ORBIT PT. The ORBIT PT works in conjunction with an TRUMPF mobile operating table to automate the patient transport process.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Item No. 1275499, Serial No. 100711351
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Trumpf Medical Systems
- Reason for Recall:
- An odor and smoke developed - The cause was determined to be a power supply unit sitting in the controller column.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
ORBIT PT. The ORBIT PT works in conjunction with an TRUMPF mobile operating table to automate the patient transport process.
Product Codes/Lot Numbers:
Item No. 1275499, Serial No. 100711351
Distribution:
Distributed in: US, NJ
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0155-2014
Related Recalls
Camera System, Television, Surgical without audio Intended for transmission and documentation of surgical interventions for the purpose of archiving and teaching.
Trumpf Medical Systems
The radiated emission from the TruVidia" Wireless Receiver is above the normative accepted limits, in which the radiated emission may interact with other devices in the operating room and cause them not to function properly.
Trumpf has identified six incidents of pinched cables.
Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters.