Constellation Vision System, REF: 8065751150
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 00380657511501, Serial Numbers: 1403028701X, 1202865201X, 0802983301X
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Alcon Research, LLC
- Reason for Recall:
- Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of cutters, scissors, and forceps, which could lead to an increased risk of microbial infection, tissue damage and progressive visual impairment.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Constellation Vision System, REF: 8065751150
Product Codes/Lot Numbers:
UDI-DI: 00380657511501, Serial Numbers: 1403028701X, 1202865201X, 0802983301X
Distribution:
Distributed in: US, GA, IL, PR
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0099-2024
Related Recalls
XXX
Alcon PRECISION1 One-Day Contact Lenses, -2.25D.
Alcon Research
Class I - Dangerous
Potential for a weak seal in some units resulting in compromise in sterility.
Due to incomplete seals in the pouch which provide the sterile barrier.