ACCOLADE DR SL MRI (Model L311)
Class I - DangerousWhat Should You Do?
- Check if you have this product: GTIN 00802526558962, Serial Numbers: 200737, 202944, 203132, 203273, 203512, 203621, 203903, 204448, 204712, 204767, 204877, 204938, 205006, 205007, 205020, 205029, 205055, 205063, 205068, 205069, 205076, 205077, 205079, 205088, 205116, 205218, 205226, 205261, 205328, 205330, 205339, 205354, 205355, 205375, 205398, 205400, 205423, 205457, 205515, 205544, 205552, 205601, 205613, 205633, 205642, 205645, 205651, 205656, 205668, 205694, 205699, 205705, 205713, 205716, 205717, 205719, 205728, 205772, 205778, 205808, 205816, 205818, 205830, 205833, 205835, 205838, 205870, 205896, 205956, 205983, 205995, 206085, 206094, 206117, 206123, 206144, 206152, 206157, 206252, 206260, 206265, 206285, 206325, 206375, 206404, 206408, 206411, 206505, 207204, 207462, 207887, 208016, 208046, 208192, 208212, 208325, 208368, 208372, 208379, 208434, 208449, 208461, 208465, 208471, 208489, 208632, 208636, 208651, 208658, 208690, 208788, 208952, 209027, 209089, 209115, 209132, 209185, 209198, 209201, 209214, 209215, 209221, 209230, 209231, 209253, 209338, 209365, 209443, 209820, 209837, 209851, 209873, 209888, 209893, 209918, 209924, 209931, 209990, 210009, 210010, 210015, 210059, 210092, 210107, 210120, 210150, 210157, 210160, 210182, 210202, 210255, 210300, 210345, 210348, 210354, 210434, 210469, 210513, 210517, 210846, 214405, 214877, 216563, 217673, 217871, 217939, 218076, 219080, 219100, 219169, 219194, 219243, 219266, 219308, 219327, 219338, 219375, 219384, 219451, 219457, 219479, 219523, 219689, 219710, 219712, 219720, 219725, 219732, 219759, 219768, 219769, 219775, 219805, 219806, 219809, 219811, 219825, 219840, 219853, 219857, 219867, 219878, 219880, 219883, 219887, 219909, 219911, 219916, 219926, 219933, 219938, 219952, 219963, 219975, 219981, 219984, 219986, 219987, 219992, 219995, 219998, 220003, 220005, 220020, 220022, 220025, 220034, 220035, 220036, 220044, 220052, 220064, 220067, 220079, 220080, 220081, 220118, 220162, 220175, 220177, 220185, 220224, 220343, 220347, 220349, 220350, 220368, 220375, 220379, 220381, 220393, 220394, 220396, 220398, 220402, 220410, 220412, 220656, 220757, 220773, 220864, 221482, 221488, 221501, 221506, 221518, 221521, 221565, 221576, 221601, 221617, 221669, 221696, 221710, 221715, 221792, 221800, 221804, 221832, 221935, 222102, 222126, 222148, 222166, 222254, 222278, 222298, 222386, 222389, 222392, 222412, 222463, 222486, 222534, 222706, 222718, 222772, 222773, 222788, 222807, 222823, 222849, 222856, 222860, 222872, 222875, 222876, 222877, 222896, 222899, 222905, 222911, 222945, 222958, 222979, 223011, 223013, 223020, 223032, 223044, 223051, 223053, 223064, 223065, 223082, 223083, 223106, 223111, 223119, 223124, 223127, 223135, 223140, 223150, 223155, 223179, 223182, 223186, 223189, 223190, 223202, 223206, 223213, 223217, 223238, 223241, 223255, 223286, 223287, 223321, 223322, 223330, 227095, 227107, 227108, 227118, 227123, 227124, 227131, 227136, 227154, 227161, 227183, 227187, 227195, 227210, 227220, 227251, 227270, 227318, 227338, 227339, 227354, 227359, 227364, 227371, 227386, 227387, 227388, 227389, 227390, 227400, 227413, 227423, 227435, 227436, 227439, 227462, 227528, 227534, 227553, 227564, 227565, 228224, 228239, 228242, 228254, 228321, 228341, 228343, 228390, 228391, 228396, 228473, 228520, 228562, 228905, 228908, 228916, 228941, 228942, 228957, 228962, 228965, 230239, 230267, 231233, 231235, 231237, 231238, 231239, 231244, 231245, 231254, 231257, 231265, 231266, 231271, 231273, 231279, 231282, 231285, 231287, 231301, 231307, 231308, 231324, 231331, 231333, 231335, 231339, 231340, 231342, 231355, 231357, 231359, 231360, 231363, 231367, 231370, 231371, 231375, 231377, 231384, 231393, 231396, 231398, 231401, 231403, 231407, 231409, 231411, 231413, 231414, 231415, 231420, 231421, 232251, 232253, 232257, 232284, 232289, 232296, 232317, 232324, 232351, 232361, 232362, 232374, 232377, 232378, 232390, 232391, 232394, 232402, 232408, 232409, 232585, 232591, 232600, 232612, 232646, 232655, 232656, 232657, 232667, 232711, 232726, 232734, 232739, 232742, 232743, 232746, 232747, 232748, 232753, 232755, 232762, 232765, 232928, 232933, 232941, 232947, 232950, 232956, 232968, 232986, 232994, 232997, 233015, 233046, 233075, 233079, 233083, 233087, 233093, 233096, 233107, 233127, 233136, 233138, 233139, 233666, 233669, 233670, 233673, 233687, 233690, 233691, 233694, 233701, 233704, 233717, 233721, 233727, 233729, 233730, 233733, 233737, 233738, 233751, 233754, 233757, 233766, 233786, 233812, 233813, 233814, 233815, 233820, 233829, 233835, 233836, 233837, 233843, 233848, 233849, 233853, 233855, 233856, 233868, 233870, 233876, 233878, 233887, 233891, 233893, 233897, 233933, 235008, 235051, 235060, 235066, 235069, 235076, 235093, 235110, 235112, 235127, 235132, 235140, 235160, 235163, 235164, 235167, 235172, 235176, 235185, 235186, 235195, 235218, 235222, 235238, 235250, 235256, 235270, 235273, 235280, 235285, 235288, 235294, 235307, 235308, 235317, 235320, 235339, 235341, 235343, 235352, 235360, 235387, 235405, 235406, 235410, 235412, 235419, 235436, 235454, 235475, 235484, 235499, 236097, 236101, 236104, 236106, 236116, 236117, 236130, 236131, 236132, 236144, 236146, 236149, 236170, 236180, 236192, 236197, 236200, 236202, 236210, 236212, 236218, 236225, 236232, 236234, 236238, 236241, 236245, 236246, 236381, 237128, 237141, 237145, 237163, 237180, 237189, 237224, 237236, 237237, 237238, 237239, 237241, 237252, 237275, 237276, 237278, 237281, 237283, 237285, 237291, 237292, 237293, 237306, 237322, 237325, 237333, 237334, 237338, 237345, 237383, 237398, 237407, 237409, 237419, 237439, 237440, 237441, 237447, 237448, 237464, 237466, 237478, 237483, 237485, 237538, 237543, 237550, 237552, 238029, 238032, 238033, 238046, 238055, 238058, 238083, 238097, 238102, 238115, 238134, 238137, 238194, 238259, 238420, 238428, 238970, 239012, 239017, 239018, 240069, 240094, 240219, 240227, 240239, 240266, 240278, 240282, 240340, 240379, 240380, 240397, 240398, 240405, 240421, 240461, 240466, 240504, 240506, 240561, 240668, 240682, 240697, 240784, 240863, 240864, 240872, 240904, 240911, 240923, 240932, 241171, 241293, 241423, 241448, 241456, 241587, 241654, 241666, 241675, 241824, 242076, 242102, 242293, 242338, 242373, 242412, 242578, 242580, 242648, 242708, 242713, 242716, 242718, 242735, 242740, 242747, 242754, 242766, 242774, 242799, 242813, 242817, 242836, 242838, 242843, 242844, 242850, 242863, 242872, 242873, 242901, 242902, 242992, 243002, 243022, 243039, 243041, 243045, 243054, 243055, 243061, 243063, 243064, 243068, 243088, 243096, 243100, 243116, 243117, 243118, 243121, 243125, 243126, 243129, 243134, 243135, 243136, 243149, 243151, 243152, 243153, 243156, 243157, 243163, 243164, 243171, 243188, 243189, 243191, 243195, 243274, 243361, 243364, 243372, 243429, 243430, 243450, 243483, 243491, 243509, 243514, 243527, 243531, 243535, 243536, 243579, 243585, 243592, 243595, 243596, 243605, 243623, 243625, 243643, 243645, 243646, 243647, 243649, 243650, 243651, 243655, 243664, 243666, 243670, 243671, 243672, 243673, 243678, 243685, 243694, 243698, 243700, 243706, 243708, 243719, 243722, 243741, 243742, 243746, 243762, 243770, 243775, 243776, 243780, 243798, 243812, 243813, 243830, 243836, 243869, 243897, 243902, 243915, 243916, 243922, 243926, 243947, 243949, 243952, 244002, 244026, 244035, 244058, 244064, 244087, 244090, 244132, 244154, 244164, 244180, 244184, 244191, 244238, 244265, 244273, 244285, 244333, 244343, 244352, 244383, 244404, 244437, 244445, 244465, 244502, 244558, 244572, 244573, 244582, 244584, 244594, 244599, 244604, 244649, 244652, 244653, 244660, 244711, 244714, 244716, 244732, 244734, 244736, 244740, 244753, 244765, 244767, 244788, 244837, 244845, 244855, 244931, 245025, 245070, 245082, 245083, 245086, 245123, 245350, 245374, 245627, 245628, 245629, 245630, 245635, 245636, 245637, 245638, 245641, 245642, 245643, 245644, 245648, 245650, 245651, 245655, 245658, 245663, 245674, 245675, 245678, 245679, 245684, 245686, 245689, 245691, 245694, 245695, 245698, 245701, 245705, 245710, 245711, 245714, 245715, 245719, 245725, 245728, 245733, 245742, 