Brand Names: Integra¿ Camino 110-4 Intracranial Pressure (ICP) Monitoring Catheters. Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively. Common Name: Intracranial Pressure and Pressure-Temperature Monitoring Kits Classification Name: Device, Monitoring, Intracranial Pressure 510(k) Number: K102875 The Camino 110-4 Intracranial Pressure monitoring catheters are sterile transducer-tipped pressure monitoring catheters that are used by neurosurgeons to rapidly determine and continuously monitor intracranial pressure or pressure/temperature.  The Camino 110-4B catheter is used for Intracranial Pressure Monitoring in the parenchyma or subarachnoid space.  The Camino 110-4G catheter is used for Post Craniotomy Subdural Pressure Monitoring.  The Camino 110-4H catheter is intended for Intracranial Pressure Monitoring and Cerebrospinal fluid drainage.  The Camino 110-4L catheter is intended for use with the Licox¿ Brain Oxygen Monitoring System to measure intracranial pressure in the parenchyma.  The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. Lot Numbers: Catalog Number 110-4B: 3050RX195484, 3050RX198473, 305000205426, 3050RX197641, 3050RX201421, 305000205648, 305000205652, 3050RX199149, 3050RX201407, 3050RX199850, 305000205976, 3050RY189979, 3050RX200862, 305000206178, 3050RY193648, 3050RX200486, 3050RX181026, 3050RX185115, 3050RY195484, 305000207061, 305000207062, 3050RY198473, 305000207330, 305000207334, 3050RY197641, 3050RY200862, 305000207580, 305000207892, 3050RY200486, 3050RY181026, 3050RY185115, 3050RX207330, 3050RX207334, 3050RX207580, 305000208008, 305000208510, 3050RY207892, 305000210060, 3050RX208008, 3050RX208510, 3050RX205976, 3050RX206178, 305000210563, 305000211723, 3050RX210060, 305000212616, 305000213007, 305000213191, 30500X210563, 3050RX207061, 305000214590, 3050RX205426, 30500X215502, 3050RX207062, 305000215531, 30500X216410, 3050RY199850, 305000216132, 30500X216627, 30500X217119, 3050RY217112, 305000216718, 3050RX217751, 3050RX217840, 3050RY217837, 305000217858, 30500X218369, 30500X218391, 30500X219054, 3050RY206178, 305000220708, 305000218578, 30500X220430, 305000219261, 30500X221221, 305000221226, 305000219989, 30500X222569, 305000221953, 305000221955, 30500X223153, 30500X223709, 305000223275, 305000223280, 305000222640, 30500X224142, 305000223867, 305000223866, 30500X224692, 305000224352, 305000224353, 30500X225386, 305000225045, 305000225044, 30500X225992, 305000225632, 305000225633, 30500X226540, 30500X227059, 305000226844, 305000226846, 305000226127, 30500X227564, 305000226836, 30500X227820, 305000227127, 305000227128, 30500X228194, 305000227790, 305000227789, 30500X228787, 305000228522, 305000228521, 305000228971, 305000228972, 30500X229560, 305000229328, 305000229329, 30500X229766, 30500X230087, 305000229825, 305000229826, 30500X230565, 305000230303, 305000230302, 30500X231053, 305000230780, 305000230781, 30500X231786, 305000231461, 305000231462, 30500X232331, 305000232098, 305000232099, 30500X232960, 305000232605, 305000232604, 305000233245, 305000233244, 305000233243, 305000233242, 305000233241, 305000233240, 30500X233415, 30500X234183, 305000233607, 305000233608, 30500X234806, 305000234303, 305000234304, 305000235022, 305000235017, 30500X235732, 305000235598, 30500X236328, 305000236105, 305000236104, 305000236655, 305000236653, 30500Y237087, 30500X237090, 305000237232, 30500X237761, 305000237824, 305000237825, 30500X238273, 305000238263, 305000238264, 30500X239095, 305000238818, 305000238819, 30500X239497, 30500X239748, 30500X239958, 30500X240334, 30500X240668, 30500X241368, 305000241516, 305000242236, 305000242792, 305000242795, 305000242796, 305000242794, 305000242793, 305000242791, 3050RY243869, 305000243251, 305000243882, 305000243881, 305000243888, 305000243889, 305000243883, 30500X244281, 30500X244880, 305000244513,30500X245306, 305000244481, 305000244480, 305000244934, 30500X245937, 30500X246067, 305000245408, 30500X246872, 305000245835, 30500X246951, 30500X247608, 305000246460, 30500X248284, 