Iso-Gard Filter S, with Expandi-Flex, Tethered Cap, REF: a) 19261T, OUS only; b) 19262T, OUS only; c) 19272T, OUS only; Microbial medical gas filter, single-use,
Class I - DangerousWhat Should You Do?
- Check if you have this product: a) 19261T, UDI: (01)04026704400492(17)250528(10)20FT51, (01)04026704400492(17)251028(10)KMH20L0409, (01)04026704400492(17)251228(10)KMH21A0086, (01)04026704400492(17)251228(10)KMH21A0190, (01)04026704400492(17)260128(10)KMH21B0163, (01)04026704400492(17)260128(10)KMH21B0343, (01)04026704400492(17)260128(10)KMH21B0355, (01)04026704400492(17)260228(10)KMH21C0177, (01)04026704400492(17)260228(10)KMH21C0387, (01)04026704400492(17)260528(10)KMH21F0156, (01)04026704400492(17)260528(10)KMH21F0337, (01)04026704400492(17)260628(10)KMH21G0426, (01)04026704400492(17)260828(10)KMH21J0020, (01)04026704400492(17)260828(10)KMH21J0245, (01)04026704400492(17)260928(10)KMH21K0142, (01)04026704400492(17)261128(10)KMH21M0182, (01)04026704400492(17)270128(10)KMH22B0151, (01)04026704400492(17)270128(10)KMH22B0248; b) 19262T, UDI: (01)04026704400508(17)250528(10)202025, (01)04026704400508(17)250528(10)202026, (01)04026704400508(17)250628(10)202030, (01)04026704400508(17)250828(10)KMZ20J0163, (01)04026704400508(17)250828(10)KMZ20J0171, (01)04026704400508(17)250828(10)KMZ20J0582, (01)04026704400508(17)250828(10)KMZ20J0586, (01)04026704400508(17)250828(10)KMZ20J0587, (01)04026704400508(17)250928(10)KMZ20K0681, (01)04026704400508(17)251028(10)KMZ20L0968, (01)04026704400508(17)251028(10)KMZ20L0977, (01)04026704400508(17)251028(10)KMZ20L1275, (01)04026704400508(17)251128(10)KMZ20M0544, (01)04026704400508(17)251128(10)KMZ20M0661, (01)04026704400508(17)251228(10)KMZ21A0175, (01)04026704400508(17)251228(10)KMZ21A0186, (01)04026704400508(17)251228(10)KMZ21A0187, (01)04026704400508(17)251228(10)KMZ21A0208, (01)04026704400508(17)260128(10)KMZ21B0483, (01)04026704400508(17)260128(10)KMZ21B0485, (01)04026704400508(17)260128(10)KMZ21B0495, (01)04026704400508(17)260128(10)KMZ21B0811, (01)04026704400508(17)260128(10)KMZ21B0918, (01)04026704400508(17)260128(10)KMZ21B1172, (01)04026704400508(17)260228(10)KMZ21C0884, (01)04026704400508(17)260228(10)KMZ21C0910, (01)04026704400508(17)260228(10)KMZ21C0929, (01)04026704400508(17)260228(10)KMZ21C0931, (01)04026704400508(17)260328(10)KMZ21D0419, (01)04026704400508(17)260528(10)KMZ21F0223, (01)04026704400508(17)260528(10)KMZ21F0260, (01)04026704400508(17)260528(10)KMZ21F0528, (01)04026704400508(17)260628(10)KMZ21G0728, (01)04026704400508(17)260628(10)KMZ21G0735, (01)04026704400508(17)260628(10)KMZ21G0739, (01)04026704400508(17)260728(10)KMZ21H0055, (01)04026704400508(17)260828(10)KMZ21J0006, (01)04026704400508(17)260828(10)KMZ21J0007, (01)04026704400508(17)260828(10)KMZ21J0008, (01)04026704400508(17)260928(10)KMZ21K0510, (01)04026704400508(17)260928(10)KMZ21K0524, (01)04026704400508(17)260928(10)KMZ21K0539, (01)04026704400508(17)260928(10)KMZ21K0572, (01