245746, 245747, 245751, 245752, 245759, 245763, 245776, 245780, 245785, 245808, 245876, 245960, 245962, 245964, 246033, 246049, 246076, 246078, 246086, 246090, 246096, 246123, 246146, 246150, 246162, 246167, 246222, 246229, 246238, 246274, 246276, 246295, 246313, 246316, 246331, 246340, 246349, 246352, 246363, 246386, 246390, 246393, 246399, 246413, 246425, 246446, 246458, 246466, 246490, 246497, 246499, 246504, 246507, 246508, 246519, 246523, 246542, 246544, 246547, 246555, 246557, 246583, 246587, 246600, 246646, 246673, 246682, 246683, 246684, 246691, 246707, 246719, 246723, 246729, 246732, 246735, 246742, 246751, 246757, 246763, 246764, 246765, 246781, 246784, 246792, 246793, 246796, 246803, 246809, 246810, 246824, 246831, 246835, 246847, 246848, 246862, 246872, 246878, 246882, 246884, 246891, 246900, 246904, 246905, 246912, 246921, 246925, 246935, 246950, 246963, 246965, 246973, 246993, 246997, 247001, 247003, 247007, 247015, 247017, 247031, 247032, 247036, 247037, 247047, 247051, 247067, 247094, 247107, 247113, 247120, 247124, 247134, 247176, 247193, 247199, 247205, 247209, 247215, 247226, 247229, 247238, 247254, 247268, 247290, 247329, 247332, 247345, 247378, 247384, 247386, 247393, 247398, 247408, 247422, 247432, 247434, 247440, 247447, 247467, 247468, 247469, 247480, 247484, 247492, 247495, 247499, 247507, 247511, 247512, 247513, 247522, 247523, 247524, 247531, 247537, 247548, 247551, 247553, 247556, 247558, 247559, 247561, 247566, 247574, 247578, 247583, 247594, 247596, 247603, 247615, 247616, 247637, 247638, 247641, 247644, 247645, 247663, 247665, 247674, 247676, 247685, 247691, 247693, 247701, 247845, 247848, 247868, 247875, 247898, 247900, 247910, 247919, 247930, 247931, 247944, 247947, 247962, 247972, 247996, 248000, 248009, 248011, 248027, 248028, 248030, 248102, 248105, 248123, 248127, 248153, 248155, 248190, 248197, 248224, 248805, 248817, 248874, 248970, 248990, 248991, 249038, 249053, 249090, 249134, 249139, 249143, 249146, 249158, 249160, 249161, 249166, 249170, 249198, 249207, 249229, 249274, 249283, 249352, 249397, 249403, 249456, 249558, 249578, 249661, 249703, 249736, 249763, 249764, 249766, 249778, 249806, 249834, 249840, 249843, 249864, 249876, 249890, 249891, 249897, 249902, 249905, 249911, 249955, 249956, 249965, 249988, 250024, 250026, 250122, 250189, 250212, 250242, 250300, 250400, 250421, 250477, 250479, 250542, 250545, 250551, 250552, 250561, 250598, 250604, 250605, 250612, 250637, 250650, 250669, 250697, 250715, 250760, 250767, 250774, 250820, 251052, 251132, 251167, 251183, 251188, 251240, 251243, 251245, 251256, 251258, 251265, 251269, 251278, 251283, 251290, 251295, 251304, 251339, 251383, 251397, 251408, 251434, 251451, 251469, 251514, 251517, 251535, 251552, 251618, 251626, 251654, 251663, 251670, 251681, 251682, 251717, 251719, 251723, 251729, 251737, 251773, 251779, 251783, 251791, 251809, 251828, 251884, 251905, 251909, 251914, 251980, 252041, 252107, 252110, 252132, 252140, 252170, 252185, 252214, 252237, 252247, 252257, 252282, 252290, 252297, 252310, 252313, 252329, 252353, 252359, 252366, 252390, 252394, 252409, 252419, 252460, 252476, 252533, 252550, 252555, 252569, 252596, 252603, 252607, 252632, 252661, 252663, 252670, 252672, 252674, 252691, 252700, 252705, 252727, 252738, 252750, 252754, 252756, 252767, 252814, 252818, 252819, 253255, 253268, 253329, 253638, 253678, 253719, 253730, 253843, 253883, 253894, 253913, 253933, 253951, 253955, 253957, 253965, 254010, 254024, 254025, 254028, 254038, 254046, 254048, 254052, 254059, 254061, 254065, 254066, 254071, 254075, 254076, 254085, 254101, 254110, 254114, 254117, 254122, 254126, 254134, 254142, 254145, 254148, 254149, 254150, 254152, 254159, 254163, 254165, 254166, 254179, 254184, 254188, 254195, 254197, 254208, 254209, 254217, 254250, 254253, 254267, 254277, 254280, 254283, 254285, 254302, 254312, 254314, 254318, 254325, 254333, 254336, 254339, 254365, 254382, 254505, 254538, 254599, 254625, 254627, 254649, 254657, 254664, 254739, 254807, 254850, 254865, 254964, 255032, 255036, 255048, 255084, 255100, 255142, 255143, 255172, 255190, 255195, 255211, 255247, 255278, 255489, 255506, 255520, 255525, 255527, 255607, 255622, 255634, 255722, 255756, 255764, 255781, 255782, 255791, 255814, 255844, 255850, 255902, 255931, 255968, 255971, 255988, 255989, 255995, 255998, 256000, 256019, 256020, 256071, 256074, 256117, 256214, 256240, 256305, 256334, 256346, 256358, 256437, 256551, 256571, 256624, 256625, 256636, 256643, 256648, 256651, 256659, 256670, 256671, 256679, 256684, 256690, 256696, 256699, 256705, 256717, 256721, 256728, 256736, 256741, 256745, 256750, 256754, 256765, 256787, 256788, 256800, 256819, 256823, 256827, 256829, 256833, 256834, 256836, 256838, 256843, 256853, 256854, 256855, 256858, 256866, 256867, 256879, 256890, 256894, 256896, 256904, 256925, 256926, 256931, 256937, 256945, 256950, 256972, 256984, 256990, 257004, 257013, 257027, 257040, 257050, 257052, 257056, 257059, 257063, 257074, 257081, 257087, 257104, 257131, 257141, 257144, 257154, 257156, 257161, 257167, 257176, 257178, 257179, 257228, 257415, 257488, 257550, 257567, 257583, 257584, 257609, 257633, 257643, 257678, 257682, 257687, 257697, 257699, 257700, 257728, 257737, 257743, 257761, 257772, 257779, 257794, 257802, 257814, 257838, 257843, 257879, 257922, 257940, 257942, 257952, 257974, 257981, 258015, 258037, 258095, 258200, 258227, 258231, 258238, 258285, 258317, 258321, 258325, 258330, 258378, 258395, 258416, 258428, 258452, 258453, 258459, 258467, 258482, 258493, 258494, 258520, 258622, 258649, 258801, 258813, 258864, 258930, 258932, 258941, 258949, 258964, 258967, 258975, 258978, 258983, 258987, 259000, 259046, 259057, 259070, 259091, 259092, 259124, 259145, 259168, 259170, 259171, 259173, 259197, 259238, 259247, 259265, 259306, 259310, 259367, 259374, 259377, 259379, 259385, 259389, 259405, 259413, 259417, 259433, 259444, 259450, 259479, 259521, 259524, 259525, 259527, 259528, 259537, 259542, 259543, 259548, 259549, 259560, 259574, 259591, 259612, 259613, 259632, 259633, 259641, 259645, 259659, 259660, 259666, 259687, 259746, 259784, 259878, 259888, 259914, 259919, 259969, 259984, 260024, 260186, 260190, 260203, 260212, 260327, 260573, 260689, 260855, 260869, 260914, 260924, 260959, 260966, 260992, 261007, 261019, 261021, 261039, 261044, 261045, 261050, 261077, 261078, 261079, 261080, 261107, 261111, 261147, 261205, 261211, 261225, 261226, 261255, 261302, 261342, 261371, 261394, 261422, 261507, 261536, 261540, 261551, 261585, 261606, 261653, 261664, 261668, 261680, 261711, 261712, 261720, 261763, 261784, 261813, 261824, 261831, 261832, 261848, 261857, 261863, 261864, 261881, 261895, 261909, 261914, 261915, 261917, 261944, 261949, 261951, 261958, 261961, 262013, 262071, 262084, 262110, 262116, 262138, 262142, 262146, 262164, 262182, 262187, 262188, 262222, 262226, 262236, 262244, 262245, 262246, 262249, 262260, 262266, 262273, 262274, 262279, 262305, 262308, 262326, 262329, 262346, 262365, 262376, 262390, 262418, 262426, 262438, 262439, 262440, 262444, 262446, 262448, 262450, 262456, 262466, 262469, 262471, 262492, 262494, 262500, 262502, 262507, 262518, 262519, 262529, 262534, 262541, 262545, 262558, 262572, 262576, 262578, 262585, 262592, 262593, 262597, 262611, 262612, 262623, 262626, 262629, 262643, 262650, 262660, 262666, 262672, 262673, 262674, 262679, 262680, 262688, 262694, 262708, 262726, 262731, 262734, 262737, 262740, 262741, 262745, 262754, 262755, 262771, 262780, 262786, 262791, 262794, 262801, 262808, 262811, 262822, 262850, 262852, 262890, 262891, 262914, 262920, 262921, 262928, 262935, 262937, 262945, 262951, 262958, 262970, 262987, 263004, 263018, 263021, 263022, 263040, 263042, 263049, 263065, 263073, 263074, 263085, 263088, 263093, 263094, 263095, 263116, 263118, 263128, 263132, 263136, 263142, 263150, 263157, 263166, 263172, 263187, 263190, 263193, 263201, 263202, 263204, 263211, 263216, 263229, 263231, 263233, 263238, 263250, 263252, 263266, 263291, 263302, 263303, 263312, 263313, 263328, 263332, 263349, 263355, 263356, 263367, 263383, 263409, 263422, 263423, 263425, 263431, 263433, 263440, 263450, 263452, 263453, 263456, 263481, 263489, 263504, 263513, 263518, 263541, 