, 1129, 1130, 1131, 1135, 1136, 1137, 1138, 1139, 1141, 1142, 1143, 2001, 2004, 2006, 2007, 2008, 2010, 2012, 2013, 2014, 2015, 2017, 2019, 2024, 2026, 2027, 2028, 2030, 2036, 2038, 2039, 2041, 2042, 2045, 2047, 2048, 2050, 2051, 2052, 2055, 2058, 2060, 2065, 2066, 2074, 2075, 2076, 2078, 2079, 2080, 2081, 2084, 2090, 2096, 2098, 2099, 2100, 2107, 2114, 2115, 2117, 2118, 2119, 2120, 2122, 2123, 2124, 2125, 2127, 2128, 2130, 3004, 3005, 3007, 3008, 3010, 3015, 3016, 3018, 3020, 3022, 3023, 3026, 3027, 3029, 3032, 3035, 3037, 3039, 3040, 3045, 3047, 3048, 3049, 3050, 3051, 3052, 3055, 3056, 3057, 3058, 3064, 3065, 3067, 3069, 3070, 3071, 3075, 3076, 3077, 3080, 3082, 3083, 3085, 3087, 3088, 3090, 3091, 3092, 3093, 3094, 3096, 3098, 3100, 3101, 3104, 3105, 3106, 3107, 3109, 3112, 3114, 3115, 3116, 3117, 3118, 3120, 3121, 3122, 3123, 3124, 3125, 3132, 3133, 3135, 3137, 3138, 3139, 3140, 3141, 4001, 4002, 4004, 4006, 4007, 4008, 4010, 4011, 4012, 4014, 4015, 4017, 4018, 4019, 4021, 4022, 4023, 4024, 4025, 4026, 4028, 4030, 4034, 4039, 4041, 4044, 4045, 4048, 4049, 4050, 4052, 4053, 4054, 4056, 4057, 4060, 4062, 4063, 4064, 4066, 4067, 4068, 4069, 4070, 4071, 4076, 4077, 4078, 4090, 10209, 10411, 12210, 12310, 31611, 40109, 4082, 4083, 4084, 4086, 4095, 4096, 4097, 4098, 4100, 4107, 4112, 4114, 4116, 4117, 4118, 4119, 4120, 4121, 4122, 4125, 4127, 4129, 4130, 4131, 4132, 4135, 4136, 4137, 4141, 4142, 4143, 4146, 4147, 4148, 4149, 4151, 4152, 4154, 4156, 4158, 4159, 4164, 4166, 4168, 4171, 4173, 4174, 4177, 4178, 4179, 4181, 4182, 4183, 4185, 4186, 4190, 4196, 4197, 4202, 4205, 4206, 4208, 4209, 4214, 4215, 4217, 4218, 4219, 4220, 4221, 4222, 4223, 4225, 4226, 5001, 5002, 5004, 5005, 5006, 5008, 5009, 5011, 5012, 5013, 5016, 5017, 5018, 5019, 5020, 5021, 5022, 5023, 5024, 5025, 5027, 5029, 5031, 5032, 5033, 5036, 5037, 5039, 5040, 5041, 5042, 5043, 5045, 5046, 5048, 5049, 5050, 5051, 5057, 5060, 5063, 5067, 5069, 5070, 5071, 5102, 5107, 5108, 5117, 5123, 6000, 60911, 62211, 83110, 83210, 92110, 4017, 4018, 4019, 4022, 4023, 4025, 4030, 4032, 4033, 4034, 4035, 4037, 4044, 4046, 4049, 5001, 5003, 5004, 5006, 5007, 5008, 5010, 5011, 5012, 5013, 5015, 5016, 5017, 5018, 5020, 5021, 5022, 5024, 5025, 5028, 5031, 5033, 5036, 5037, 5039, 5040, 5043, 3, 6006, 6007, 6009, 6010, 7001, 7004, 7006, 7409, 7501, 7502, 7507, 7509, 6053, 6501, 7386, 7502, 7504, 7505, 7507, 7508, 7510, 7512, 7513, 7515, 7516, 7517, 7518, 7519, 7520, 7521, 7522, 7523, 7525, 7526, 7529, 7531, 7533, 7534, 7535, 7536, 7538, 7539, 7540, 7542, 7545, 7547, 7548, 7549, 7550, 7551, 7553, 7555, 7556, 7557, 7558, 7561, 7562, 7563, 7564, 7566, 7568, 7569, 7573, 7574, 7576, 7577, 7578, 7579, 7581, 7582, 7584, 7587, 7588, 7589, 7590, 7592, 7595, 7596, 7599, 7600, 7602, 7605, 7606, 7608, 7610, 7611, 7614, 7615, 7616, 7617, 7618, 7618*, 7620, 7621, 7622, 7623, 7624, 7626, 7627, 7628, 7629, 7631, 7635, 7636, 7637, 7638, 7651, 7861, 8001, 8003, 8005, 8006, 8007, 8009, 8010, 8011, 8014, 8015, 8017, 8020, 8021, 8023, 8024, 8025, 8027, 8028, 8029, 8030, 8031, 8032, 8035, 8036, 8037, 8039, 8040, 8044, 8045, 8047, 8049, 8050, 8051, 8053, 8055, 8056, 8057, 8058, 8059, 8060, 8063, 8064, 8067, 8069, 8070, 8071, 8072, 8073, 8075, 8078, 8079, 8080, 8081, 8082, 8085, 8087, 8088, 8089, 8090, 8092, 8094, 8095, 8096, 8097, 8098, 8099, 8100, 8102, 8105, 8106, 8108, 8109, 8110, 8111, 8112, 8113, 8114, 8115, 8116, 8117, 