Medical convenience kits Item Number/Description BFCT04X CATARACT PACK - 205986 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 PERP25AQ PEI RETINAL PACK-206025 RFEY50AC EYE PACK-206026 UIVT47AS VITRECTOMY SUPPLY
Class I - DangerousWhat Should You Do?
- Check if you have this product: BFCT04X UDI-DI 191072225955 Lot 7992111 bag serial number N/A; CMEY17AI UDI-DI 191072211514 lot 8082511 bag serial numbers 17390960 17390970 17391031; PERP25AQ UDI-DI 191072199720 lot 8078511 bag serial number 17507000; RFEY50AC UDI-DI 191072202918 lot 8048811 bag serial number 17483686; UIVT47AS UDI-DI 191072224569 lot 7969711 bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- American Contract Systems Inc.
- Reason for Recall:
- Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Medical convenience kits Item Number/Description BFCT04X CATARACT PACK - 205986 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 PERP25AQ PEI RETINAL PACK-206025 RFEY50AC EYE PACK-206026 UIVT47AS VITRECTOMY SUPPLY
Product Codes/Lot Numbers:
BFCT04X UDI-DI 191072225955 Lot 7992111 bag serial number N/A; CMEY17AI UDI-DI 191072211514 lot 8082511 bag serial numbers 17390960 17390970 17391031; PERP25AQ UDI-DI 191072199720 lot 8078511 bag serial number 17507000; RFEY50AC UDI-DI 191072202918 lot 8048811 bag serial number 17483686; UIVT47AS UDI-DI 191072224569 lot 7969711 bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0185-2026
Related Recalls
Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK CORONARY ANGIO PACK- 205991 RCAN45C ANGIO CUSTOM RCH UDCO10F CVL CORONARY PACK 219224
American Contract Systems
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Medical convenience kits Item Number/Description BUCY78E CYSTO PACK CMCP10X CYSTO PACK - 205972 IHCY10Y CYSTO PACK MHCY35AE CYSTO PACK- 212487 PLBY52E CYSTO PACK UDCY31Z CYSTO PACK - 206047 UICY28L MOR CYSTO PACK
American Contract Systems
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Medical convenience kits Item number/ Product description AHLD15X LABOR DELIVERY PACK - 213387 AKDV96B VAGINAL DELIVERY AKDV96B VAGINAL DELIVERY EVBH28D BIRTHING ROOM PACK UPVD01K VAGINAL DELIVERY PACK
American Contract Systems
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.