263542, 263578, 263579, 263605, 263626, 263628, 263638, 263655, 263676, 263684, 263714, 263731, 263734, 263766, 263790, 263817, 263829, 263832, 263860, 263875, 263881, 263889, 263890, 263915, 263919, 263920, 263928, 264055, 264060, 264076, 264101, 264103, 264112, 264133, 264161, 264191, 264200, 264213, 264215, 264222, 264226, 264231, 264258, 264266, 264269, 264286, 264292, 264293, 264309, 264325, 264331, 264337, 264350, 264368, 264380, 264387, 264388, 264403, 264406, 264407, 264414, 264418, 264419, 264424, 264426, 264427, 264449, 264450, 264451, 264452, 264495, 264498, 264501, 264541, 264543, 264550, 264551, 264554, 264567, 264581, 264583, 264585, 264586, 264593, 264595, 264598, 264668, 264679, 264688, 264694, 264721, 264725, 264733, 264748, 264766, 264773, 264818, 264824, 264842, 264848, 264897, 264936, 265614, 265641, 265652, 266438, 267244, 267342, 267357, 267406, 267409, 267410, 267429, 267488, 267508, 267513, 267517, 267566, 267581, 267615, 267616, 267664, 267842, 267908, 267957, 267958, 267971, 267982, 267984, 267994, 267999, 268029, 268031, 268051, 268071, 268072, 268104, 268117, 268144, 268145, 268170, 268188, 268202, 268210, 268212, 268240, 268246, 268247, 268251, 268267, 268270, 268275, 268281, 268300, 268310, 268332, 268371, 268373, 268379, 268391, 268400, 268407, 268409, 268444, 268453, 268454, 268457, 268459, 268487, 268489, 268529, 268532, 268533, 268535, 268537, 268538, 268541, 268556, 268583, 268590, 268602, 268633, 268666, 268694, 268739, 268743, 268789, 268822, 268823, 268824, 268844, 268858, 268861, 268876, 268880, 268883, 268898, 268902, 268918, 268919, 268956, 268967, 268970, 268978, 268986, 268991, 269010, 269060, 269100, 269114, 269115, 269152, 269154, 269166, 269224, 269227, 269284, 269323, 269338, 269346, 269352, 269359, 269388, 269390, 269408, 269409, 269428, 269438, 269440, 269445, 269455, 269458, 269465, 269482, 269484, 269490, 269496, 269502, 269509, 269517, 269522, 269535, 269541, 269545, 269572, 269576, 269582, 269646, 269649, 269668, 269669, 269680, 269683, 269693, 269694, 269697, 269703, 269707, 269719, 269731, 269747, 269749, 269754, 269755, 269758, 269760, 269774, 269780, 269782, 269787, 269789, 269794, 269832, 269834, 269835, 269836, 269837, 269838, 269840, 269841, 269842, 269843, 269847, 269858, 269866, 269871, 269872, 269874, 269878, 269881, 269889, 269934, 269970, 269984, 270004, 270025, 270032, 270038, 270052, 270061, 270102, 270103, 270105, 270139, 270143, 270147, 270155, 270162, 270185, 270212, 270213, 270229, 270236, 270239, 270244, 270245, 270246, 270247, 270249, 270250, 270262, 270263, 270267, 270273, 270275, 270293, 270294, 270297, 270298, 270310, 270311, 270366, 270371, 270383, 270390, 270404, 270412, 270415, 270417, 270420, 270422, 270434, 270465, 270466, 270467, 270470, 270471, 270485, 270493, 270510, 270523, 270530, 270532, 270533, 270537, 270541, 270573, 270582, 270586, 270613, 270624, 270635, 270643, 270700, 270717, 270740, 270779, 270824, 270825, 270827, 270840, 270841, 270847, 270983, 271007, 271036, 271041, 271096, 271125, 271137, 271140, 271152, 271153, 271155, 271160, 271168, 271173, 271174, 271175, 271209, 271211, 271212, 271248, 271261, 271266, 271267, 271278, 271282, 271283, 271286, 271296, 271299, 271300, 271301, 271306, 271307, 271308, 271328, 271332, 271336, 271344, 271349, 271350, 271352, 271355, 271358, 271359, 271368, 271370, 271376, 271378, 271381, 271382, 271384, 271386, 271396, 271401, 271403, 271409, 271415, 271425, 271434, 271449, 271452, 271456, 271457, 271464, 271475, 271482, 271498, 271501, 271503, 271510, 271517, 271519, 271520, 271541, 271544, 271560, 271562, 271570, 271585, 271591, 271617, 271624, 271626, 271632, 271641, 271655, 271660, 271665, 271669, 271675, 271679, 271687, 271691, 271693, 271694, 271703, 271706, 271708, 271710, 271711, 271712, 271731, 271743, 271750, 271757, 271762, 271768, 271771, 271772, 271784, 271791, 271795, 271827, 271831, 271839, 271843, 271848, 271859, 271869, 271871, 271875, 271886, 271888, 271891, 271895, 271901, 271902, 271903, 271915, 271942, 271946, 271949, 271962, 271972, 271975, 271981, 271987, 271990, 271998, 272004, 272005, 272027, 272039, 272051, 272066, 272083, 272086, 272094, 272106, 272109, 272111, 272131, 272146, 272147, 272148, 272170, 272180, 272194, 272198, 272204, 272209, 272220, 272224, 272247, 272287, 272308, 272314, 272365, 272390, 272424, 272431, 272438, 272440, 272443, 272455, 272469, 272477, 272492, 272497, 272500, 272584, 272586, 272593, 272602, 272634, 272658, 272674, 272679, 272682, 272709, 272712, 272734, 272753, 272757, 272765, 272815, 272828, 272839, 272848, 272878, 272892, 272898, 272951, 272955, 272956, 273028, 273052, 273081, 273097, 273105, 273108, 273119, 273123, 273134, 273145, 273172, 273177, 273180, 273210, 273218, 273226, 273231, 273232, 273237, 273241, 273244, 273248, 273249, 273253, 273264, 273267, 273277, 273281, 273282, 273289, 273294, 273310, 273315, 273324, 273335, 273337, 273340, 273358, 273380, 273395, 273404, 273415, 273417, 273423, 273427, 273445, 273476, 273485, 273488, 273494, 273497, 273514, 273521, 273522, 273523, 273547, 273548, 273614, 273693, 273729, 273735, 273771, 273827, 273847, 273851, 273858, 273868, 273882, 273902, 273938, 273960, 273964, 273996, 274023, 274038, 274046, 274056, 274078, 274099, 274101, 274142, 274183, 274230, 274247, 274277, 274285, 274288, 274324, 274334, 274339, 274345, 274388, 274411, 274415, 274441, 274456, 274499, 274549, 274617, 274619, 274656, 274657, 274666, 274683, 274706, 274718, 274719, 274721, 274733, 274741, 274745, 274746, 274751, 274763, 274768, 274771, 274780, 274783, 274789, 274795, 274796, 274805, 274806, 274826, 274828, 274837, 274839, 274852, 274854, 274867, 274877, 274886, 274892, 274900, 274901, 274939, 274948, 274949, 274964, 274966, 274968, 274975, 274976, 275002, 275008, 275023, 275031, 275037, 275039, 275046, 275047, 275053, 275054, 275070, 275082, 275088, 275091, 275119, 275123, 275128, 275142, 275150, 275156, 275161, 275169, 275171, 275173, 275185, 275189, 275201, 275206, 275213, 275232, 275237, 275238, 275244, 275245, 275257, 275262, 275271, 275293, 275297, 275329, 275352, 275368, 275394, 275395, 275415, 275417, 275427, 275436, 275459, 275471, 275473, 275488, 275492, 275510, 275515, 275544, 275612, 275623, 275637, 275639, 275640, 275641, 275662, 275685, 275717, 275730, 275772, 275780, 275787, 275789, 275790, 275800, 275801, 275825, 275852, 275854, 275857, 275895, 275904, 275922, 275947, 275948, 275971, 275972, 275987, 275991, 275997, 275999, 276019, 276020, 276029, 276033, 276056, 276059, 276115, 276186, 276187, 276212, 276229, 276232, 276279, 276294, 276330, 276610, 276726, 276807, 276906, 276917, 276960, 277235, 277266, 277314, 277376, 277382, 277408, 277469, 277502, 277529, 277532, 277552, 277589, 277609, 277620, 277645, 277662, 277765, 277847, 277879, 277903, 277933, 277966, 277972, 277988, 277999, 278048, 278063, 278075, 278107, 278115, 278121, 278154, 278168, 278173, 278235, 278237, 278243, 278245, 278264, 278267, 278290, 278298, 278321, 278340, 278344, 278377, 278391, 278408, 278415, 278453, 278496, 278542, 278582, 278586, 278590, 278595, 278617, 278625, 278634, 278642, 278643, 278648, 278664, 278665, 278671, 278673, 278676, 278678, 278680, 278681, 278684, 278687, 278690, 278697, 278702, 278705, 278706, 278712, 278715, 278736, 278737, 278739, 278749, 278766, 278770, 278789, 278810, 278820, 278837, 278867, 278876, 278884, 278887, 278888, 278891, 278907, 278908, 278930, 278950, 278956, 279015, 279030, 279039, 279056, 279111, 279116, 279159, 279254, 279284, 279294, 279305, 279340, 279348, 279351, 279376, 279378, 279408, 279437, 279463, 279464, 279507, 279511, 279531, 279551, 279738, 279818, 279846, 279884, 279918, 279972, 280052, 280163, 280291, 280367, 280369, 280387, 280469, 280485, 280520, 280524, 280550, 280620, 280665, 280673, 280717, 280741, 280755, 280773, 280794, 280810, 280852, 280854, 280858, 280867, 280878, 280884, 280906, 280908, 280919, 280924, 280955, 280968, 280974, 280994, 280999, 281013, 281024, 281037, 281059, 281062, 281078, 281094, 281095, 281107, 281129, 281135, 281156, 281173, 281179, 281187, 281192, 281198, 281199, 281210, 281226, 281229, 281231, 281248, 281249, 281250, 281251, 281267, 281273, 281279, 281280, 281285, 281286, 281298, 