XiO Radiation Treatment Planning System. Used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Version 4.51 and higher
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Elekta, Inc.
- Reason for Recall:
- Incorrect Treatment Delivery Using Third Party Fixed Wedges.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
XiO Radiation Treatment Planning System. Used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed.
Product Codes/Lot Numbers:
Version 4.51 and higher
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0130-2016
Related Recalls
Software may fail to load adapted baseline shift plan during treatment delivery when a communication error occurs.
Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.
Due to adapter not being properly aligned or locked in place.