1.5T SIGNA HDxt MR System nuclear magnetic resonance imaging system

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Software version HD29.1; and HD16.2 (China only)
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
GE Healthcare, LLC
Reason for Recall:
Under certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

1.5T SIGNA HDxt MR System nuclear magnetic resonance imaging system

Product Codes/Lot Numbers:

Software version HD29.1; and HD16.2 (China only)

Distribution:

Distributed in: AZ, CA, FL, GA, IL, LA, MA, MD, MN, NC, NH, NJ, NY, OH, OK, TX, VA, WA, UK

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0141-2022

Related Recalls

Class I - Dangerous

Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.

Feb 27, 2026 Other Medical Devices Nationwide View Details →

Under certain workflows, patient information shown in the viewer may not match the images displayed on Centricity Universal Viewer Zero Footprint Client (ZFP) with versions ZFP v6.0: SP11.6, SP11.6.0.1 and SP11.6.1 on systems with Patient Timeline installed,

Jan 16, 2026 Infusion Pumps Nationwide View Details →

There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.

Feb 19, 2025 Infusion Pumps View Details →