Monoject 0.9% Sodium Chloride, Flush Prefilled Syringe, 10 mL, REF: 8881570121 inside the following Aligned Medical Solutions Custom Convenience Kits: AMS-9041CP Leaderflex Insertion Kit with Ultrasound, AMS-9046CP-1 Insertion Tray-RX, AMS8939A Universal Procedure Pack w/Split Drape, AMS9957A Port Insertion Pack, and AMS12149 Procedure Pack
Class I - DangerousWhat Should You Do?
- Check if you have this product: Syringe Lots: All Lots manufactured and distributed between July 1, 2019 and July 2021. Kit Model/Lots: AMS-9041CP/ 136833; AMS-9046CP-1/ 137794; AMS8939A/ 140390, 141088, 141096, 141100, 141102, 146039, 146349, 147902, 147903, 148414, 148452, 148640, 149132, 149135, 149485, 141087, 141097, 145615, 146351, 146766, 146768, 146905, 146908, 148638, 149137, 149138, 149484, 149486, 141101, 148410, 148411, 148639, 149133, 141098, 146037, 148409, 148413, 148450, 149131, 149134, 149136, 148412, 148451, 148636, 141717 AMS9957A/ 147773; AMS12149/ 145718, 146897, 148612
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Windstone Medical Packaging, Inc.
- Reason for Recall:
- Monoject 0.9% Sodium Chloride Flush Syringe Manufactured by Cardinal Health and placed into kits. Potential for the plunger to draw back after air has been expelled and reintroduce air back into the syringe.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Monoject 0.9% Sodium Chloride, Flush Prefilled Syringe, 10 mL, REF: 8881570121 inside the following Aligned Medical Solutions Custom Convenience Kits: AMS-9041CP Leaderflex Insertion Kit with Ultrasound, AMS-9046CP-1 Insertion Tray-RX, AMS8939A Universal Procedure Pack w/Split Drape, AMS9957A Port Insertion Pack, and AMS12149 Procedure Pack
Product Codes/Lot Numbers:
Syringe Lots: All Lots manufactured and distributed between July 1, 2019 and July 2021. Kit Model/Lots: AMS-9041CP/ 136833; AMS-9046CP-1/ 137794; AMS8939A/ 140390, 141088, 141096, 141100, 141102, 146039, 146349, 147902, 147903, 148414, 148452, 148640, 149132, 149135, 149485, 141087, 141097, 145615, 146351, 146766, 146768, 146905, 146908, 148638, 149137, 149138, 149484, 149486, 141101, 148410, 148411, 148639, 149133, 141098, 146037, 148409, 148413, 148450, 149131, 149134, 149136, 148412, 148451, 148636, 141717 AMS9957A/ 147773; AMS12149/ 145718, 146897, 148612
Distribution:
Distributed in: FL, PA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0119-2022
Related Recalls
Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes) Angio Pack Part numbers AMS6908E, AMS6908F
Windstone Medical Packaging
Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full disconnection between syringe and manifold. If unwinding occurs, there is a potential for biohazard exposure, blood loss, infection. or air embolism.
Custom Convenience Kits Pre Op Kit - NS, Part Number AMS14994
Windstone Medical Packaging
The product correction was initiated because B Braun is issuing a voluntary Urgent Medical Device Correction for gravity IV administration sets as well as pump administration sets utilized with BBMI s Infusomat Space Large Volume Pump, Outlook Pump and Vista Basic Pump due to potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers or other connections and the ability to prime (occlusion).
Custom Convenience Kits Pre Op Kit - NS, Part Number AMS14994A
Windstone Medical Packaging
The product correction was initiated because B Braun is issuing a voluntary Urgent Medical Device Correction for gravity IV administration sets as well as pump administration sets utilized with BBMI s Infusomat Space Large Volume Pump, Outlook Pump and Vista Basic Pump due to potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers or other connections and the ability to prime (occlusion).