281306, 281319, 281336, 281343, 281346, 281348, 281351, 281355, 281362, 281363, 281378, 281389, 281398, 281403, 281412, 281413, 281428, 281431, 281433, 281435, 281436, 281444, 281445, 281446, 281457, 281461, 281476, 281491, 281495, 281504, 281536, 281556, 281572, 281581, 281638, 281678, 281682, 281704, 281730, 281894, 281919, 281968, 281970, 281991, 281999, 282060, 282068, 282071, 282080, 282083, 282125, 282128, 282133, 282134, 282139, 282156, 282203, 282214, 282218, 282238, 282242, 282245, 282251, 282256, 282262, 282270, 282275, 282276, 282279, 282282, 282286, 282292, 282316, 282330, 282331, 282337, 282345, 282356, 282358, 282359, 282380, 282381, 282384, 282387, 282393, 282397, 282403, 282405, 282408, 282418, 282430, 282435, 282461, 282464, 282466, 282471, 282478, 282498, 282502, 282503, 282507, 282541, 282589, 282607, 282609, 282615, 282616, 282623, 282653, 282675, 282695, 282697, 282741, 282754, 282765, 282780, 282783, 282795, 282811, 282824, 282828, 282855, 282862, 282873, 282881, 282887, 282889, 282915, 282916, 282942, 282950, 282958, 282965, 282967, 282969, 282970, 282985, 283016, 283019, 283020, 283073, 283075, 283087, 283094, 283095, 283131, 283188, 283189, 283190, 283214, 283239, 283254, 283259, 283266, 283274, 283276, 283278, 283279, 283283, 283325, 283327, 283329, 283346, 283424, 283439, 283490, 283507, 283508, 283608, 283677, 283754, 283756, 283770, 283798, 283806, 283811, 283812, 283819, 283832, 283848, 283858, 283877, 283881, 283899, 283911, 283912, 283918, 283919, 284007, 284046, 284049, 284056, 284062, 284078, 284079, 284085, 284086, 284212, 284222, 284270, 284299, 284310, 284331, 284422, 284424, 284425, 284429, 284466, 284478, 284479, 284484, 284491, 284507, 284516, 284518, 284550, 284552, 284554, 284593, 284627, 284631, 284641, 284661, 284664, 284669, 284676, 284683, 284688, 284689, 284690, 284691, 284692, 284696, 284707, 284713, 284722, 284724, 284725, 284741, 284746, 284790, 284927, 284954, 284957, 285088, 285127, 285176, 285439, 285452, 285469, 285485, 285507, 285511, 285525, 285535, 285540, 285545, 285563, 285581, 285588, 285590, 285635, 285644, 285658, 285659, 285671, 285673, 285674, 285689, 285690, 285691, 285732, 285742, 285744, 285792, 285806, 285888, 285904, 285934, 285991, 286008, 286026, 286125, 286151, 286177, 286222, 286226, 286238, 286242, 286246, 286254, 286277, 286278, 286281, 286297, 286328, 286331, 286376, 286379, 286397, 286419, 286428, 286438, 286446, 286456, 286467, 286475, 286491, 286496, 286507, 286538, 286539, 286588, 286650, 286654, 286665, 286684, 286688, 286699, 286720, 286737, 286760, 286780, 286794, 286820, 286826, 286857, 286872, 286905, 286958, 286960, 286963, 286987, 286992, 287007, 287029, 287036, 287039, 287083, 287138, 287142, 287143, 287155, 287156, 287157, 287160, 287168, 287175, 287176, 287202, 287214, 287218, 287235, 287245, 287246, 287247, 287257, 287265, 287266, 287271, 287274, 287277, 287288, 287289, 287334, 287373, 287392, 287411, 287426, 287440, 287468, 287493, 287526, 287532, 287585, 287604, 287607, 287612, 287614, 287616, 287623, 287632, 287647, 287648, 287676, 287681, 287703, 287711, 287751, 287808, 287822, 287838, 287850, 287860, 287862, 287864, 287884, 287889, 287897, 287903, 287909, 287910, 287917, 287920, 287921, 287927, 287929, 287933, 287944, 287968, 287982, 287996, 288010, 288022, 288031, 288037, 288039, 288045, 288074, 288076, 288087, 288096, 288105, 288106, 288141, 288144, 288146, 288156, 288158, 288161, 288162, 288168, 288180, 288201, 288208, 288212, 288236, 288239, 288242, 288261, 288263, 288264, 288271, 288281, 288282, 288286, 288296, 288298, 288316, 288333, 288360, 288389, 288391, 288430, 288431, 288435, 288439, 288444, 288450, 288462, 288468, 288478, 288488, 288495, 288511, 288512, 288518, 288526, 288529, 288547, 288548, 288550, 288554, 288565, 288572, 288578, 288583, 288594, 288610, 288640, 288647, 288650, 288651, 288652, 288653, 288655, 288661, 288665, 288697, 288740, 288752, 288762, 288811, 288820, 288821, 288826, 288949, 289559, 289634, 289657, 289658, 289717, 289742, 289743, 289750, 289758, 289871, 289882, 289917, 289950, 290017, 290026, 290044, 290068, 290080, 290101, 290111, 290117, 290123, 290137, 290147, 290151, 290170, 290176, 290181, 290197, 290199, 290200, 290205, 290208, 290231, 290301, 290318, 290323, 290324, 290328, 290340, 290347, 290350, 290353, 290357, 290360, 290364, 290419, 290424, 290496, 290537, 290551, 290622, 290637, 290647, 290665, 290819, 290869, 290874, 290900, 290921, 291037, 291042, 291043, 291044, 291051, 291056, 291061, 291065, 291069, 291165, 291243, 291249, 291262, 291267, 291274, 291275, 291281, 291303, 291308, 291312, 291315, 291331, 291337, 291339, 291344, 291354, 291360, 291362, 291364, 291365, 291384, 291386, 291390, 291452, 291455, 291463, 291486, 291513, 291518, 291528, 291534, 291554, 291563, 291597, 291663, 291749, 291826, 291829, 291841, 291842, 291865, 291866, 291893, 291908, 291911, 291912, 291940, 291943, 292001, 292019, 292027, 292030, 292039, 292053, 292068, 292070, 292073, 292076, 292078, 292079, 292084, 292086, 292087, 292090, 292091, 292092, 292094, 292097, 292118, 292152, 292168, 292229, 292258, 292280, 292281, 292292, 292303, 292308, 292309, 292310, 292314, 292315, 292329, 292331, 292352, 292354, 292371, 292377, 292381, 292386, 292387, 292393, 292445, 292474, 292477, 292503, 292519, 292545, 292683, 292694, 292713, 292731, 292734, 292735, 292743, 292760, 292770, 292780, 292781, 292782, 292784, 292789, 292792, 292793, 292796, 292799, 292803, 292804, 292809, 292814, 292838, 292855, 292857, 292866, 292873, 292882, 292890, 292891, 292900, 292910, 292914, 292938, 292941, 292953, 292975, 292976, 292987, 292995, 293033, 293065, 293084, 293127, 293133, 293140, 293192, 293200, 293204, 293205, 293206, 293219, 293227, 293294, 293298, 293382, 293548, 293589, 293642, 293875, 293891, 293898, 293901, 293927, 293957, 293987, 293994, 294001, 294003, 294009, 294011, 294024, 294029, 294033, 294036, 294054, 294055, 294056, 294088, 294091, 294093, 294094, 294095, 294098, 294099, 294101, 294110, 294118, 294135, 294136, 294141, 294144, 294146, 294149, 294150, 294158, 294161, 294163, 294164, 294169, 294182, 294183, 294184, 294187, 294188, 294193, 294194, 294199, 294202, 294205, 294206, 294209, 294214, 294218, 294224, 294232, 294233, 294236, 294240, 294250, 294252, 294254, 294255, 294258, 294263, 294264, 294265, 294267, 294272, 294276, 294286, 294288, 294305, 294308, 294314, 294316, 294335, 294336, 294339, 294341, 294342, 294348, 294349, 294350, 294362, 294363, 294364, 294365, 294370, 294377, 294384, 294388, 294389, 294391, 294397, 294399, 294402, 294403, 294405, 294406, 294408, 294412, 294416, 294417, 294420, 294421, 294433, 294434, 294446, 294451, 294453, 294455, 294456, 294457, 294459, 294460, 294465, 294466, 294471, 294480, 294484, 294487, 294490, 294497, 294498, 294502, 294503, 294507, 294508, 294511, 294513, 294514, 294516, 294518, 294527, 294532, 294533, 294535, 294536, 294537, 294538, 294544, 294549, 294550, 294557, 294560, 294573, 294577, 294579, 294584, 294588, 294590, 294597, 294602, 294604, 294605, 294607, 294610, 294612, 294614, 294619, 294624, 294628, 294631, 294642, 294650, 294651, 294656, 294658, 294659, 294661, 294663, 294664, 294666, 294667, 294669, 294670, 294671, 294672, 294677, 294678, 294679, 294681, 294682, 294686, 294690, 294691, 294693, 294694, 294701, 294702, 294705, 294714, 294715, 294716, 294717, 294722, 294724, 294725, 294729, 294731, 294733, 294735, 294738, 294739, 294740, 294742, 294743, 294744, 294745, 294746, 294749, 294750, 294753, 294755, 294756, 294757, 294758, 294759, 294766, 294767, 294768, 294769, 294772, 294773, 294774, 294775, 294776, 294779, 294789, 294796, 294797, 294800, 294801, 294802, 294803, 294804, 294805, 294806, 294808, 294809, 294813, 294816, 294817, 294819, 294822, 294827, 294832, 294834, 294835, 294839, 294840, 294841, 294842, 294843, 294853, 294879, 294881, 294883, 294885, 294888, 294889, 294890, 294895, 294896, 294898, 294904, 294915, 294951, 294952, 294958, 294959, 294960, 294995, 295058, 295070, 295077, 295082, 295085, 295092, 295102, 29
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Boston Scientific Corporation
- Reason for Recall:
- Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