8118, 8119, 8120, 8121, 8122, 8124, 8125, 8126, 8127, 8128, 8130, 8131, 8132, 8133, 8135, 8137, 8138, 8141, 8142, 8144, 8145, 8146, 8147, 8501, 8502, 8503, 8504, 8505, 8506, 8508, 8510, 8511, 8512, 8513, 8514, 8515, 8516, 8517, 8518, 8519, 8520, 8521, 8522, 8523, 8524, 8525, 8526, 8527, 8528, 8529, 8530, 8532, 8533, 8534, 8535, 8536, 8538, 8540, 8541, 8543, 8544, 8545, 8548, 8549, 8550, 8552, 8553, 8556, 8558, 8559, 8560, 8561, 8562, 8563, 8564, 8565, 8567, 8568, 8569, 8570, 8571, 8572, 8573, 8574, 8575, 8576, 8577, 8578, 8580, 8581, 8583, 8585, 8586, 8587, 8588, 8589, 8590, 8592, 8593, 8595, 8596, 8598, 8600, 8603, 8604, 8605, 8606, 8607, 8608, 8609, 8610, 8611, 8612, 8613, 8615, 8616, 8617, 8618, 8619, 8620, 8621, 8622, 8624, 8625, 8626, 8628, 8629, 8630, 8631, 8632, 8633, 8634, 8635, 8636, 8637, 8638, 8639, 8640, 8641, 8642, 8643, 8645, 8646, 8647, 8650, 8651, 8652, 8653, 8654, 8655, 8656, 8657, 8658, 8659, 8660, 8661, 8662, 8665, 8666, 8667, 8668, 8669, 8670, 8671, 8672, 8673, 8674, 8675, 8676, 8677, 8678, 8679, 8681, 8683, 8684, 8685, 8686, 8687, 8688, 8689, 8690, 8691, 8692, 8693, 8694, 8695, 8696, 8697, 8698, 8699, 8700, 8701, 8702, 8703, 8704, 8705, 8706, 8707, 8708, 8709, 8710, 8711, 8712, 8714, 8716, 8717, 8718, 8719, 8720, 8721, 8722, 8723, 8724, 8725, 8726, 8727, 8728, 8729, 8730, 8731, 8732, 8733, 8734, 8735, 8736, 8738, 8739, 8740, 8741, 8743, 8744, 8745, 8746, 8747, 8748, 8749, 8750, 8754, 8755, 8756, 8757, 8758, 8759, 8760, 8761, 8763, 8766, 8767, 8769, 8770, 8771, 8773, 8778, 8779, 8780, 8781, 8782, 8783, 8784, 8786, 8788, 8789, 8790, 8791, 8792, 8793, 8794, 8799, 8801, 8802, 8803, 8804, 8805, 8807, 8808, 8809, 8810, 8811, 8812, 8813, 8814, 8816, 8817, 8818, 8819, 8820, 8822, 8823, 8824, 8825, 8827, 8828, 8829, 8830, 8831, 8832, 8833, 8834, 8835, 8836, 8837, 8838, 8839, 8841, 8842, 8844, 5045, 7502, 7503, 7505, 7506, 7509, 7511, 8001, 8003, 8004, 8007, 8009, 8010, 8012, 8013, 8014, 8015, 8016, 8017, 8019, 8021, 8022, 8024, 8025, 8027, 8028, 8030, 8032, 8501, 8503, 8504, 8506, 8507, 8508, 8509, 8511, 8512, 8513, 8514, 8517, 8518, 8519, 8521, 8523, 8524, 8525, 8526, 8528, 8530, 8532, 8534, 8535, 8536, 8537, 8539, 8540, 8543, 8545, 8547, 8548, 8549, 8550, 8552, 8553, 8554, 8555, 8556, 7090, 7502, 7504, 7505, 8001, 8003, 8005, 8021, 8501, 8502, 8503, 8504, 8505, 8506, 1001, 13044, 5039, 5505, 5507, 5508, 5510, 5511, 5512, 5514, 5517, 5518, 5519, 5521, 5522, 5523, 5524, 5525, 5526, 5527, 5528, 5529, 5530, 5531, 5532, 5533, 5535, 5536, 5537, 5538, 5539, 5540, 5541, 5542, 5546, 5547, 5548, 5550, 5551, 5553, 5555, 5556, 5557, 5558, 5559, 5560, 5561, 5562, 5563, 5565, 5567, 5570, 5571, 5572, 5573, 5575, 5578, 5579, 5580, 5581, 5584, 5585, 5586, 5587, 5588, 5589, 5590, 5591, 5593, 5594, 5595, 5597, 5599, 5600, 5601, 5602, 5603, 5604, 5605, 5606, 5607, 5608, 5609, 5610, 5611, 5612, 5613, 5614, 5616, 5617, 5618, 5620, 5622, 5623, 5624, 5625, 5629, 5631, 5632, 5634, 5636, 5637, 5639, 5640, 5642, 5644, 5645, 5646, 5647, 5649, 5650, 5658, 5671, 5672, 5673, 5674, 5675, 5676, 5677, 5678, 5679, 7005, 7006, 7007, 7009, 7011, 7012, 7013, 7014, 7016, 7017, 7018, 7019, 7020, 7022, 7025, 7027, 7028, 7029, 7030, 7031, 7032, 7033, 7034, 7036, 7037, 7039, 7040, 7041, 7043, 7044, 7045, 7046, 7047, 7048, 7049, 7050, 7051, 7056, 7058, 7059, 7060, 7061, 7062, 7063, 7065, 7067, 7068, 7069, 7070, 7073, 7076, 7078, 7080, 7082, 7083, 7084, 7086, 7088, 7090, 