)04026704400508(17)260928(10)KMZ21K0578, (01)04026704400508(17)260928(10)KMZ21K0592, (01)04026704400508(17)260928(10)KMZ21K0952, (01)04026704400508(17)261028(10)KMZ21L0218, (01)04026704400508(17)261128(10)KMZ21M0674, (01)04026704400508(17)270128(10)KMZ22B0125, (01)04026704400508(17)270128(10)KMZ22B0345, (01)04026704400508(17)270128(10)KMZ22B0809, (01)04026704400508(17)270128(10)KMZ22B0828, (01)04026704400508(17)270128(10)KMZ22B0838, (01)04026704400508(17)270128(10)KMZ22B0869, (01)04026704400508(17)270228(10)KMZ22C0659; c) 19272T, UDI: (01)04026704400515(17)250528(10)202025, (01)04026704400515(17)250528(10)202026, (01)04026704400515(17)250628(10)202027, (01)04026704400515(17)250628(10)202028, (01)04026704400515(17)250628(10)202029, (01)04026704400515(17)250628(10)202030, (01)04026704400515(17)250628(10)202031, (01)04026704400515(17)250628(10)KMZ20G0005, (01)04026704400515(17)250628(10)KMZ20G0026, (01)04026704400515(17)250628(10)KMZ20G0105, (01)04026704400515(17)250628(10)KMZ20G0112, (01)04026704400515(17)250728(10)KMZ20H0081, (01)04026704400515(17)250728(10)KMZ20H0096, (01)04026704400515(17)250728(10)KMZ20H0158, (01)04026704400515(17)250728(10)KMZ20H0379, (01)04026704400515(17)250728(10)KMZ20H0530, (01)04026704400515(17)250728(10)KMZ20H0540, (01)04026704400515(17)250728(10)KMZ20H0564, (01)04026704400515(17)250728(10)KMZ20H0812, (01)04026704400515(17)250728(10)KMZ20H0841, (01)04026704400515(17)250728(10)KMZ20H1021, (01)04026704400515(17)250728(10)KMZ20H1024, (01)04026704400515(17)250728(10)KMZ20H1031, (01)04026704400515(17)250828(10)KMZ20J0092, (01)04026704400515(17)250828(10)KMZ20J0102, (01)04026704400515(17)250828(10)KMZ20J0227, (01)04026704400515(17)250828(10)KMZ20J0441, (01)04026704400515(17)250828(10)KMZ20J0535, (01)04026704400515(17)250828(10)KMZ20J0551, (01)04026704400515(17)250828(10)KMZ20J0566, (01)04026704400515(17)250828(10)KMZ20J0640, (01)04026704400515(17)250828(10)KMZ20J0654, (01)04026704400515(17)250828(10)KMZ20J0663, (01)04026704400515(17)250828(10)KMZ20J0893, (01)04026704400515(17)250828(10)KMZ20J0958, (01)04026704400515(17)250928(10)KMZ20K0034, (01)04026704400515(17)250928(10)KMZ20K0196, (01)04026704400515(17)250928(10)KMZ20K0236, (01)04026704400515(17)250928(10)KMZ20K0430, (01)04026704400515(17)250928(10)KMZ20K0659, (01)04026704400515(17)250928(10)KMZ20K0662, (01)04026704400515(17)250928(10)KMZ20K0838, (01)04026704400515(17)250928(10)KMZ20K0946, (01)04026704400515(17)250928(10)KMZ20K1260, (01)04026704400515(17)250928(10)KMZ20K1289, (01)04026704400515(17)250928(10)KMZ20K1294, (01)04026704400515(17)250928(10)KMZ20K1310, (01)04026704400515(17)250928(10)KMZ20K1339, (01)04026704400515(17)250928(10)KMZ20K1340, (01)04026704400515(17)251028(10)KMZ20L0003, (01)04026704400515(17)251028(10)KMZ20L0030, (01)04026704400515(17)251028(10)KMZ20L0031, (01)04026704400515(17)251028(10)KMZ20L0117, (01)04026704400515(17)251028(10)KMZ20L0214, (01)04026704400515(17)251028(10)KMZ20L0351, (01)