ACCOLADE DR SL MRI (Model L311)
Product Codes/Lot Numbers:
GTIN 00802526558962, Serial Numbers: 200737, 202944, 203132, 203273, 203512, 203621, 203903, 204448, 204712, 204767, 204877, 204938, 205006, 205007, 205020, 205029, 205055, 205063, 205068, 205069, 205076, 205077, 205079, 205088, 205116, 205218, 205226, 205261, 205328, 205330, 205339, 205354, 205355, 205375, 205398, 205400, 205423, 205457, 205515, 205544, 205552, 205601, 205613, 205633, 205642, 205645, 205651, 205656, 205668, 205694, 205699, 205705, 205713, 205716, 205717, 205719, 205728, 205772, 205778, 205808, 205816, 205818, 205830, 205833, 205835, 205838, 205870, 205896, 205956, 205983, 205995, 206085, 206094, 206117, 206123, 206144, 206152, 206157, 206252, 206260, 206265, 206285, 206325, 206375, 206404, 206408, 206411, 206505, 207204, 207462, 207887, 208016, 208046, 208192, 208212, 208325, 208368, 208372, 208379, 208434, 208449, 208461, 208465, 208471, 208489, 208632, 208636, 208651, 208658, 208690, 208788, 208952, 209027, 209089, 209115, 209132, 209185, 209198, 209201, 209214, 209215, 209221, 209230, 209231, 209253, 209338, 209365, 209443, 209820, 209837, 209851, 209873, 209888, 209893, 209918, 209924, 209931, 209990, 210009, 210010, 210015, 210059, 210092, 210107, 210120, 210150, 210157, 210160, 210182, 210202, 210255, 210300, 210345, 210348, 210354, 210434, 210469, 210513, 210517, 210846, 214405, 214877, 216563, 217673, 217871, 217939, 218076, 219080, 219100, 219169, 219194, 219243, 219266, 219308, 219327, 219338, 219375, 219384, 219451, 219457, 219479, 219523, 219689, 219710, 219712, 219720, 219725, 219732, 219759, 219768, 219769, 219775, 219805, 219806, 219809, 219811, 219825, 219840, 219853, 219857, 219867, 219878, 219880, 219883, 219887, 219909, 219911, 219916, 219926, 219933, 219938, 219952, 219963, 219975, 219981, 219984, 219986, 219987, 219992, 219995, 219998, 220003, 220005, 220020, 220022, 220025, 220034, 220035, 220036, 220044, 220052, 220064, 220067, 220079, 220080, 220081, 220118, 220162, 220175, 220177, 220185, 220224, 220343, 220347, 220349, 220350, 220368, 220375, 220379, 220381, 220393, 220394, 220396, 220398, 220402, 220410, 220412, 220656, 220757, 220773, 220864, 221482, 221488, 221501, 221506, 221518, 221521, 221565, 221576, 221601, 221617, 221669, 221696, 221710, 221715, 221792, 221800, 221804, 221832, 221935, 222102, 222126, 222148, 222166, 222254, 222278, 222298, 222386, 222389, 222392, 222412, 222463, 222486, 222534, 222706, 222718, 222772, 222773, 222788, 222807, 222823, 222849, 222856, 222860, 222872, 222875, 222876, 222877, 222896, 222899, 222905, 222911, 222945, 222958, 222979, 223011, 223013, 223020, 223032, 223044, 223051, 223053, 223064, 223065, 223082, 223083, 223106, 223111, 223119, 223124, 223127, 223135, 223140, 223150, 223155, 223179, 223182, 223186, 223189, 223190, 223202, 223206, 223213, 223217, 223238, 223241, 223255, 223286, 223287, 223321, 223322, 223330, 227095, 227107, 227108, 227118, 227123, 227124, 227131, 227136, 227154, 227161, 227183, 227187, 227195, 227210, 227220, 227251, 227270, 227318, 227338, 227339, 227354, 227359, 227364, 227371, 227386, 227387, 227388, 227389, 227390, 227400, 227413, 227423, 227435, 227436, 227439, 227462, 227528, 227534, 227553, 227564, 227565, 228224, 228239, 228242, 228254, 228321, 228341, 228343, 228390, 228391, 228396, 228473, 228520, 228562, 228905, 228908, 228916, 228941, 228942, 228957, 228962, 228965, 230239, 230267, 231233, 231235, 231237, 231238, 231239, 231244, 231245, 231254, 231257, 231265, 231266, 231271, 231273, 231279, 231282, 231285, 231287, 231301, 231307, 231308, 231324, 231331, 231333, 231335, 231339, 231340, 231342, 231355, 231357, 231359, 231360, 231363, 231367, 231370, 231371, 231375, 231377, 231384, 231393, 231396, 231398, 231401, 231403, 231407, 231409, 231411, 231413, 231414, 231415, 231420, 231421, 232251, 232253, 232257, 232284, 232289, 232296, 232317, 232324, 232351, 232361, 232362, 232374, 232377, 232378, 232390, 232391, 232394, 232402, 232408, 232409, 232585, 232591, 232600, 232612, 232646, 232655, 232656, 232657, 232667, 232711, 232726, 232734, 232739, 232742, 232743, 232746, 232747, 232748, 232753, 232755, 232762, 232765, 232928, 232933, 232941, 232947, 232950, 232956, 232968, 232986, 232994, 232997, 233015, 233046, 233075, 233079, 233083, 233087, 233093, 233096, 233107, 233127, 233136, 233138, 233139, 233666, 233669, 233670, 233673, 233687, 233690, 233691, 233694, 233701, 233704, 233717, 233721, 233727, 233729, 233730, 233733, 233737, 233738, 233751, 233754, 233757, 233766, 233786, 233812, 233813, 233814, 233815, 233820, 233829, 233835, 233836, 233837, 233843, 233848, 233849, 233853, 233855, 233856, 233868, 233870, 233876, 233878, 233887, 233891, 233893, 233897, 233933, 235008, 235051, 235060, 235066, 235069, 235076, 235093, 235110, 235112, 235127, 235132, 235140, 235160, 235163, 235164, 235167, 235172, 235176, 235185, 235186, 235195, 235218, 235222, 235238, 235250, 235256, 235270, 235273, 235280, 235285, 235288, 235294, 235307, 235308, 235317, 235320, 235339, 235341, 235343, 235352, 235360, 235387, 235405, 235406, 235410, 235412, 235419, 235436, 235454, 235475, 235484, 235499, 236097, 236101, 236104, 236106, 236116, 236117, 236130, 236131, 236132, 236144, 236146, 236149, 236170, 236180, 236192, 236197, 236200, 236202, 236210, 236212, 236218, 236225, 236232, 236234, 236238, 236241, 236245, 236246, 236381, 237128, 237141, 237145, 237163, 237180, 237189, 237224, 237236, 237237, 237238, 237239, 237241, 237252, 237275, 237276, 237278, 237281, 237283, 237285, 237291, 237292, 237293, 237306, 237322, 237325, 237333, 237334, 237338, 237345, 237383, 237398, 237407, 237409, 237419, 237439, 237440, 237441, 237447, 237448, 237464, 237466, 237478, 237483, 237485, 237538, 237543, 237550, 237552, 238029, 238032, 238033, 238046, 238055, 238058, 238083, 238097, 238102, 238115, 238134, 238137, 238194, 238259, 238420, 238428, 238970, 239012, 239017, 239018, 240069, 240094, 240219, 240227, 240239, 240266, 240278, 240282, 240340, 240379, 240380, 240397, 240398, 240405, 240421, 240461, 240466, 240504, 240506, 240561, 240668, 240682, 240697, 240784, 240863, 240864, 240872, 240904, 240911, 240923, 240932, 241171, 241293, 241423, 241448, 241456, 241587, 241654, 241666, 241675, 241824, 242076, 242102, 242293, 242338, 242373, 242412, 242578, 242580, 242648, 242708, 242713, 242716, 242718, 242735, 242740, 242747, 242754, 242766, 242774, 242799, 242813, 242817, 242836, 242838, 242843, 242844, 242850, 242863, 242872, 242873, 242901, 242902, 242992, 243002, 243022, 243039, 243041, 243045, 243054, 243055, 243061, 243063, 243064, 243068, 243088, 243096, 243100, 243116, 243117, 243118, 243121, 243125, 243126, 243129, 243134, 243135, 243136, 243149, 243151, 243152, 243153, 243156, 243157, 243163, 243164, 243171, 243188, 243189, 243191, 243195, 243274, 243361, 243364, 243372, 243429, 243430, 243450, 243483, 243491, 243509, 243514, 243527, 243531, 243535, 243536, 243579, 243585, 243592, 243595, 243596, 243605, 243623, 243625, 243643, 243645, 243646, 243647, 243649, 243650, 243651, 243655, 243664, 243666, 243670, 243671, 243672, 243673, 243678, 243685, 243694, 243698, 243700, 243706, 243708, 243719, 243722, 243741, 243742, 243746, 243762, 243770, 243775, 243776, 243780, 243798, 243812, 243813, 243830, 243836, 243869, 243897, 243902, 243915, 243916, 243922, 243926, 243947, 243949, 243952, 244002, 244026, 244035, 244058, 244064, 244087, 244090, 244132, 244154, 244164, 244180, 244184, 244191, 244238, 244265, 244273, 244285, 244333, 244343, 244352, 244383, 244404, 244437, 244445, 244465, 244502, 244558, 244572, 244573, 244582, 244584, 244594, 244599, 244604, 244649, 244652, 244653, 244660, 244711, 244714, 244716, 244732, 244734, 244736, 244740, 244753, 244765, 244767, 244788, 244837, 244845, 244855, 244931, 245025, 245070, 245082, 245083, 245086, 245123, 245350, 245374, 245627, 245628, 245629, 245630, 245635, 245636, 245637, 245638, 245641, 245642, 245643, 245644, 245648, 245650, 245651, 245655, 245658, 245663, 245674, 245675, 245678, 245679, 245684, 245686, 245689, 245691, 245694, 245695, 245698, 245701, 245705, 245710, 245711, 245714, 245715, 245719, 245725, 245728, 245733, 245742, 245746, 245747, 245751, 245752, 245759, 245763, 