Terumo Custom Cardiovascular and Perfusion Procedure Kits Catalog Codes: Kits with a 5-digit code beginning with the digit 6: Catalog Number: 6XXXX ; Catalog Codes: Kits with a 5-digit code beginning with the digit 7: Catalog Number: 7XXXX Product usage: The Cardioplegia Delivery Sets are indicated for delivery of cardioplegic solutions to the heart during cardiopulmonary bypass surgery. The Cardiovascular Procedure Kits are indicated for use only in the extracorporeal circuit for the cardiopulmonary bypass procedure(s) for which the user designed it.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Catalog Number: 6XXXX Lot Numbers: ML07, MM18, MM25, MN08, MN IS, MN22. Manufacturing dates September 7, 20 I 0 through November 22, 20 I 0. Catalog Number: 7XXXX Lot Numbers: ML07,ML20, MM18, MM25, MNO1, MN08, MN15, MN29,MP06, MP13. Manufacturing dates September 7, 2010 through December 13,2010.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Terumo Cardiovascular Systems Corp
- Reason for Recall:
- Defect in the luer connector can compromise the seal on the female side of the connector and the compromised seal can introduce air in the table line.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Terumo Custom Cardiovascular and Perfusion Procedure Kits Catalog Codes: Kits with a 5-digit code beginning with the digit 6: Catalog Number: 6XXXX ; Catalog Codes: Kits with a 5-digit code beginning with the digit 7: Catalog Number: 7XXXX Product usage: The Cardioplegia Delivery Sets are indicated for delivery of cardioplegic solutions to the heart during cardiopulmonary bypass surgery. The Cardiovascular Procedure Kits are indicated for use only in the extracorporeal circuit for the cardiopulmonary bypass procedure(s) for which the user designed it.
Product Codes/Lot Numbers:
Catalog Number: 6XXXX Lot Numbers: ML07, MM18, MM25, MN08, MN IS, MN22. Manufacturing dates September 7, 20 I 0 through November 22, 20 I 0. Catalog Number: 7XXXX Lot Numbers: ML07,ML20, MM18, MM25, MNO1, MN08, MN15, MN29,MP06, MP13. Manufacturing dates September 7, 2010 through December 13,2010.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0159-2015
Related Recalls
CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial Kit
Terumo Cardiovascular Systems
Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.
Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754
Terumo Cardiovascular Systems
During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be triggered to the user, due to the dashed-out value.
Pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmHg versus 400mmHg, potential myocardial tissue damage potentially requiring surgical or medical/pharmacological intervention