P2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model number: P2131; NSN: 6530-01-641-4641; Serial Numbers AFS-0001 through AFS-0386. Note: AFS-0297 was not manufactured.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Fort Defiance Industries, LLC
- Reason for Recall:
- The rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
P2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641
Product Codes/Lot Numbers:
Model number: P2131; NSN: 6530-01-641-4641; Serial Numbers AFS-0001 through AFS-0386. Note: AFS-0297 was not manufactured.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0121-2022
Related Recalls
During long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system that can potentially interfere with the rear sight gauge and / or the water liquid level switch.