7091, 7092, 7094, 7095, 7096, 7097, 7098, 7102, 7103, 7104, 7105, 7109, 7110, 7111, 7112, 7117, 7120, 7121, 7124, 7126, 7127, 7501, 7502, 7503, 7504, 7505, 7506, 7508, 7509, 7510, 7512, 7513, 7515, 7516, 7517, 7519, 7520, 7521, 7523, 7524, 7525, 7526, 7530, 7531, 7532, 7533, 7534, 7535, 7536, 7537, 7538, 7539, 7541, 7544, 7545, 7551, 7552, 7553, 7555, 7556, 7557, 7560, 7561, 7562, CAX271/U2B004, 5501, 5502, 5503, 5506, 5511, 5512, 5513, 5514, 5516, 5517, 5519, 5521, 5523, 5525, 5526, 5527, 5529, 7003, 7004, 7005, 7006, 7008, 7011, 7013, 7015, 7017, 7018, 7019, 7020, 7021, 7022, 7024, 7025, 7026, 7501, 7504, 7506, 7507, 7508, 7509, 7513, 5101, 5102, 5104, 5109, 5110, 5112, 5113, 5115, 5116, 5118, 5121, 5125, 5126, 5127, 5128, 5131, 5132, 5133, 5135, 5136, 5142, 5144, 5146, 5147, 5148, 5149, 5150, 5151, 5153, 5157, 5158, 5159, 5160, 5162, 5165, 5167, 5168, 5170, 5171, 5172, 5173, 5174, 5502, 1003, 548, 550, 558, 564, 579, 581, 588, 589, 590, 592, 593, 595, RC528, RC537, RC549, 551, 584, RC521, RC536, 10610, 21511, 510, 529, 530, 557, 559, 573, 62310, RC498, RC543, 556, 572, RC493, RC524, RC538, RC541, RC545, RC546, 8501, 8503, 8504, 8505, 8506, 8508, 8509, 8510, 8511, 100003, 1234, 1234578, 4916, 60151, 63010, 7409, 7502, 8191, 8501, 8502, 8503, 8506, 8507, 8508, 8509, 8512, 8514, 8515, 8517, 8519, 8520, 8521, 8522, 8523, 8524, 8525, 8529, 8530, 8531, 8532, 8533, 8534, 8537, 8538, 8539, 8540, 8541, 8542, 8543, 8544, 8546, 8547, 8548, 8549, 8550, 8551, 8552, 8553, 8554, 8555, 8556, 8557, 8560, 8561, 8562, 8563, 8564, 8565, 8566, 8567, 8568, 8569, 8570, 8571, 8572, 8573, 8574, 8577, 8578, 8579, 8580, 8581, 8583, 9003, 9004, 9006, 9007, 9008, 9009, 9010, 9011, 9012, 9013, 9014, 9015, 9016, 9017, 9018, 9019, 9022, 9023, 9024, 9026, 9027, 9028, 9029, 9030, 9031, 9032, 9033, 9034, 9035, 9036, 9037, 9038, 9039, 9041, 9043, 8473046345, 8503, 8205, 8505, 8506, 8509, 8510, 8511, 8513, 8514, 8515, 8516, 8517, 8518, 8519, 8520, 9821, 8501, 8501*, 8502, 8507, 8508, 10000, 10001, 10002, 10003, 10005, 10006, 10007, 10008, 10009, 10010, 10013, 10014, 10015, 10016, 10017, 10019, 10020, 10022, 10023, 10024, 10025, 10026, 10028, 10029, 10030, 10031, 10032, 10033, 10034, 10035, 10036, 10037, 10038, 10040, 10041, 10042, 10044, 10046, 10047, 10048, 10049, 10050, 10051, 10052, 10053, 10054, 10055, 10056, 10057, 10058, 10059, 10060, 10061, 10062, 10063, 10065, 10066, 10067, 10068, 10069, 10070, 10071, 10072, 10073, 10074, 10076, 10077, 10078, 10079, 10080, 10082, 10083, 10084, 10085, 10087, 10088, 10089, 10090, 10091, 10092, 10093, 10094, 10096, 10098, 10099, 10100, 10101, 10102, 10103, 10104, 10106, 10107, 10108, 10109, 10111, 10112, 10113, 10114, 10115, 10117, 10118, 10119, 10120, 10121, 10122, 10124, 10125, 10126, 10127, 10128, 10129, 10130, 10131, 10132, 10133, 10135, 10136, 10137, 10138, 10139, 10140, 10141, 10142, 10143, 10144, 10147, 10149, 10150, 10151, 10153, 10154, 10155, 10156, 10157, 10158, 10159, 10160, 10161, 10162, 10163, 10164, 10165, 10166, 10167, 10168, 10169, 10170, 10171, 10172, 10173, 10174, 10175, 10176, 10177, 10178, 10179, 10180, 10181, 10182, 10183, 10184, 10185, 10186, 10187, 10188, 10191, 10192, 10193, 10194, 10195, 10198, 10199, 10200, 10201, 10202, 10203, 10204, 10205, 10206, 10207, 10208, 10209, 10211, 10212, 10213, 10214, 10215, 10216, 10217, 10218, 10219, 10220, 10221, 10222, 10224, 10225, 10226, 10227, 10228, 10230, 10231, 10232, 10233, 