305000247007, 3050RX227127, 3050RX234303, 305000247717, 305000249072, 3050RX242794, 3050RX242793, 3050RX237825, 3050RX241516, 30500X249488, 305000248479, 30500X250423, 305000249743, 30500X250964, 305000250456, 30500X251060, 305000251055, 30500X251560, 30500X252133, 3050RY252136, 305000251453, 30500X252771, 305000252058, 3050RX244513, 305000252646, 305000253785, 30500X254929, 3050RY255145, 30500X255141, 305000254589, 305000255549, 305000255551, 305000255038, 30500X256614, 305000256192, 305000257281, 305000256709, 305000257280, 305000258048, 30500X259385, 305000258379, 305000258921, 305000258922, 30500X259557, 305000259431, 305000259432, 30500X260159, 305000259966, 305000259960, 305000260514, 305000260527, 30500Y261748, 3050RX253785, 305000261325, 305000261326, 305000261737, 305000261736, 3050RX256192, 30500X262441, 305000262328, 305000262327, 305000262978, 305000262977, 3050RY263508, 30500X263559, 305000263504, 305000264073, 305000264074, 305000264625, 305000264624, 3050RY265348, 30500X265422, 30500X265976, 3050RY265975, 305000265530, 305000265527, 30500X266142, 305000266134, 305000266135, 30500X266789, 305000267193, 305000267194, 3050RX261325, 30500X267468, 305000267421, 305000267422, 30500X267990, 305000267673, 305000267674, 305000268294, 305000268293, 305000266784, 305000268914, 305000268915, 3050RX261736, 3050RY269955, 30500X269963, 305000269220, 305000269221, 305000270081, 3050RY271045, 30500X271047, 305000271042, 305000271716, 305000271715, 30500X272367, 305000272273, 305000272274, 305000272852, 305000272851, 30500X274204, 305000273466, 305000273463, 305000274200, 305000274201, 30500X275297, 30500X275393, 305000274816, 30500X276594, 3050RX268914, 305000277761, 3050RY282196, 305000282304, 30500X282815, 3050RY283371, 3050RX283315, 3050RX283314 Catalog Number 110-4G: 3050RX195650, 3050RX198690, 3050RY193969, 3050RY197837, 3050RX199151, 3050RX199942, 3050RX201552, 3050RY185259, 3050RX200538, 305000206177, 3050RX200989, 305000206556, 305000206557, 3050RX200990, 3050RY199151, 3050RY199942, 305000207465, 3050RY200538, 3050RY200989, 305000208007, 3050RY200990, 3050RX207465, 3050RX206557, 3050RX206556, 305000209417, 305000209105, 3050RX208007, 305000210891, 305000211419, 3050RX206177, 30500X209105, 3050RX209417, 305000212298, 30500X210891, 305000213240, 305000214103, 30500X215501, 30500X215577, 3050RX202201, 305000215628, 3050RY195650, 3050RY198690, 30500X216621, 305000216246,3050RY217113, 30500X217118, 305000217140, 30500X217835, 3050RY217839, 305000217464, 30500X218393, 305000218009, 30500X219053, 305000218225, 305000219118, 30500X220001, 30500X220429, 305000219990, 30500X221225, 305000221227, 30500X222567, 305000222642, 305000223279, 30500X223707, 305000223868, 30500X224685, 305000224354, 30500X225383, 305000225046, 305000225634, 30500X226000, 30500X226541, 305000226731, 305000226869, 305000226867, 30500X227567, 305000227129, 30500X228784, 305000228523, 305000228973, 30500X229561, 305000229330, 30500X230084, 305000229827, 305000230304, 30500X231054,305000230782, 30500X231787, 305000231463, 30500X232333, 305000232101, 30500X232776, 305000232606, 30500X233419, 305000233246, 305000233609, 30500X234690, 30500X234807, 305000234306, 305000235566, 30500X236329, 30500Y237088, 30500X237767, 305000237233, 305000238820, 30500X239496, 305000239476, 30500X240336, 30500X241357, 305000241518, 30500X242172, 305000242249, 305000243252, 305000244364, 30500X245304, 305000244935, 305000245409, 305000245838, 30500X246522, 305000246308, 30500X246873, 30500X247609, 30500X248281, 305000249073, 3050RX241518, 3050RX233246, 3050RX242249, 3050RX228973, 30500X251061, 305000251056, 30500X251563, 3050RY252137, 305000252059, 305000252648, 305000254591, 30500X254931, 30500X255143, 305000255039, 305000255548, 305000256193, 305000256737, 305000257244, 305000257246, 305000257921, 305000257920, 305000258303, 305000258290, 30500X259383, 305000258793, 305000259573, 30500X260163, 305000259958, 305000259959, 3050RX254591, 3050RX255039, 305000260513, 305000260512, 30500Y261771, 