IMRIS Operating Suite, iMRI 3T S, containing Siemens 3 Tesla MRI Magnet, Skyra
Class I - DangerousWhat Should You Do?
- Check if you have this product: IMRIS Part Number(s): 116053-000, UDI/DI 00857534006035; 116117-000, UDI/DI 00857534006042; 116118-000, UDI/DI 00857534006059; 116119-000, UDI/DI 00857534006066.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- IMPRIS Imaging Inc
- Reason for Recall:
- There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
IMRIS Operating Suite, iMRI 3T S, containing Siemens 3 Tesla MRI Magnet, Skyra
Product Codes/Lot Numbers:
IMRIS Part Number(s): 116053-000, UDI/DI 00857534006035; 116117-000, UDI/DI 00857534006042; 116118-000, UDI/DI 00857534006059; 116119-000, UDI/DI 00857534006066.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0162-2026
Related Recalls
IMRIS Operating Suite, Neuro III-SV iMRI System, iMRX System, containing Siemens 3 Tesla MRI Magnet, Verio
IMPRIS Imaging
There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
IMRIS Operating Suite, InVision 3T Operating Suite, containing Siemens 3 Tesla MRI Magnet, Vida
IMPRIS Imaging
There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.