10234, 10235, 10236, 10237, 10238, 10239, 10240, 10241, 10243, 10245, 10246, 10248, 10249, 10250, 10251, 10252, 10253, 10254, 10255, 10256, 10257, 10258, 10259, 10260, 10261, 10262, 10263, 10264, 10265, 10266, 10267, 10268, 10269, 10270, 10271, 10272, 10273, 10274, 10275, 10276, 10277, 10278, 10280, 10282, 10283, 10284, 10285, 10286, 10287, 10288, 10289, 10290, 10291, 10292, 10293, 10294, 10295, 10296, 10297, 10298, 10299, 10300, 10301, 10302, 10303, 10304, 10305, 10306, 10307, 10308, 10309, 10310, 10311, 10312, 10313, 10314, 10315, 10317, 10318, 10319, 10320, 10321, 10322, 10323, 10324, 10325, 10326, 10327, 10328, 10329, 10330, 10331, 10332, 10333, 10334, 10336, 10337, 10338, 10339, 10340, 10341, 10343, 10344, 5382, 5517, 5764, 8490, 9001, 9002, 9003, 9004, 9005, 9006, 9008, 9009, 9010, 9012, 9013, 9014, 9015, 9016, 9017, 9018, 9019, 9022, 9023, 9024, 9025, 9026, 9027, 9028, 9031, 9033, 9034, 9035, 9036, 9038, 9039, 9040, 9041, 9043, 9044, 9045, 9046, 9047, 9049, 9050, 9051, 9053, 9054, 9055, 9057, 9058, 9059, 9060, 9061, 9062, 9063, 9064, 9066, 9068, 9069, 9070, 9073, 9074, 9076, 9077, 9078, 9079, 9080, 9081, 9082, 9083, 9084, 9085, 9086, 9087, 9089, 9090, 9091, 9092, 9094, 9095, 9096, 9097, 9098, 9099, 9100, 9101, 9102, 9103, 9104, 9106, 9107, 9108, 9109, 9110, 9111, 9112, 9113, 9114, 9115, 9116, 9117, 9120, 9122, 9123, 9124, 9125, 9126, 9127, 9129, 9130, 9132, 9133, 9134, 9135, 9136, 9137, 9138, 9139, 9142, 9143, 9144, 9145, 9146, 9147, 9148, 9149, 9150, 9151, 9152, 9154, 9155, 9156, 9157, 9158, 9159, 9161, 9162, 9163, 9164, 9165, 9166, 9167, 9169, 9170, 9171, 9172, 9173, 9174, 9175, 9176, 9177, 9179, 9180, 9181, 9182, 9183, 9184, 9185, 9186, 9187, 9189, 9190, 9191, 9193, 9194, 9198, 9199, 9200, 9201, 9204, 9206, 9207, 9210, 9212, 9213, 9214, 9215, 9216, 9219, 9220, 9221, 9223, 9224, 9225, 9226, 9227, 9228, 9230, 9231, 9232, 9233, 9234, 9235, 9238, 9239, 9240, 9241, 9242, 9243, 9245, 9246, 9247, 9248, 9251, 9252, 9253, 9254, 9255, 9256, 9259, 9260, 9261, 9262, 9263, 9264, 9265, 9266, 9267, 9268, 9269, 9270, 9271, 9272, 9273, 9274, 9275, 9276, 9277, 9279, 9280, 9281, 9282, 9283, 9284, 9285, 9287, 9288, 9289, 9290, 9292, 9293, 9294, 9295, 9296, 9297, 9298, 9299, 9300, 9301, 9302, 9303, 9304, 9305, 9306, 9307, 9308, 9309, 9311, 9313, 10001, 10002, 10003, 10004, 10005, 10006, 10007, 10008, 10009, 10011, 10012, 10014, 10015, 10016, 10017, 10018, 10019, 10022, 10023, 10025, 10026, 10027, 10028, 10029, 10030, 10031, 10032, 10034, 10035, 10036, 10038, 10039, 10040, 10041, 10044, 10045, 10046, 10047, 10048, 10049, 10050, 10051, 10052, 10055, 10056, 10057, 10058, 10059, 10061, 8505, 8509, 9001, 9002, 9003, 9004, 9005, 9007, 9008, 9009, 9010, 9011, 9012, 9013, 9014, 9015, 9016, 9017, 9019, 9020, 9021, 9022, 9024, 9026, 9031, 9032, 9033, 9034, 9036, 9037, 9038, 9039, 9040, 9041, 9042, 9043, 9044, 9045, 9001, 9002, 9005, 9006, 9007, 9009, 9010, 9011, 9013, 9014, 9016, and 9017.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Medical Solutions USA, Inc.
Reason for Recall:
Potential failure mode related to the radial brake, where the detector was able to drift, led to a re-design of the brake redundancy circuit. The redesigned Brake Redundancy Circuit will force even a defective brake to engage in the unpowered state by applying a reverse charge to the brake and ensuring the brake has demagnetized.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

