Animas Vibe Insulin Infusion Pump. This product is indicated for continuous subcutaneous infusion of insulin for the treatment of diabetes and has a continuous glucose monitoring feature.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model Number(s): 11220008 PMPW/OKIT,VIBE,BLU,mg,IL 11220108 PMPW/OKIT,VIBE,BLK,mg,IL 11220208 PMPW/OKIT,VIBE,SIL,mg,IL 11220408 PMPW/OKIT,VIBE,PGL,mg,IL 11220508 PMPW/OKIT,VIBE,LGN,mg,IL 11251008 REV KIT,VIBE,BLU,mg,IL 11251108 REV KIT,VIBE,BLK,mg,IL 11251208 REV KIT,VIBE,SIL,mg,IL 11251408 REV KIT,VIBE,PGL,mg,IL 11251508 REV KIT,VIBE,LGN,mg,IL 11451008 REV KIT,VIBEPL,BLU,mg,IL 11451108 REV KIT,VIBEPL,BLK,mg,IL 11451208 REV KIT,VIBEPL,SIL,mg,IL 11451408 REV KIT,VIBEPL,PGL,mg, IL 11451508 REV KIT,VIBEPL,LGN,mg,IL 11420008 PMPW/OKIT,VIBEPL,BLU,mg,IL 11420108 PMPW/OKIT,VIBEPL,BLK,mg,IL 11420208 PMPW/OKIT,VIBEPL,SIL,mg,IL 11420408 PMPW/OKIT,VIBEPL,PGL,mg,IL 11420508 PMPW/OKIT,VIBEPL,LGN,mg,IL
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Animas Corporation
Reason for Recall:
Animas determined that the Hebrew language translation in the Vibe Insulin Infusion Pump was incorrect and could lead to potential patient confusion. In addition, one line of Hebrew text was omitted from the display screen.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Animas Vibe Insulin Infusion Pump. This product is indicated for continuous subcutaneous infusion of insulin for the treatment of diabetes and has a continuous glucose monitoring feature.

Product Codes/Lot Numbers:

Model Number(s): 11220008 PMPW/OKIT,VIBE,BLU,mg,IL 11220108 PMPW/OKIT,VIBE,BLK,mg,IL 11220208 PMPW/OKIT,VIBE,SIL,mg,IL 11220408 PMPW/OKIT,VIBE,PGL,mg,IL 11220508 PMPW/OKIT,VIBE,LGN,mg,IL 11251008 REV KIT,VIBE,BLU,mg,IL 11251108 REV KIT,VIBE,BLK,mg,IL 11251208 REV KIT,VIBE,SIL,mg,IL 11251408 REV KIT,VIBE,PGL,mg,IL 11251508 REV KIT,VIBE,LGN,mg,IL 11451008 REV KIT,VIBEPL,BLU,mg,IL 11451108 REV KIT,VIBEPL,BLK,mg,IL 11451208 REV KIT,VIBEPL,SIL,mg,IL 11451408 REV KIT,VIBEPL,PGL,mg, IL 11451508 REV KIT,VIBEPL,LGN,mg,IL 11420008 PMPW/OKIT,VIBEPL,BLU,mg,IL 11420108 PMPW/OKIT,VIBEPL,BLK,mg,IL 11420208 PMPW/OKIT,VIBEPL,SIL,mg,IL 11420408 PMPW/OKIT,VIBEPL,PGL,mg,IL 11420508 PMPW/OKIT,VIBEPL,LGN,mg,IL

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0136-2015

Related Recalls

Calibration factors in the pump overwritten during a programming step. The force sensor could send a lower signal value to the pump processor, with loss of prime warnings, occlusion alarms and the pump unable to detect a cartridge during the prime sequence. Field action initiated 8/29/2011.

Sep 6, 2011 Diagnostic Equipment View Details →

Animas determined that keypad wear over time was contributing to an increased complaint rate associated with damaged keypads on the Vibe Insulin Infusion Pump and System. Animas has implemented an improvement to the keypad related to the material used to reduce susceptibility to damage and increase durability.

Oct 25, 2012 Infusion Pumps View Details →

Animas has identified a component issue affecting a small supply of the Animas(R) 2020 insulin pumps. The component issue may trigger the pumps to sound a false alarm or warning related to one of the following: - Loss of prime; -Occlusion; or -No cartridge detected.

Jan 3, 2013 Infusion Pumps Nationwide View Details →