3050RX257244, 3050RX257920, 305000261159, 305000261160, 3050RY262444, 30500X262443, 305000261734, 305000261732, 305000262325, 305000262326, 3050RX258303, 305000262976, 305000262975, 3050RY263510, 30500X263560, 305000263507, 305000263506, 305000264075, 305000264076, 305000264627, 305000264628, 30500X265424, 305000265525, 305000265524, 30500X265978, 30500X266143, 305000266063, 305000266064, 30500X266787, 305000266638, 305000266637, 305000267191, 30500X267470, 305000267423, 305000267675, 3050RY269957, 30500X274206 Catalog Number 110-4H: 3050RX198281, 3050RY194144, 3050RY198281, 3050RY200861, 30500X223154, 305000220948, 305000223622, 30500X225385, 30500X233417, 305000231714, 305000241350, 30500X243874, 305000250959, 30500X254932, 305000263557 Catalog Number 110-4L: 305000221383, 30500X222566, 305000222762, 30500X223971,305000223969, 305000225078, 30500X226539, 305000226221, 30500X227058, 305000226848, 305000226749, 305000226847, 305000226871, 305000226,870, 30500X228783, 305000229450, 305000229909, 305000230871 30500X232775, 305000232199, 30500X232961, 305000233311, 305000234438, 30500X235731, 305000235181, 305000236201, 305000237368, 305000237369, 305000237799, 305000238539, 305000239102, 30500X240245, 30500X240246, 30500X240247, 305000239736, 305000239737, 30500X241365, 30500X241522, 305000241082, 30500X242170, 30500X245765, 30500X246521, 3050RX237799, 3050RX226749, 3050RX229450, 3050RX241082, 3050RX232199, 3050RY252775, 3050RX240446, 3050RY255148, 305000259553, 305000260626, 305000261835, 305000262436, 3050RX259553, 305000263146, 3050RX260626, 305000264077, 3050RX262436, 3050RX261835, 305000264175, 3050RY265351, 3050RY265766, 305000265531, 30500X266144, 3050RY266146, 30500X266522, 305000266062, 305000267297, 30500X267473, 30500X268133, 305000267784, 305000268412, 3050RX265531, 305000268916, 3050RY271046, 305000271872, 30500X272374, 305000272591, 30500X274210, 30500X274963, 305000281820 Catalog Number 110-4BC: 176044, 177018, 178013, 178898, 179757, 180139, 180955, 182279, 182908, 184079, 184687, 185284, 186277, 187293, 190452, 187780, 190451, 191891, 193414, 194074, 195161, 195608, 196987, 197492, 198189, 198512, 200069, 200490, 199609, 201039, 201595, 202423, 203565, 207077, 208532 Catalog Number 110-4HC: 179758, 180956, 183071, 184686, 186278, 187781, 190450, 192246, 194993, 195450, 198191, 199708, 201040, 201596, 202009 Catalog Number 110-4LC: 162523, 167450, 167786, 175720, 177188, 178547, 182130, 187495, 195245, 197493.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Integra Limited
Reason for Recall:
Integra LifeSciences has identified through an internal evaluation that in order to comply with US FDA requirements, certain Camino¿ Intracranial Pressure Monitoring Kits require additional labelling to indicate they are UNSAFE in an MR environment. Only the following Camino¿ Intracranial Pressure Monitoring Kits are affected: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Brand Names: Integra¿ Camino 110-4 Intracranial Pressure (ICP) Monitoring Catheters. Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively. Common Name: Intracranial Pressure and Pressure-Temperature Monitoring Kits Classification Name: Device, Monitoring, Intracranial Pressure 510(k) Number: K102875 The Camino 110-4 Intracranial Pressure monitoring catheters are sterile transducer-tipped pressure monitoring catheters that are used by neurosurgeons to rapidly determine and continuously monitor intracranial pressure or pressure/temperature.  The Camino 110-4B catheter is used for Intracranial Pressure Monitoring in the parenchyma or subarachnoid space.  The Camino 110-4G catheter is used for Post Craniotomy Subdural Pressure Monitoring.  The Camino 110-4H catheter is intended for Intracranial Pressure Monitoring and Cerebrospinal fluid drainage.  The Camino 110-4L catheter is intended for use with the Licox¿ Brain Oxygen Monitoring System to measure intracranial pressure in the parenchyma.  The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively.