e.cam Gamma Camera, an emission computed tomography system. It is intended to detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV.

Product Codes/Lot Numbers:

E.cam Model numbers affected : 4380213, 4380221, 5242826, 5077066, 5077074, 5984005, 5989079, 5989087, 5989095, 5991109, 5991117, 5992099, 7324119, 7324135, 7324143, 7324150, 7760809, 7760932, 7760161, 7761161, 7823920, 7823938, 7823946, 7823953, 7823961, 7823979, 10151531, 10151532, 10520745. This recall applies only to ecam systems that were installed before December 31st 2004. Serial Numbers 7438, 7440, 81010, 1011, 8509, 8510, 1, 1007, 1008, 1010, 1019, 1020, 2001, 2004, 2006, 2007, 2009, 2010, 2014, 2017, 2019, 2022, 2023, 2025, 3002, 3006, 3008, 3010, 3011, 3012, 3013, 3014, 3017, 3021, 3022, 3023, 3024, 3029, 3031, 3034, 4001, 4002, 4004, 4008, 4009, 4010, 4011, 4013, 4014, 4016, 1006, 1009, 1015, 1016, 1020, 1021, 1024, 1025, 1029, 1030, 1031, 1032, 1033, 1041, 1044, 1046, 1051, 1056, 1057, 1063, 1068, 1069, 1070, 1071, 10711, 1074, 1076, 1077, 1079, 1084, 1092, 1095, 1096, 1098, 1100, 1102, 1105, 1106, 1111, 1116, 1117, 1118, 1122, 1123, 1124, 1127, 1130, 1132, 1133, 1134, 1135, 1136, 1137, 1142, 1143, 1146, 1148, 1150, 1151, 1153, 1155, 1157, 1158, 1160, 1161, 1162, 1163, 1165, 1166, 1169, 1172, 1174, 1176, 1180, 1181, 1182, 1184, 1186, 1187, 1188, 1189, 1190, 1191, 1192, 1196, 1197, 1199, 1200, 12009, 1201, 1202, 1205, 1206, 1207, 1208, 121010, 1211, 1234, 126110, 21909, 26011, 31609, 31711, 31827, 41207, 62111, 62310, 7030, 71410, 71411, 7440, 7502, 7582, 8149, 82409, 9, 918073, 1002, 1006, 1009, 1013, 1014, 1015, 1016, 1019, 1020, 1021, 1024, 1029, 1034, 1036, 1037, 1038, 1039, 1046, 1049, 1052, 1056, 1058, 1059, 1062, 1063, 1065, 1067, 1071, 1077, 1080, 1081, 1082, 1083, 1088, 1089, 1094, 1095, 1099, 1101, 1103, 1104, 1105, 1107, 1109, 1110, 1113, 1114, 1116, 1118, 1121, 1129, 1130, 1131, 1135, 1136, 1137, 1138, 1139, 1141, 1142, 1143, 2001, 2004, 2006, 2007, 2008, 2010, 2012, 2013, 2014, 2015, 2017, 2019, 2024, 2026, 2027, 2028, 2030, 2036, 2038, 2039, 2041, 2042, 2045, 2047, 2048, 2050, 2051, 2052, 2055, 2058, 2060, 2065, 2066, 2074, 2075, 2076, 2078, 2079, 2080, 2081, 2084, 2090, 2096, 2098, 2099, 2100, 2107, 2114, 2115, 2117, 2118, 2119, 2120, 2122, 2123, 2124, 2125, 2127, 2128, 2130, 3004, 3005, 3007, 3008, 3010, 3015, 3016, 3018, 3020, 3022, 3023, 3026, 3027, 3029, 3032, 3035, 3037, 3039, 3040, 3045, 3047, 3048, 3049, 3050, 3051, 3052, 3055, 3056, 3057, 3058, 3064, 3065, 3067, 3069, 3070, 3071, 3075, 3076, 3077, 3080, 3082, 3083, 3085, 3087, 3088, 3090, 3091, 3092, 3093, 3094, 3096, 3098, 3100, 3101, 3104, 3105, 3106, 3107, 3109, 3112, 3114, 3115, 3116, 3117, 3118, 3120, 3121, 3122, 3123, 3124, 3125, 3132, 3133, 3135, 3137, 3138, 3139, 3140, 3141, 4001, 4002, 4004, 4006, 4007, 4008, 4010, 4011, 4012, 4014, 4015, 4017, 4018, 4019, 4021, 4022, 4023, 4024, 4025, 4026, 4028, 4030, 4034, 4039, 4041, 4044, 4045, 4048, 4049, 4050, 4052, 4053, 4054, 4056, 4057, 4060, 4062, 4063, 4064, 4066, 4067, 4068, 4069, 4070, 4071, 4076, 4077, 4078, 4090, 10209, 10411, 12210, 12310, 31611, 40109, 4082, 4083, 4084, 4086, 4095, 4096, 4097, 4098, 4100, 4107, 4112, 4114, 4116, 4117, 4118, 4119, 4120, 4121, 4122, 4125, 4127, 4129, 4130, 4131, 4132, 4135, 4136, 4137, 4141, 4142, 4143, 4146, 4147, 4148, 4149, 4151, 4152, 4154, 4156, 4158, 4159, 4164, 4166, 4168, 4171, 4173, 4174, 4177, 4178, 4179, 4181, 4182, 4183, 4185, 4186, 4190, 4196, 4197, 4202, 4205, 4206, 4208, 4209, 4214, 4215, 4217, 4218, 4219, 4220, 4221, 4222, 4223, 4225, 4226, 5001, 5002, 5004, 5005, 5006, 5008, 5009, 5011, 5012, 5013, 5016, 5017, 5018, 5019, 5020, 5021, 5022, 5023, 5024, 5025, 5027, 5029, 5031, 5032, 5033, 5036, 5037, 5039, 5040, 5041, 5042, 5043, 5045, 5046, 5048, 5049, 5050, 5051, 5057, 5060, 5063, 5067, 5069, 5070, 5071, 5102, 5107, 5108, 5117, 5123, 6000, 60911, 62211, 83110, 83210, 92110, 4017, 4018, 