Dimension Vista Protein 1 Control, PROT 1 CON L, Siemens Material Number (SMN) 10445919, Product Code KC715; IVD
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot numbers: 6BQL01, 6BQL01A, 6BQL01B, 6BQL01C, 6BQL01D (exp. date 2017-06-13); 6JQL01A, 6JQL01B, 6JQL01C (exp. date 2018-01-31). --- Lot # (UDI): 6BQL01 (008427680063336BQL0120170613), 6BQL01A (008427680063406BQM01A20170617), 6BQL01B (008427680063406BQM01B20170617), 6BQL01C (008427680063406BQM01C20170617), 6BQL01D (008427680063406BQM01D20170617), 6JQL01A (008427680063336JQL01A20180131), 6JQL01B (008427680063336JQL01B20180131), 6JQL01C (008427680063336JQL01C20180131).
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Healthcare Diagnostics, Inc.
- Reason for Recall:
- Complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Dimension Vista Protein 1 Control, PROT 1 CON L, Siemens Material Number (SMN) 10445919, Product Code KC715; IVD
Product Codes/Lot Numbers:
Lot numbers: 6BQL01, 6BQL01A, 6BQL01B, 6BQL01C, 6BQL01D (exp. date 2017-06-13); 6JQL01A, 6JQL01B, 6JQL01C (exp. date 2018-01-31). --- Lot # (UDI): 6BQL01 (008427680063336BQL0120170613), 6BQL01A (008427680063406BQM01A20170617), 6BQL01B (008427680063406BQM01B20170617), 6BQL01C (008427680063406BQM01C20170617), 6BQL01D (008427680063406BQM01D20170617), 6JQL01A (008427680063336JQL01A20180131), 6JQL01B (008427680063336JQL01B20180131), 6JQL01C (008427680063336JQL01C20180131).
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0174-2018
Related Recalls
Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.