Zero-Gravity Radiation Protection System Radiology Personnel protective shield.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All serial numbers for all models listed below manufactured/distributed between June 28, 2010 August 25, 2016. Part No. : Description ZGM-6-.5H : Zero-Gravity floor unit ZGM-6-5ROUNDH : Zero-Gravity floor unit ZGM-6Demo : Zero-Gravity demo floor unit ZGCM-48 : Zero-Gravity ceiling mounted monorail with 48 boom arm ZGCM-48-SPECIAL : Zero-Gravity ceiling mounted monorail with 48 boom arm with monorail extension ZG48 : Zero-Gravity 48 boom arm assembly and body shield for monorail ZGCM-66 : Zero-Gravity ceiling mounted monorail with 66 boom arm ZGCM-HSA : Zero-Gravity ceiling mounted monorail with hinged swing arm ZGCH : Zero-Gravity ceiling mounted monorail Contour track with 48 boom arm ZGCH48 : Zero-Gravity 48 boom arm assembly and body shield for Contour track ZGHSA : Zero-Gravity single point, ceiling mount with hinged swing arm ZGCSP : Zero-Gravity single point, ceiling mount with hinged swing arm 29858 : Zero-Gravity floor unit
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Contour Fabricators Inc
- Reason for Recall:
- Two customer complaints suggest that excessive rotation of the cable in one direction (clockwise or counter-clockwise) without allowing the cable to unwind to its normal condition may cause the cable to break suddenly and the body shield to fall. While no injuries or deaths occurred in either failure, the Zero-GravityTM body shield weighs 55 pounds, and in the event of a cable failure, the body shield will fall immediately. A fall of this nature places the user, associated staff and the patient at risk for a potential injury.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Zero-Gravity Radiation Protection System Radiology Personnel protective shield.
Product Codes/Lot Numbers:
All serial numbers for all models listed below manufactured/distributed between June 28, 2010 August 25, 2016. Part No. : Description ZGM-6-.5H : Zero-Gravity floor unit ZGM-6-5ROUNDH : Zero-Gravity floor unit ZGM-6Demo : Zero-Gravity demo floor unit ZGCM-48 : Zero-Gravity ceiling mounted monorail with 48 boom arm ZGCM-48-SPECIAL : Zero-Gravity ceiling mounted monorail with 48 boom arm with monorail extension ZG48 : Zero-Gravity 48 boom arm assembly and body shield for monorail ZGCM-66 : Zero-Gravity ceiling mounted monorail with 66 boom arm ZGCM-HSA : Zero-Gravity ceiling mounted monorail with hinged swing arm ZGCH : Zero-Gravity ceiling mounted monorail Contour track with 48 boom arm ZGCH48 : Zero-Gravity 48 boom arm assembly and body shield for Contour track ZGHSA : Zero-Gravity single point, ceiling mount with hinged swing arm ZGCSP : Zero-Gravity single point, ceiling mount with hinged swing arm 29858 : Zero-Gravity floor unit
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0080-2017