4019, 4022, 4023, 4025, 4030, 4032, 4033, 4034, 4035, 4037, 4044, 4046, 4049, 5001, 5003, 5004, 5006, 5007, 5008, 5010, 5011, 5012, 5013, 5015, 5016, 5017, 5018, 5020, 5021, 5022, 5024, 5025, 5028, 5031, 5033, 5036, 5037, 5039, 5040, 5043, 3, 6006, 6007, 6009, 6010, 7001, 7004, 7006, 7409, 7501, 7502, 7507, 7509, 6053, 6501, 7386, 7502, 7504, 7505, 7507, 7508, 7510, 7512, 7513, 7515, 7516, 7517, 7518, 7519, 7520, 7521, 7522, 7523, 7525, 7526, 7529, 7531, 7533, 7534, 7535, 7536, 7538, 7539, 7540, 7542, 7545, 7547, 7548, 7549, 7550, 7551, 7553, 7555, 7556, 7557, 7558, 7561, 7562, 7563, 7564, 7566, 7568, 7569, 7573, 7574, 7576, 7577, 7578, 7579, 7581, 7582, 7584, 7587, 7588, 7589, 7590, 7592, 7595, 7596, 7599, 7600, 7602, 7605, 7606, 7608, 7610, 7611, 7614, 7615, 7616, 7617, 7618, 7618*, 7620, 7621, 7622, 7623, 7624, 7626, 7627, 7628, 7629, 7631, 7635, 7636, 7637, 7638, 7651, 7861, 8001, 8003, 8005, 8006, 8007, 8009, 8010, 8011, 8014, 8015, 8017, 8020, 8021, 8023, 8024, 8025, 8027, 8028, 8029, 8030, 8031, 8032, 8035, 8036, 8037, 8039, 8040, 8044, 8045, 8047, 8049, 8050, 8051, 8053, 8055, 8056, 8057, 8058, 8059, 8060, 8063, 8064, 8067, 8069, 8070, 8071, 8072, 8073, 8075, 8078, 8079, 8080, 8081, 8082, 8085, 8087, 8088, 8089, 8090, 8092, 8094, 8095, 8096, 8097, 8098, 8099, 8100, 8102, 8105, 8106, 8108, 8109, 8110, 8111, 8112, 8113, 8114, 8115, 8116, 8117, 8118, 8119, 8120, 8121, 8122, 8124, 8125, 8126, 8127, 8128, 8130, 8131, 8132, 8133, 8135, 8137, 8138, 8141, 8142, 8144, 8145, 8146, 8147, 8501, 8502, 8503, 8504, 8505, 8506, 8508, 8510, 8511, 8512, 8513, 8514, 8515, 8516, 8517, 8518, 8519, 8520, 8521, 8522, 8523, 8524, 8525, 8526, 8527, 8528, 8529, 8530, 8532, 8533, 8534, 8535, 8536, 8538, 8540, 8541, 8543, 8544, 8545, 8548, 8549, 8550, 8552, 8553, 8556, 8558, 8559, 8560, 8561, 8562, 8563, 8564, 8565, 8567, 8568, 8569, 8570, 8571, 8572, 8573, 8574, 8575, 8576, 8577, 8578, 8580, 8581, 8583, 8585, 8586, 8587, 8588, 8589, 8590, 8592, 8593, 8595, 8596, 8598, 8600, 8603, 8604, 8605, 8606, 8607, 8608, 8609, 8610, 8611, 8612, 8613, 8615, 8616, 8617, 8618, 8619, 8620, 8621, 8622, 8624, 8625, 8626, 8628, 8629, 8630, 8631, 8632, 8633, 8634, 8635, 8636, 8637, 8638, 8639, 8640, 8641, 8642, 8643, 8645, 8646, 8647, 8650, 8651, 8652, 8653, 8654, 8655, 8656, 8657, 8658, 8659, 8660, 8661, 8662, 8665, 8666, 8667, 8668, 8669, 8670, 8671, 8672, 8673, 8674, 8675, 8676, 8677, 8678, 8679, 8681, 8683, 8684, 8685, 8686, 8687, 8688, 8689, 8690, 8691, 8692, 8693, 8694, 8695, 8696, 8697, 8698, 8699, 8700, 8701, 8702, 8703, 8704, 8705, 8706, 8707, 8708, 8709, 8710, 8711, 8712, 8714, 8716, 8717, 8718, 8719, 8720, 8721, 8722, 8723, 8724, 8725, 8726, 8727, 8728, 8729, 8730, 8731, 8732, 8733, 8734, 8735, 8736, 8738, 8739, 8740, 8741, 8743, 8744, 8745, 8746, 8747, 8748, 8749, 8750, 8754, 8755, 8756, 8757, 8758, 8759, 8760, 8761, 8763, 8766, 8767, 8769, 8770, 8771, 8773, 8778, 8779, 8780, 8781, 8782, 8783, 8784, 8786, 8788, 8789, 8790, 8791, 8792, 8793, 8794, 8799, 8801, 8802, 8803, 8804, 8805, 8807, 8808, 8809, 8810, 8811, 8812, 8813, 8814, 8816, 8817, 8818, 8819, 8820, 8822, 8823, 8824, 8825, 8827, 8828, 8829, 8830, 8831, 8832, 8833, 8834, 8835, 8836, 8837, 8838, 8839, 8841, 8842, 8844, 5045, 7502, 7503, 7505, 7506, 7509, 7511, 8001, 8003, 8004, 8007, 8009, 8010, 8012, 8013, 8014, 8015, 8016, 8017, 8019, 8021, 8022, 8024, 8025, 8027, 8028, 8030, 8032, 8501, 8503, 8504, 8506, 8507, 8508, 8509, 8511, 8512, 8513, 8514, 8517, 8518, 8519, 8521, 8523, 8524, 8525, 8526, 8528, 8530, 8532, 8534, 8535, 8536, 8537, 8539, 8540, 8543, 8545, 8547, 8548, 8549, 8550, 8552, 8553, 8554, 8555, 8556, 7090, 7502, 7504, 7505, 8001, 8003, 8005, 8021, 8501, 8502, 8503, 8504, 8505, 8506, 1001, 13044, 5039, 5505, 5507, 5508, 5510, 5511, 5512, 5514, 5517, 5518, 5519, 5521, 5522, 5523, 5524, 5525, 5526, 5527, 5528, 5529, 5530, 5531, 5532, 5533, 5535, 5536, 5537, 5538, 5539, 5540, 5541, 5542, 5546, 5547, 5548, 