245776, 245780, 245785, 245808, 245876, 245960, 245962, 245964, 246033, 246049, 246076, 246078, 246086, 246090, 246096, 246123, 246146, 246150, 246162, 246167, 246222, 246229, 246238, 246274, 246276, 246295, 246313, 246316, 246331, 246340, 246349, 246352, 246363, 246386, 246390, 246393, 246399, 246413, 246425, 246446, 246458, 246466, 246490, 246497, 246499, 246504, 246507, 246508, 246519, 246523, 246542, 246544, 246547, 246555, 246557, 246583, 246587, 246600, 246646, 246673, 246682, 246683, 246684, 246691, 246707, 246719, 246723, 246729, 246732, 246735, 246742, 246751, 246757, 246763, 246764, 246765, 246781, 246784, 246792, 246793, 246796, 246803, 246809, 246810, 246824, 246831, 246835, 246847, 246848, 246862, 246872, 246878, 246882, 246884, 246891, 246900, 246904, 246905, 246912, 246921, 246925, 246935, 246950, 246963, 246965, 246973, 246993, 246997, 247001, 247003, 247007, 247015, 247017, 247031, 247032, 247036, 247037, 247047, 247051, 247067, 247094, 247107, 247113, 247120, 247124, 247134, 247176, 247193, 247199, 247205, 247209, 247215, 247226, 247229, 247238, 247254, 247268, 247290, 247329, 247332, 247345, 247378, 247384, 247386, 247393, 247398, 247408, 247422, 247432, 247434, 247440, 247447, 247467, 247468, 247469, 247480, 247484, 247492, 247495, 247499, 247507, 247511, 247512, 247513, 247522, 247523, 247524, 247531, 247537, 247548, 247551, 247553, 247556, 247558, 247559, 247561, 247566, 247574, 247578, 247583, 247594, 247596, 247603, 247615, 247616, 247637, 247638, 247641, 247644, 247645, 247663, 247665, 247674, 247676, 247685, 247691, 247693, 247701, 247845, 247848, 247868, 247875, 247898, 247900, 247910, 247919, 247930, 247931, 247944, 247947, 247962, 247972, 247996, 248000, 248009, 248011, 248027, 248028, 248030, 248102, 248105, 248123, 248127, 248153, 248155, 248190, 248197, 248224, 248805, 248817, 248874, 248970, 248990, 248991, 249038, 249053, 249090, 249134, 249139, 249143, 249146, 249158, 249160, 249161, 249166, 249170, 249198, 249207, 249229, 249274, 249283, 249352, 249397, 249403, 249456, 249558, 249578, 249661, 249703, 249736, 249763, 249764, 249766, 249778, 249806, 249834, 249840, 249843, 249864, 249876, 249890, 249891, 249897, 249902, 249905, 249911, 249955, 249956, 249965, 249988, 250024, 250026, 250122, 250189, 250212, 250242, 250300, 250400, 250421, 250477, 250479, 250542, 250545, 250551, 250552, 250561, 250598, 250604, 250605, 250612, 250637, 250650, 250669, 250697, 250715, 250760, 250767, 250774, 250820, 251052, 251132, 251167, 251183, 251188, 251240, 251243, 251245, 251256, 251258, 251265, 251269, 251278, 251283, 251290, 251295, 251304, 251339, 251383, 251397, 251408, 251434, 251451, 251469, 251514, 251517, 251535, 251552, 251618, 251626, 251654, 251663, 251670, 251681, 251682, 251717, 251719, 251723, 251729, 251737, 251773, 251779, 251783, 251791, 251809, 251828, 251884, 251905, 251909, 251914, 251980, 252041, 252107, 252110, 252132, 252140, 252170, 252185, 252214, 252237, 252247, 252257, 252282, 252290, 252297, 252310, 252313, 252329, 252353, 252359, 252366, 252390, 252394, 252409, 252419, 252460, 252476, 252533, 252550, 252555, 252569, 252596, 252603, 252607, 252632, 252661, 252663, 252670, 252672, 252674, 252691, 252700, 252705, 252727, 252738, 252750, 252754, 252756, 252767, 252814, 252818, 252819, 253255, 253268, 253329, 253638, 253678, 253719, 253730, 253843, 253883, 253894, 253913, 253933, 253951, 253955, 253957, 253965, 254010, 254024, 254025, 254028, 254038, 254046, 254048, 254052, 254059, 254061, 254065, 254066, 254071, 254075, 254076, 254085, 254101, 254110, 254114, 254117, 254122, 254126, 254134, 254142, 254145, 254148, 254149, 254150, 254152, 254159, 254163, 254165, 254166, 254179, 254184, 254188, 254195, 254197, 254208, 254209, 254217, 254250, 254253, 254267, 254277, 254280, 254283, 254285, 254302, 254312, 254314, 254318, 254325, 254333, 254336, 254339, 254365, 254382, 254505, 254538, 254599, 254625, 254627, 254649, 254657, 254664, 254739, 254807, 254850, 254865, 254964, 255032, 255036, 255048, 255084, 255100, 255142, 255143, 255172, 255190, 255195, 255211, 255247, 255278, 255489, 255506, 255520, 255525, 255527, 255607, 255622, 255634, 255722, 255756, 255764, 255781, 255782, 255791, 255814, 255844, 255850, 255902, 255931, 255968, 255971, 255988, 255989, 255995, 255998, 256000, 256019, 256020, 256071, 256074, 256117, 256214, 256240, 256305, 256334, 256346, 256358, 256437, 256551, 256571, 256624, 256625, 256636, 256643, 256648, 256651, 256659, 256670, 256671, 256679, 256684, 256690, 256696, 256699, 256705, 256717, 256721, 256728, 256736, 256741, 256745, 256750, 256754, 256765, 256787, 256788, 256800, 256819, 256823, 256827, 256829, 256833, 256834, 256836, 256838, 256843, 256853, 256854, 256855, 256858, 256866, 256867, 256879, 256890, 256894, 256896, 256904, 256925, 256926, 256931, 256937, 256945, 256950, 256972, 256984, 256990, 257004, 257013, 257027, 257040, 257050, 257052, 257056, 257059, 257063, 257074, 257081, 257087, 257104, 257131, 257141, 257144, 257154, 257156, 257161, 257167, 257176, 257178, 257179, 257228, 257415, 257488, 257550, 257567, 257583, 257584, 257609, 257633, 257643, 257678, 257682, 257687, 257697, 257699, 257700, 257728, 257737, 257743, 257761, 257772, 257779, 257794, 257802, 257814, 257838, 257843, 257879, 257922, 257940, 257942, 257952, 257974, 257981, 258015, 258037, 258095, 258200, 258227, 258231, 258238, 258285, 258317, 258321, 258325, 258330, 258378, 258395, 258416, 258428, 258452, 258453, 258459, 258467, 258482, 258493, 258494, 258520, 258622, 258649, 258801, 258813, 258864, 258930, 258932, 258941, 258949, 258964, 258967, 258975, 258978, 258983, 258987, 259000, 259046, 259057, 259070, 259091, 259092, 259124, 259145, 259168, 259170, 259171, 259173, 259197, 259238, 259247, 259265, 259306, 259310, 259367, 259374, 259377, 259379, 259385, 259389, 259405, 259413, 259417, 259433, 259444, 259450, 259479, 259521, 259524, 259525, 259527, 259528, 259537, 259542, 259543, 259548, 259549, 259560, 259574, 259591, 259612, 259613, 259632, 259633, 259641, 259645, 259659, 259660, 259666, 259687, 259746, 259784, 259878, 259888, 259914, 259919, 259969, 259984, 260024, 260186, 260190, 260203, 260212, 260327, 260573, 260689, 260855, 260869, 260914, 260924, 260959, 260966, 260992, 261007, 261019, 261021, 261039, 261044, 261045, 261050, 261077, 261078, 261079, 261080, 261107, 261111, 261147, 261205, 261211, 261225, 261226, 261255, 261302, 261342, 261371, 261394, 261422, 261507, 261536, 261540, 261551, 261585, 261606, 261653, 261664, 261668, 261680, 261711, 261712, 261720, 261763, 261784, 261813, 261824, 261831, 261832, 261848, 261857, 261863, 261864, 261881, 261895, 261909, 261914, 261915, 261917, 261944, 261949, 261951, 261958, 261961, 262013, 262071, 262084, 262110, 262116, 262138, 262142, 262146, 262164, 262182, 262187, 262188, 262222, 262226, 262236, 262244, 262245, 262246, 262249, 262260, 262266, 262273, 262274, 262279, 262305, 262308, 262326, 262329, 262346, 262365, 262376, 262390, 262418, 262426, 262438, 262439, 262440, 262444, 262446, 262448, 262450, 262456, 262466, 262469, 262471, 262492, 262494, 262500, 262502, 262507, 262518, 262519, 262529, 262534, 262541, 262545, 262558, 262572, 262576, 262578, 262585, 262592, 262593, 262597, 262611, 262612, 262623, 262626, 262629, 262643, 262650, 262660, 262666, 262672, 262673, 262674, 262679, 262680, 262688, 262694, 262708, 262726, 262731, 262734, 262737, 262740, 262741, 262745, 262754, 262755, 262771, 262780, 262786, 262791, 262794, 262801, 262808, 262811, 262822, 262850, 262852, 262890, 262891, 262914, 262920, 262921, 262928, 262935, 262937, 262945, 262951, 262958, 262970, 262987, 263004, 263018, 263021, 263022, 263040, 263042, 263049, 263065, 263073, 263074, 263085, 263088, 263093, 263094, 263095, 263116, 263118, 263128, 263132, 263136, 263142, 263150, 263157, 263166, 263172, 263187, 263190, 263193, 263201, 263202, 263204, 263211, 263216, 263229, 263231, 263233, 263238, 263250, 263252, 263266, 263291, 263302, 263303, 263312, 263313, 263328, 263332, 263349, 263355, 263356, 263367, 263383, 263409, 263422, 263423, 263425, 263431, 263433, 263440, 263450, 263452, 263453, 263456, 263481, 263489, 263504, 263513, 263518, 263541, 263542, 263578, 263579, 263605, 263626, 263628, 263638, 