Product Codes/Lot Numbers:

Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. Lot Numbers: Catalog Number 110-4B: 3050RX195484, 3050RX198473, 305000205426, 3050RX197641, 3050RX201421, 305000205648, 305000205652, 3050RX199149, 3050RX201407, 3050RX199850, 305000205976, 3050RY189979, 3050RX200862, 305000206178, 3050RY193648, 3050RX200486, 3050RX181026, 3050RX185115, 3050RY195484, 305000207061, 305000207062, 3050RY198473, 305000207330, 305000207334, 3050RY197641, 3050RY200862, 305000207580, 305000207892, 3050RY200486, 3050RY181026, 3050RY185115, 3050RX207330, 3050RX207334, 3050RX207580, 305000208008, 305000208510, 3050RY207892, 305000210060, 3050RX208008, 3050RX208510, 3050RX205976, 3050RX206178, 305000210563, 305000211723, 3050RX210060, 305000212616, 305000213007, 305000213191, 30500X210563, 3050RX207061, 305000214590, 3050RX205426, 30500X215502, 3050RX207062, 305000215531, 30500X216410, 3050RY199850, 305000216132, 30500X216627, 30500X217119, 3050RY217112, 305000216718, 3050RX217751, 3050RX217840, 3050RY217837, 305000217858, 30500X218369, 30500X218391, 30500X219054, 3050RY206178, 305000220708, 305000218578, 30500X220430, 305000219261, 30500X221221, 305000221226, 305000219989, 30500X222569, 305000221953, 305000221955, 30500X223153, 30500X223709, 305000223275, 305000223280, 305000222640, 30500X224142, 305000223867, 305000223866, 30500X224692, 305000224352, 305000224353, 30500X225386, 305000225045, 305000225044, 30500X225992, 305000225632, 305000225633, 30500X226540, 30500X227059, 305000226844, 305000226846, 305000226127, 30500X227564, 305000226836, 30500X227820, 305000227127, 305000227128, 30500X228194, 305000227790, 305000227789, 30500X228787, 305000228522, 305000228521, 305000228971, 305000228972, 30500X229560, 305000229328, 305000229329, 30500X229766, 30500X230087, 305000229825, 305000229826, 30500X230565, 305000230303, 305000230302, 30500X231053, 305000230780, 305000230781, 30500X231786, 305000231461, 305000231462, 30500X232331, 305000232098, 305000232099, 30500X232960, 305000232605, 305000232604, 305000233245, 305000233244, 305000233243, 305000233242, 305000233241, 305000233240, 30500X233415, 30500X234183, 305000233607, 305000233608, 30500X234806, 305000234303, 305000234304, 305000235022, 305000235017, 30500X235732, 305000235598, 30500X236328, 305000236105, 305000236104, 305000236655, 305000236653, 30500Y237087, 30500X237090, 305000237232, 30500X237761, 305000237824, 305000237825, 30500X238273, 305000238263, 305000238264, 30500X239095, 305000238818, 305000238819, 30500X239497, 30500X239748, 30500X239958, 30500X240334, 30500X240668, 30500X241368, 305000241516, 305000242236, 305000242792, 305000242795, 305000242796, 305000242794, 305000242793, 305000242791, 3050RY243869, 305000243251, 305000243882, 305000243881, 305000243888, 305000243889, 305000243883, 30500X244281, 30500X244880, 305000244513,30500X245306, 305000244481, 305000244480, 305000244934, 30500X245937, 30500X246067, 305000245408, 30500X246872, 305000245835, 30500X246951, 30500X247608, 305000246460, 30500X248284, 305000247007, 3050RX227127, 3050RX234303, 305000247717, 305000249072, 3050RX242794, 3050RX242793, 3050RX237825, 3050RX241516, 30500X249488, 305000248479, 