5550, 5551, 5553, 5555, 5556, 5557, 5558, 5559, 5560, 5561, 5562, 5563, 5565, 5567, 5570, 5571, 5572, 5573, 5575, 5578, 5579, 5580, 5581, 5584, 5585, 5586, 5587, 5588, 5589, 5590, 5591, 5593, 5594, 5595, 5597, 5599, 5600, 5601, 5602, 5603, 5604, 5605, 5606, 5607, 5608, 5609, 5610, 5611, 5612, 5613, 5614, 5616, 5617, 5618, 5620, 5622, 5623, 5624, 5625, 5629, 5631, 5632, 5634, 5636, 5637, 5639, 5640, 5642, 5644, 5645, 5646, 5647, 5649, 5650, 5658, 5671, 5672, 5673, 5674, 5675, 5676, 5677, 5678, 5679, 7005, 7006, 7007, 7009, 7011, 7012, 7013, 7014, 7016, 7017, 7018, 7019, 7020, 7022, 7025, 7027, 7028, 7029, 7030, 7031, 7032, 7033, 7034, 7036, 7037, 7039, 7040, 7041, 7043, 7044, 7045, 7046, 7047, 7048, 7049, 7050, 7051, 7056, 7058, 7059, 7060, 7061, 7062, 7063, 7065, 7067, 7068, 7069, 7070, 7073, 7076, 7078, 7080, 7082, 7083, 7084, 7086, 7088, 7090, 7091, 7092, 7094, 7095, 7096, 7097, 7098, 7102, 7103, 7104, 7105, 7109, 7110, 7111, 7112, 7117, 7120, 7121, 7124, 7126, 7127, 7501, 7502, 7503, 7504, 7505, 7506, 7508, 7509, 7510, 7512, 7513, 7515, 7516, 7517, 7519, 7520, 7521, 7523, 7524, 7525, 7526, 7530, 7531, 7532, 7533, 7534, 7535, 7536, 7537, 7538, 7539, 7541, 7544, 7545, 7551, 7552, 7553, 7555, 7556, 7557, 7560, 7561, 7562, CAX271/U2B004, 5501, 5502, 5503, 5506, 5511, 5512, 5513, 5514, 5516, 5517, 5519, 5521, 5523, 5525, 5526, 5527, 5529, 7003, 7004, 7005, 7006, 7008, 7011, 7013, 7015, 7017, 7018, 7019, 7020, 7021, 7022, 7024, 7025, 7026, 7501, 7504, 7506, 7507, 7508, 7509, 7513, 5101, 5102, 5104, 5109, 5110, 5112, 5113, 5115, 5116, 5118, 5121, 5125, 5126, 5127, 5128, 5131, 5132, 5133, 5135, 5136, 5142, 5144, 5146, 5147, 5148, 5149, 5150, 5151, 5153, 5157, 5158, 5159, 5160, 5162, 5165, 5167, 5168, 5170, 5171, 5172, 5173, 5174, 5502, 1003, 548, 550, 558, 564, 579, 581, 588, 589, 590, 592, 593, 595, RC528, RC537, RC549, 551, 584, RC521, RC536, 10610, 21511, 510, 529, 530, 557, 559, 573, 62310, RC498, RC543, 556, 572, RC493, RC524, RC538, RC541, RC545, RC546, 8501, 8503, 8504, 8505, 8506, 8508, 8509, 8510, 8511, 100003, 1234, 1234578, 4916, 60151, 63010, 7409, 7502, 8191, 8501, 8502, 8503, 8506, 8507, 8508, 8509, 8512, 8514, 8515, 8517, 8519, 8520, 8521, 8522, 8523, 8524, 8525, 8529, 8530, 8531, 8532, 8533, 8534, 8537, 8538, 8539, 8540, 8541, 8542, 8543, 8544, 8546, 8547, 8548, 8549, 8550, 8551, 8552, 8553, 8554, 8555, 8556, 8557, 8560, 8561, 8562, 8563, 8564, 8565, 8566, 8567, 8568, 8569, 8570, 8571, 8572, 8573, 8574, 8577, 8578, 8579, 8580, 8581, 8583, 9003, 9004, 9006, 9007, 9008, 9009, 9010, 9011, 9012, 9013, 9014, 9015, 9016, 9017, 9018, 9019, 9022, 9023, 9024, 9026, 9027, 9028, 9029, 9030, 9031, 9032, 9033, 9034, 9035, 9036, 9037, 9038, 9039, 9041, 9043, 8473046345, 8503, 8205, 8505, 8506, 8509, 8510, 8511, 8513, 8514, 8515, 8516, 8517, 8518, 8519, 8520, 9821, 8501, 8501*, 8502, 8507, 8508, 10000, 10001, 10002, 10003, 10005, 10006, 10007, 10008, 10009, 10010, 10013, 10014, 10015, 10016, 10017, 10019, 10020, 10022, 10023, 10024, 10025, 10026, 10028, 10029, 10030, 10031, 10032, 10033, 10034, 10035, 10036, 10037, 10038, 10040, 10041, 10042, 10044, 10046, 10047, 10048, 10049, 10050, 10051, 10052, 10053, 10054, 10055, 10056, 10057, 10058, 10059, 10060, 10061, 10062, 10063, 10065, 10066, 10067, 10068, 10069, 10070, 10071, 10072, 10073, 10074, 10076, 10077, 10078, 10079, 10080, 10082, 10083, 10084, 10085, 10087, 10088, 10089, 10090, 10091, 10092, 10093, 10094, 10096, 10098, 10099, 10100, 10101, 10102, 10103, 10104, 10106, 10107, 10108, 10109, 10111, 10112, 10113, 10114, 10115, 10117, 10118, 10119, 10120, 10121, 10122, 10124, 10125, 10126, 10127, 10128, 10129, 10130, 10131, 10132, 10133, 10135, 10136, 10137, 10138, 10139, 10140, 10141, 10142, 10143, 10144, 10147, 10149, 10150, 10151, 10153, 10154, 10155, 10156, 10157, 10158, 10159, 10160, 10161, 10162, 10163, 10164, 10165, 10166, 10167, 10168, 10169, 10170, 10171, 10172, 10173, 10174, 10175, 10176, 