263655, 263676, 263684, 263714, 263731, 263734, 263766, 263790, 263817, 263829, 263832, 263860, 263875, 263881, 263889, 263890, 263915, 263919, 263920, 263928, 264055, 264060, 264076, 264101, 264103, 264112, 264133, 264161, 264191, 264200, 264213, 264215, 264222, 264226, 264231, 264258, 264266, 264269, 264286, 264292, 264293, 264309, 264325, 264331, 264337, 264350, 264368, 264380, 264387, 264388, 264403, 264406, 264407, 264414, 264418, 264419, 264424, 264426, 264427, 264449, 264450, 264451, 264452, 264495, 264498, 264501, 264541, 264543, 264550, 264551, 264554, 264567, 264581, 264583, 264585, 264586, 264593, 264595, 264598, 264668, 264679, 264688, 264694, 264721, 264725, 264733, 264748, 264766, 264773, 264818, 264824, 264842, 264848, 264897, 264936, 265614, 265641, 265652, 266438, 267244, 267342, 267357, 267406, 267409, 267410, 267429, 267488, 267508, 267513, 267517, 267566, 267581, 267615, 267616, 267664, 267842, 267908, 267957, 267958, 267971, 267982, 267984, 267994, 267999, 268029, 268031, 268051, 268071, 268072, 268104, 268117, 268144, 268145, 268170, 268188, 268202, 268210, 268212, 268240, 268246, 268247, 268251, 268267, 268270, 268275, 268281, 268300, 268310, 268332, 268371, 268373, 268379, 268391, 268400, 268407, 268409, 268444, 268453, 268454, 268457, 268459, 268487, 268489, 268529, 268532, 268533, 268535, 268537, 268538, 268541, 268556, 268583, 268590, 268602, 268633, 268666, 268694, 268739, 268743, 268789, 268822, 268823, 268824, 268844, 268858, 268861, 268876, 268880, 268883, 268898, 268902, 268918, 268919, 268956, 268967, 268970, 268978, 268986, 268991, 269010, 269060, 269100, 269114, 269115, 269152, 269154, 269166, 269224, 269227, 269284, 269323, 269338, 269346, 269352, 269359, 269388, 269390, 269408, 269409, 269428, 269438, 269440, 269445, 269455, 269458, 269465, 269482, 269484, 269490, 269496, 269502, 269509, 269517, 269522, 269535, 269541, 269545, 269572, 269576, 269582, 269646, 269649, 269668, 269669, 269680, 269683, 269693, 269694, 269697, 269703, 269707, 269719, 269731, 269747, 269749, 269754, 269755, 269758, 269760, 269774, 269780, 269782, 269787, 269789, 269794, 269832, 269834, 269835, 269836, 269837, 269838, 269840, 269841, 269842, 269843, 269847, 269858, 269866, 269871, 269872, 269874, 269878, 269881, 269889, 269934, 269970, 269984, 270004, 270025, 270032, 270038, 270052, 270061, 270102, 270103, 270105, 270139, 270143, 270147, 270155, 270162, 270185, 270212, 270213, 270229, 270236, 270239, 270244, 270245, 270246, 270247, 270249, 270250, 270262, 270263, 270267, 270273, 270275, 270293, 270294, 270297, 270298, 270310, 270311, 270366, 270371, 270383, 270390, 270404, 270412, 270415, 270417, 270420, 270422, 270434, 270465, 270466, 270467, 270470, 270471, 270485, 270493, 270510, 270523, 270530, 270532, 270533, 270537, 270541, 270573, 270582, 270586, 270613, 270624, 270635, 270643, 270700, 270717, 270740, 270779, 270824, 270825, 270827, 270840, 270841, 270847, 270983, 271007, 271036, 271041, 271096, 271125, 271137, 271140, 271152, 271153, 271155, 271160, 271168, 271173, 271174, 271175, 271209, 271211, 271212, 271248, 271261, 271266, 271267, 271278, 271282, 271283, 271286, 271296, 271299, 271300, 271301, 271306, 271307, 271308, 271328, 271332, 271336, 271344, 271349, 271350, 271352, 271355, 271358, 271359, 271368, 271370, 271376, 271378, 271381, 271382, 271384, 271386, 271396, 271401, 271403, 271409, 271415, 271425, 271434, 271449, 271452, 271456, 271457, 271464, 271475, 271482, 271498, 271501, 271503, 271510, 271517, 271519, 271520, 271541, 271544, 271560, 271562, 271570, 271585, 271591, 271617, 271624, 271626, 271632, 271641, 271655, 271660, 271665, 271669, 271675, 271679, 271687, 271691, 271693, 271694, 271703, 271706, 271708, 271710, 271711, 271712, 271731, 271743, 271750, 271757, 271762, 271768, 271771, 271772, 271784, 271791, 271795, 271827, 271831, 271839, 271843, 271848, 271859, 271869, 271871, 271875, 271886, 271888, 271891, 271895, 271901, 271902, 271903, 271915, 271942, 271946, 271949, 271962, 271972, 271975, 271981, 271987, 271990, 271998, 272004, 272005, 272027, 272039, 272051, 272066, 272083, 272086, 272094, 272106, 272109, 272111, 272131, 272146, 272147, 272148, 272170, 272180, 272194, 272198, 272204, 272209, 272220, 272224, 272247, 272287, 272308, 272314, 272365, 272390, 272424, 272431, 272438, 272440, 272443, 272455, 272469, 272477, 272492, 272497, 272500, 272584, 272586, 272593, 272602, 272634, 272658, 272674, 272679, 272682, 272709, 272712, 272734, 272753, 272757, 272765, 272815, 272828, 272839, 272848, 272878, 272892, 272898, 272951, 272955, 272956, 273028, 273052, 273081, 273097, 273105, 273108, 273119, 273123, 273134, 273145, 273172, 273177, 273180, 273210, 273218, 273226, 273231, 273232, 273237, 273241, 273244, 273248, 273249, 273253, 273264, 273267, 273277, 273281, 273282, 273289, 273294, 273310, 273315, 273324, 273335, 273337, 273340, 273358, 273380, 273395, 273404, 273415, 273417, 273423, 273427, 273445, 273476, 273485, 273488, 273494, 273497, 273514, 273521, 273522, 273523, 273547, 273548, 273614, 273693, 273729, 273735, 273771, 273827, 273847, 273851, 273858, 273868, 273882, 273902, 273938, 273960, 273964, 273996, 274023, 274038, 274046, 274056, 274078, 274099, 274101, 274142, 274183, 274230, 274247, 274277, 274285, 274288, 274324, 274334, 274339, 274345, 274388, 274411, 274415, 274441, 274456, 274499, 274549, 274617, 274619, 274656, 274657, 274666, 274683, 274706, 274718, 274719, 274721, 274733, 274741, 274745, 274746, 274751, 274763, 274768, 274771, 274780, 274783, 274789, 274795, 274796, 274805, 274806, 274826, 274828, 274837, 274839, 274852, 274854, 274867, 274877, 274886, 274892, 274900, 274901, 274939, 274948, 274949, 274964, 274966, 274968, 274975, 274976, 275002, 275008, 275023, 275031, 275037, 275039, 275046, 275047, 275053, 275054, 275070, 275082, 275088, 275091, 275119, 275123, 275128, 275142, 275150, 275156, 275161, 275169, 275171, 275173, 275185, 275189, 275201, 275206, 275213, 275232, 275237, 275238, 275244, 275245, 275257, 275262, 275271, 275293, 275297, 275329, 275352, 275368, 275394, 275395, 275415, 275417, 275427, 275436, 275459, 275471, 275473, 275488, 275492, 275510, 275515, 275544, 275612, 275623, 275637, 275639, 275640, 275641, 275662, 275685, 275717, 275730, 275772, 275780, 275787, 275789, 275790, 275800, 275801, 275825, 275852, 275854, 275857, 275895, 275904, 275922, 275947, 275948, 275971, 275972, 275987, 275991, 275997, 275999, 276019, 276020, 276029, 276033, 276056, 276059, 276115, 276186, 276187, 276212, 276229, 276232, 276279, 276294, 276330, 276610, 276726, 276807, 276906, 276917, 276960, 277235, 277266, 277314, 277376, 277382, 277408, 277469, 277502, 277529, 277532, 277552, 277589, 277609, 277620, 277645, 277662, 277765, 277847, 277879, 277903, 277933, 277966, 277972, 277988, 277999, 278048, 278063, 278075, 278107, 278115, 278121, 278154, 278168, 278173, 278235, 278237, 278243, 278245, 278264, 278267, 278290, 278298, 278321, 278340, 278344, 278377, 278391, 278408, 278415, 278453, 278496, 278542, 278582, 278586, 278590, 278595, 278617, 278625, 278634, 278642, 278643, 278648, 278664, 278665, 278671, 278673, 278676, 278678, 278680, 278681, 278684, 278687, 278690, 278697, 278702, 278705, 278706, 278712, 278715, 278736, 278737, 278739, 278749, 278766, 278770, 278789, 278810, 278820, 278837, 278867, 278876, 278884, 278887, 278888, 278891, 278907, 278908, 278930, 278950, 278956, 279015, 279030, 279039, 279056, 279111, 279116, 279159, 279254, 279284, 279294, 279305, 279340, 279348, 279351, 279376, 279378, 279408, 279437, 279463, 279464, 279507, 279511, 279531, 279551, 279738, 279818, 279846, 279884, 279918, 279972, 280052, 280163, 280291, 280367, 280369, 280387, 280469, 280485, 280520, 280524, 280550, 280620, 280665, 280673, 280717, 280741, 280755, 280773, 280794, 280810, 280852, 280854, 280858, 280867, 280878, 280884, 280906, 280908, 280919, 280924, 280955, 280968, 280974, 280994, 280999, 281013, 281024, 281037, 281059, 281062, 281078, 281094, 281095, 281107, 281129, 281135, 281156, 281173, 281179, 281187, 281192, 281198, 281199, 281210, 281226, 281229, 281231, 281248, 281249, 281250, 281251, 281267, 281273, 281279, 281280, 281285, 281286, 281298, 281306, 281319, 281336, 281343, 281346, 281348, 