30500X250423, 305000249743, 30500X250964, 305000250456, 30500X251060, 305000251055, 30500X251560, 30500X252133, 3050RY252136, 305000251453, 30500X252771, 305000252058, 3050RX244513, 305000252646, 305000253785, 30500X254929, 3050RY255145, 30500X255141, 305000254589, 305000255549, 305000255551, 305000255038, 30500X256614, 305000256192, 305000257281, 305000256709, 305000257280, 305000258048, 30500X259385, 305000258379, 305000258921, 305000258922, 30500X259557, 305000259431, 305000259432, 30500X260159, 305000259966, 305000259960, 305000260514, 305000260527, 30500Y261748, 3050RX253785, 305000261325, 305000261326, 305000261737, 305000261736, 3050RX256192, 30500X262441, 305000262328, 305000262327, 305000262978, 305000262977, 3050RY263508, 30500X263559, 305000263504, 305000264073, 305000264074, 305000264625, 305000264624, 3050RY265348, 30500X265422, 30500X265976, 3050RY265975, 305000265530, 305000265527, 30500X266142, 305000266134, 305000266135, 30500X266789, 305000267193, 305000267194, 3050RX261325, 30500X267468, 305000267421, 305000267422, 30500X267990, 305000267673, 305000267674, 305000268294, 305000268293, 305000266784, 305000268914, 305000268915, 3050RX261736, 3050RY269955, 30500X269963, 305000269220, 305000269221, 305000270081, 3050RY271045, 30500X271047, 305000271042, 305000271716, 305000271715, 30500X272367, 305000272273, 305000272274, 305000272852, 305000272851, 30500X274204, 305000273466, 305000273463, 305000274200, 305000274201, 30500X275297, 30500X275393, 305000274816, 30500X276594, 3050RX268914, 305000277761, 3050RY282196, 305000282304, 30500X282815, 3050RY283371, 3050RX283315, 3050RX283314 Catalog Number 110-4G: 3050RX195650, 3050RX198690, 3050RY193969, 3050RY197837, 3050RX199151, 3050RX199942, 3050RX201552, 3050RY185259, 3050RX200538, 305000206177, 3050RX200989, 305000206556, 305000206557, 3050RX200990, 3050RY199151, 3050RY199942, 305000207465, 3050RY200538, 3050RY200989, 305000208007, 3050RY200990, 3050RX207465, 3050RX206557, 3050RX206556, 305000209417, 305000209105, 3050RX208007, 305000210891, 305000211419, 3050RX206177, 30500X209105, 3050RX209417, 305000212298, 30500X210891, 305000213240, 305000214103, 30500X215501, 30500X215577, 3050RX202201, 305000215628, 3050RY195650, 3050RY198690, 30500X216621, 305000216246,3050RY217113, 30500X217118, 305000217140, 30500X217835, 3050RY217839, 305000217464, 30500X218393, 305000218009, 30500X219053, 305000218225, 305000219118, 30500X220001, 30500X220429, 305000219990, 30500X221225, 305000221227, 30500X222567, 305000222642, 305000223279, 30500X223707, 305000223868, 30500X224685, 305000224354, 30500X225383, 305000225046, 305000225634, 30500X226000, 30500X226541, 305000226731, 305000226869, 305000226867, 30500X227567, 305000227129, 30500X228784, 305000228523, 305000228973, 30500X229561, 305000229330, 30500X230084, 305000229827, 305000230304, 30500X231054,305000230782, 30500X231787, 305000231463, 30500X232333, 305000232101, 30500X232776, 305000232606, 30500X233419, 305000233246, 305000233609, 30500X234690, 30500X234807, 305000234306, 305000235566, 30500X236329, 30500Y237088, 30500X237767, 305000237233, 305000238820, 30500X239496, 305000239476, 30500X240336, 30500X241357, 305000241518, 30500X242172, 