Robotics Pack, REF SJVO56B
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI/DI 00191072122339, Lot Numbers: 965231
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- American Contract Systems, Inc.
- Reason for Recall:
- During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Robotics Pack, REF SJVO56B
Product Codes/Lot Numbers:
UDI/DI 00191072122339, Lot Numbers: 965231
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0178-2024
Related Recalls
Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ
American Contract Systems
Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.
Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK CORONARY ANGIO PACK- 205991 RCAN45C ANGIO CUSTOM RCH UDCO10F CVL CORONARY PACK 219224
American Contract Systems
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.