281351, 281355, 281362, 281363, 281378, 281389, 281398, 281403, 281412, 281413, 281428, 281431, 281433, 281435, 281436, 281444, 281445, 281446, 281457, 281461, 281476, 281491, 281495, 281504, 281536, 281556, 281572, 281581, 281638, 281678, 281682, 281704, 281730, 281894, 281919, 281968, 281970, 281991, 281999, 282060, 282068, 282071, 282080, 282083, 282125, 282128, 282133, 282134, 282139, 282156, 282203, 282214, 282218, 282238, 282242, 282245, 282251, 282256, 282262, 282270, 282275, 282276, 282279, 282282, 282286, 282292, 282316, 282330, 282331, 282337, 282345, 282356, 282358, 282359, 282380, 282381, 282384, 282387, 282393, 282397, 282403, 282405, 282408, 282418, 282430, 282435, 282461, 282464, 282466, 282471, 282478, 282498, 282502, 282503, 282507, 282541, 282589, 282607, 282609, 282615, 282616, 282623, 282653, 282675, 282695, 282697, 282741, 282754, 282765, 282780, 282783, 282795, 282811, 282824, 282828, 282855, 282862, 282873, 282881, 282887, 282889, 282915, 282916, 282942, 282950, 282958, 282965, 282967, 282969, 282970, 282985, 283016, 283019, 283020, 283073, 283075, 283087, 283094, 283095, 283131, 283188, 283189, 283190, 283214, 283239, 283254, 283259, 283266, 283274, 283276, 283278, 283279, 283283, 283325, 283327, 283329, 283346, 283424, 283439, 283490, 283507, 283508, 283608, 283677, 283754, 283756, 283770, 283798, 283806, 283811, 283812, 283819, 283832, 283848, 283858, 283877, 283881, 283899, 283911, 283912, 283918, 283919, 284007, 284046, 284049, 284056, 284062, 284078, 284079, 284085, 284086, 284212, 284222, 284270, 284299, 284310, 284331, 284422, 284424, 284425, 284429, 284466, 284478, 284479, 284484, 284491, 284507, 284516, 284518, 284550, 284552, 284554, 284593, 284627, 284631, 284641, 284661, 284664, 284669, 284676, 284683, 284688, 284689, 284690, 284691, 284692, 284696, 284707, 284713, 284722, 284724, 284725, 284741, 284746, 284790, 284927, 284954, 284957, 285088, 285127, 285176, 285439, 285452, 285469, 285485, 285507, 285511, 285525, 285535, 285540, 285545, 285563, 285581, 285588, 285590, 285635, 285644, 285658, 285659, 285671, 285673, 285674, 285689, 285690, 285691, 285732, 285742, 285744, 285792, 285806, 285888, 285904, 285934, 285991, 286008, 286026, 286125, 286151, 286177, 286222, 286226, 286238, 286242, 286246, 286254, 286277, 286278, 286281, 286297, 286328, 286331, 286376, 286379, 286397, 286419, 286428, 286438, 286446, 286456, 286467, 286475, 286491, 286496, 286507, 286538, 286539, 286588, 286650, 286654, 286665, 286684, 286688, 286699, 286720, 286737, 286760, 286780, 286794, 286820, 286826, 286857, 286872, 286905, 286958, 286960, 286963, 286987, 286992, 287007, 287029, 287036, 287039, 287083, 287138, 287142, 287143, 287155, 287156, 287157, 287160, 287168, 287175, 287176, 287202, 287214, 287218, 287235, 287245, 287246, 287247, 287257, 287265, 287266, 287271, 287274, 287277, 287288, 287289, 287334, 287373, 287392, 287411, 287426, 287440, 287468, 287493, 287526, 287532, 287585, 287604, 287607, 287612, 287614, 287616, 287623, 287632, 287647, 287648, 287676, 287681, 287703, 287711, 287751, 287808, 287822, 287838, 287850, 287860, 287862, 287864, 287884, 287889, 287897, 287903, 287909, 287910, 287917, 287920, 287921, 287927, 287929, 287933, 287944, 287968, 287982, 287996, 288010, 288022, 288031, 288037, 288039, 288045, 288074, 288076, 288087, 288096, 288105, 288106, 288141, 288144, 288146, 288156, 288158, 288161, 288162, 288168, 288180, 288201, 288208, 288212, 288236, 288239, 288242, 288261, 288263, 288264, 288271, 288281, 288282, 288286, 288296, 288298, 288316, 288333, 288360, 288389, 288391, 288430, 288431, 288435, 288439, 288444, 288450, 288462, 288468, 288478, 288488, 288495, 288511, 288512, 288518, 288526, 288529, 288547, 288548, 288550, 288554, 288565, 288572, 288578, 288583, 288594, 288610, 288640, 288647, 288650, 288651, 288652, 288653, 288655, 288661, 288665, 288697, 288740, 288752, 288762, 288811, 288820, 288821, 288826, 288949, 289559, 289634, 289657, 289658, 289717, 289742, 289743, 289750, 289758, 289871, 289882, 289917, 289950, 290017, 290026, 290044, 290068, 290080, 290101, 290111, 290117, 290123, 290137, 290147, 290151, 290170, 290176, 290181, 290197, 290199, 290200, 290205, 290208, 290231, 290301, 290318, 290323, 290324, 290328, 290340, 290347, 290350, 290353, 290357, 290360, 290364, 290419, 290424, 290496, 290537, 290551, 290622, 290637, 290647, 290665, 290819, 290869, 290874, 290900, 290921, 291037, 291042, 291043, 291044, 291051, 291056, 291061, 291065, 291069, 291165, 291243, 291249, 291262, 291267, 291274, 291275, 291281, 291303, 291308, 291312, 291315, 291331, 291337, 291339, 291344, 291354, 291360, 291362, 291364, 291365, 291384, 291386, 291390, 291452, 291455, 291463, 291486, 291513, 291518, 291528, 291534, 291554, 291563, 291597, 291663, 291749, 291826, 291829, 291841, 291842, 291865, 291866, 291893, 291908, 291911, 291912, 291940, 291943, 292001, 292019, 292027, 292030, 292039, 292053, 292068, 292070, 292073, 292076, 292078, 292079, 292084, 292086, 292087, 292090, 292091, 292092, 292094, 292097, 292118, 292152, 292168, 292229, 292258, 292280, 292281, 292292, 292303, 292308, 292309, 292310, 292314, 292315, 292329, 292331, 292352, 292354, 292371, 292377, 292381, 292386, 292387, 292393, 292445, 292474, 292477, 292503, 292519, 292545, 292683, 292694, 292713, 292731, 292734, 292735, 292743, 292760, 292770, 292780, 292781, 292782, 292784, 292789, 292792, 292793, 292796, 292799, 292803, 292804, 292809, 292814, 292838, 292855, 292857, 292866, 292873, 292882, 292890, 292891, 292900, 292910, 292914, 292938, 292941, 292953, 292975, 292976, 292987, 292995, 293033, 293065, 293084, 293127, 293133, 293140, 293192, 293200, 293204, 293205, 293206, 293219, 293227, 293294, 293298, 293382, 293548, 293589, 293642, 293875, 293891, 293898, 293901, 293927, 293957, 293987, 293994, 294001, 294003, 294009, 294011, 294024, 294029, 294033, 294036, 294054, 294055, 294056, 294088, 294091, 294093, 294094, 294095, 294098, 294099, 294101, 294110, 294118, 294135, 294136, 294141, 294144, 294146, 294149, 294150, 294158, 294161, 294163, 294164, 294169, 294182, 294183, 294184, 294187, 294188, 294193, 294194, 294199, 294202, 294205, 294206, 294209, 294214, 294218, 294224, 294232, 294233, 294236, 294240, 294250, 294252, 294254, 294255, 294258, 294263, 294264, 294265, 294267, 294272, 294276, 294286, 294288, 294305, 294308, 294314, 294316, 294335, 294336, 294339, 294341, 294342, 294348, 294349, 294350, 294362, 294363, 294364, 294365, 294370, 294377, 294384, 294388, 294389, 294391, 294397, 294399, 294402, 294403, 294405, 294406, 294408, 294412, 294416, 294417, 294420, 294421, 294433, 294434, 294446, 294451, 294453, 294455, 294456, 294457, 294459, 294460, 294465, 294466, 294471, 294480, 294484, 294487, 294490, 294497, 294498, 294502, 294503, 294507, 294508, 294511, 294513, 294514, 294516, 294518, 294527, 294532, 294533, 294535, 294536, 294537, 294538, 294544, 294549, 294550, 294557, 294560, 294573, 294577, 294579, 294584, 294588, 294590, 294597, 294602, 294604, 294605, 294607, 294610, 294612, 294614, 294619, 294624, 294628, 294631, 294642, 294650, 294651, 294656, 294658, 294659, 294661, 294663, 294664, 294666, 294667, 294669, 294670, 294671, 294672, 294677, 294678, 294679, 294681, 294682, 294686, 294690, 294691, 294693, 294694, 294701, 294702, 294705, 294714, 294715, 294716, 294717, 294722, 294724, 294725, 294729, 294731, 294733, 294735, 294738, 294739, 294740, 294742, 294743, 294744, 294745, 294746, 294749, 294750, 294753, 294755, 294756, 294757, 294758, 294759, 294766, 294767, 294768, 294769, 294772, 294773, 294774, 294775, 294776, 294779, 294789, 294796, 294797, 294800, 294801, 294802, 294803, 294804, 294805, 294806, 294808, 294809, 294813, 294816, 294817, 294819, 294822, 294827, 294832, 294834, 294835, 294839, 294840, 294841, 294842, 294843, 294853, 294879, 294881, 294883, 294885, 294888, 294889, 294890, 294895, 294896, 294898, 294904, 294915, 294951, 294952, 294958, 294959, 294960, 294995, 295058, 295070, 295077, 295082, 295085, 295092, 295102, 29
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0082-2026
Related Recalls
LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5
Boston Scientific
For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia