Hillrom Hospital Bed Accessory, Watch Care Incontinence Management System for VersaCare Bed Rev. A-J, Product Code/Part Number P00697901

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI/DI 00887761998865, all serial numbers
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Baxter Healthcare Corporation
Reason for Recall:
Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Hillrom Hospital Bed Accessory, Watch Care Incontinence Management System for VersaCare Bed Rev. A-J, Product Code/Part Number P00697901

Product Codes/Lot Numbers:

UDI/DI 00887761998865, all serial numbers

Distribution:

Distributed in: US

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0145-2023

Related Recalls

Certain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication on the cam, which could lead to premature wear of the mechanism assembly, resulting in over-infusion or a free-flow situation.

Nov 28, 2025 Infusion Pumps Nationwide View Details →