10177, 10178, 10179, 10180, 10181, 10182, 10183, 10184, 10185, 10186, 10187, 10188, 10191, 10192, 10193, 10194, 10195, 10198, 10199, 10200, 10201, 10202, 10203, 10204, 10205, 10206, 10207, 10208, 10209, 10211, 10212, 10213, 10214, 10215, 10216, 10217, 10218, 10219, 10220, 10221, 10222, 10224, 10225, 10226, 10227, 10228, 10230, 10231, 10232, 10233, 10234, 10235, 10236, 10237, 10238, 10239, 10240, 10241, 10243, 10245, 10246, 10248, 10249, 10250, 10251, 10252, 10253, 10254, 10255, 10256, 10257, 10258, 10259, 10260, 10261, 10262, 10263, 10264, 10265, 10266, 10267, 10268, 10269, 10270, 10271, 10272, 10273, 10274, 10275, 10276, 10277, 10278, 10280, 10282, 10283, 10284, 10285, 10286, 10287, 10288, 10289, 10290, 10291, 10292, 10293, 10294, 10295, 10296, 10297, 10298, 10299, 10300, 10301, 10302, 10303, 10304, 10305, 10306, 10307, 10308, 10309, 10310, 10311, 10312, 10313, 10314, 10315, 10317, 10318, 10319, 10320, 10321, 10322, 10323, 10324, 10325, 10326, 10327, 10328, 10329, 10330, 10331, 10332, 10333, 10334, 10336, 10337, 10338, 10339, 10340, 10341, 10343, 10344, 5382, 5517, 5764, 8490, 9001, 9002, 9003, 9004, 9005, 9006, 9008, 9009, 9010, 9012, 9013, 9014, 9015, 9016, 9017, 9018, 9019, 9022, 9023, 9024, 9025, 9026, 9027, 9028, 9031, 9033, 9034, 9035, 9036, 9038, 9039, 9040, 9041, 9043, 9044, 9045, 9046, 9047, 9049, 9050, 9051, 9053, 9054, 9055, 9057, 9058, 9059, 9060, 9061, 9062, 9063, 9064, 9066, 9068, 9069, 9070, 9073, 9074, 9076, 9077, 9078, 9079, 9080, 9081, 9082, 9083, 9084, 9085, 9086, 9087, 9089, 9090, 9091, 9092, 9094, 9095, 9096, 9097, 9098, 9099, 9100, 9101, 9102, 9103, 9104, 9106, 9107, 9108, 9109, 9110, 9111, 9112, 9113, 9114, 9115, 9116, 9117, 9120, 9122, 9123, 9124, 9125, 9126, 9127, 9129, 9130, 9132, 9133, 9134, 9135, 9136, 9137, 9138, 9139, 9142, 9143, 9144, 9145, 9146, 9147, 9148, 9149, 9150, 9151, 9152, 9154, 9155, 9156, 9157, 9158, 9159, 9161, 9162, 9163, 9164, 9165, 9166, 9167, 9169, 9170, 9171, 9172, 9173, 9174, 9175, 9176, 9177, 9179, 9180, 9181, 9182, 9183, 9184, 9185, 9186, 9187, 9189, 9190, 9191, 9193, 9194, 9198, 9199, 9200, 9201, 9204, 9206, 9207, 9210, 9212, 9213, 9214, 9215, 9216, 9219, 9220, 9221, 9223, 9224, 9225, 9226, 9227, 9228, 9230, 9231, 9232, 9233, 9234, 9235, 9238, 9239, 9240, 9241, 9242, 9243, 9245, 9246, 9247, 9248, 9251, 9252, 9253, 9254, 9255, 9256, 9259, 9260, 9261, 9262, 9263, 9264, 9265, 9266, 9267, 9268, 9269, 9270, 9271, 9272, 9273, 9274, 9275, 9276, 9277, 9279, 9280, 9281, 9282, 9283, 9284, 9285, 9287, 9288, 9289, 9290, 9292, 9293, 9294, 9295, 9296, 9297, 9298, 9299, 9300, 9301, 9302, 9303, 9304, 9305, 9306, 9307, 9308, 9309, 9311, 9313, 10001, 10002, 10003, 10004, 10005, 10006, 10007, 10008, 10009, 10011, 10012, 10014, 10015, 10016, 10017, 10018, 10019, 10022, 10023, 10025, 10026, 10027, 10028, 10029, 10030, 10031, 10032, 10034, 10035, 10036, 10038, 10039, 10040, 10041, 10044, 10045, 10046, 10047, 10048, 10049, 10050, 10051, 10052, 10055, 10056, 10057, 10058, 10059, 10061, 8505, 8509, 9001, 9002, 9003, 9004, 9005, 9007, 9008, 9009, 9010, 9011, 9012, 9013, 9014, 9015, 9016, 9017, 9019, 9020, 9021, 9022, 9024, 9026, 9031, 9032, 9033, 9034, 9036, 9037, 9038, 9039, 9040, 9041, 9042, 9043, 9044, 9045, 9001, 9002, 9005, 9006, 9007, 9009, 9010, 9011, 9013, 9014, 9016, and 9017.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0208-2016

Related Recalls

MAMMOMAT Inspiration;

Siemens Medical Solutions USA

Class I - Dangerous

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Oct 24, 2025 Other Medical Devices Nationwide View Details →

MAMMOMAT Fusion;

Siemens Medical Solutions USA

Class I - Dangerous

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Oct 24, 2025 Other Medical Devices Nationwide View Details →

MAMMOMAT Revelation;

Siemens Medical Solutions USA

Class I - Dangerous

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Oct 24, 2025 Other Medical Devices Nationwide View Details →