305000242249, 305000243252, 305000244364, 30500X245304, 305000244935, 305000245409, 305000245838, 30500X246522, 305000246308, 30500X246873, 30500X247609, 30500X248281, 305000249073, 3050RX241518, 3050RX233246, 3050RX242249, 3050RX228973, 30500X251061, 305000251056, 30500X251563, 3050RY252137, 305000252059, 305000252648, 305000254591, 30500X254931, 30500X255143, 305000255039, 305000255548, 305000256193, 305000256737, 305000257244, 305000257246, 305000257921, 305000257920, 305000258303, 305000258290, 30500X259383, 305000258793, 305000259573, 30500X260163, 305000259958, 305000259959, 3050RX254591, 3050RX255039, 305000260513, 305000260512, 30500Y261771, 3050RX257244, 3050RX257920, 305000261159, 305000261160, 3050RY262444, 30500X262443, 305000261734, 305000261732, 305000262325, 305000262326, 3050RX258303, 305000262976, 305000262975, 3050RY263510, 30500X263560, 305000263507, 305000263506, 305000264075, 305000264076, 305000264627, 305000264628, 30500X265424, 305000265525, 305000265524, 30500X265978, 30500X266143, 305000266063, 305000266064, 30500X266787, 305000266638, 305000266637, 305000267191, 30500X267470, 305000267423, 305000267675, 3050RY269957, 30500X274206 Catalog Number 110-4H: 3050RX198281, 3050RY194144, 3050RY198281, 3050RY200861, 30500X223154, 305000220948, 305000223622, 30500X225385, 30500X233417, 305000231714, 305000241350, 30500X243874, 305000250959, 30500X254932, 305000263557 Catalog Number 110-4L: 305000221383, 30500X222566, 305000222762, 30500X223971,305000223969, 305000225078, 30500X226539, 305000226221, 30500X227058, 305000226848, 305000226749, 305000226847, 305000226871, 305000226,870, 30500X228783, 305000229450, 305000229909, 305000230871 30500X232775, 305000232199, 30500X232961, 305000233311, 305000234438, 30500X235731, 305000235181, 305000236201, 305000237368, 305000237369, 305000237799, 305000238539, 305000239102, 30500X240245, 30500X240246, 30500X240247, 305000239736, 305000239737, 30500X241365, 30500X241522, 305000241082, 30500X242170, 30500X245765, 30500X246521, 3050RX237799, 3050RX226749, 3050RX229450, 3050RX241082, 3050RX232199, 3050RY252775, 3050RX240446, 3050RY255148, 305000259553, 305000260626, 305000261835, 305000262436, 3050RX259553, 305000263146, 3050RX260626, 305000264077, 3050RX262436, 3050RX261835, 305000264175, 3050RY265351, 3050RY265766, 305000265531, 30500X266144, 3050RY266146, 30500X266522, 305000266062, 305000267297, 30500X267473, 30500X268133, 305000267784, 305000268412, 3050RX265531, 305000268916, 3050RY271046, 305000271872, 30500X272374, 305000272591, 30500X274210, 30500X274963, 305000281820 Catalog Number 110-4BC: 176044, 177018, 178013, 178898, 179757, 180139, 180955, 182279, 182908, 184079, 184687, 185284, 186277, 187293, 190452, 187780, 190451, 191891, 193414, 194074, 195161, 195608, 196987, 197492, 198189, 198512, 200069, 200490, 199609, 201039, 201595, 202423, 203565, 207077, 208532 Catalog Number 110-4HC: 179758, 180956, 183071, 184686, 186278, 187781, 190450, 192246, 194993, 195450, 198191, 199708, 201040, 201596, 202009 Catalog Number 110-4LC: 162523, 167450, 167786, 175720, 177188, 178547, 182130, 187495, 195245, 197493.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0117-2014