04026704400515(17)251028(10)KMZ20L0359, (01)04026704400515(17)251028(10)KMZ20L0499, (01)04026704400515(17)251028(10)KMZ20L0537, (01)04026704400515(17)251028(10)KMZ20L0706, (01)04026704400515(17)251028(10)KMZ20L0724, (01)04026704400515(17)251028(10)KMZ20L0885, (01)04026704400515(17)251028(10)KMZ20L0886, (01)04026704400515(17)251028(10)KMZ20L0933, (01)04026704400515(17)251028(10)KMZ20L0945, (01)04026704400515(17)251028(10)KMZ20L1056, (01)04026704400515(17)251028(10)KMZ20L1091, (01)04026704400515(17)251028(10)KMZ20L1103, (01)04026704400515(17)251028(10)KMZ20L1104, (01)04026704400515(17)251128(10)KMZ20M0387, (01)04026704400515(17)251128(10)KMZ20M0777, (01)04026704400515(17)251128(10)KMZ20M0778, (01)04026704400515(17)251128(10)KMZ20M0787, (01)04026704400515(17)251128(10)KMZ20M0822, (01)04026704400515(17)251128(10)KMZ20M0842, (01)04026704400515(17)251128(10)KMZ20M0938, (01)04026704400515(17)251128(10)KMZ20M0967, (01)04026704400515(17)251128(10)KMZ20M0990, (01)04026704400515(17)251128(10)KMZ20M1249, (01)04026704400515(17)251128(10)KMZ20M1250, (01)04026704400515(17)251128(10)KMZ20M1257, (01)04026704400515(17)251128(10)KMZ20M1267, (01)04026704400515(17)251228(10)KMZ21A0012, (01)04026704400515(17)251228(10)KMZ21A0024, (01)04026704400515(17)251228(10)KMZ21A0117, (01)04026704400515(17)251228(10)KMZ21A0126, (01)04026704400515(17)251228(10)KMZ21A0155, (01)04026704400515(17)251228(10)KMZ21A0162, (01)04026704400515(17)251228(10)KMZ21A0164, (01)04026704400515(17)251228(10)KMZ21A0203, (01)04026704400515(17)251228(10)KMZ21A0218, (01)04026704400515(17)251228(10)KMZ21A0226, (01)04026704400515(17)251228(10)KMZ21A0368, (01)04026704400515(17)251228(10)KMZ21A0618, (01)04026704400515(17)251228(10)KMZ21A0635, (01)04026704400515(17)251228(10)KMZ21A0693, (01)04026704400515(17)251228(10)KMZ21A0706, (01)04026704400515(17)251228(10)KMZ21A0714, (01)04026704400515(17)251228(10)KMZ21A0725, (01)04026704400515(17)251228(10)KMZ21A0734, (01)04026704400515(17)251228(10)KMZ21A0950, (01)04026704400515(17)260128(10)KMZ21B0101, (01)04026704400515(17)260128(10)KMZ21B0163, (01)04026704400515(17)260128(10)KMZ21B0181, (01)04026704400515(17)260128(10)KMZ21B0183, (01)04026704400515(17)260128(10)KMZ21B0185, (01)04026704400515(17)260128(10)KMZ21B0189, (01)04026704400515(17)260128(10)KMZ21B0227, (01)04026704400515(17)260128(10)KMZ21B0234, (01)04026704400515(17)260128(10)KMZ21B0305, (01)04026704400515(17)260128(10)KMZ21B0320, (01)04026704400515(17)260128(10)KMZ21B0417, (01)04026704400515(17)260128(10)KMZ21B0421, (01)04026704400515(17)260128(10)KMZ21B0435, (01)04026704400515(17)260128(10)KMZ21B0465, (01)04026704400515(17)260128(10)KMZ21B0471, (01)04026704400515(17)260128(10)KMZ21B0479, (01)04026704400515(17)260128(10)KMZ21B0533, (01)04026704400515(17)260128(10)KMZ21B0689, (01)04026704400515(17)260128(10)KMZ21B1171, (01)04026704400515(17)260228(10)KMZ21C0270, (01)04026704400515(17)260228(10)KMZ21C0388, (01)04026704400515(17)260228(10)KMZ21C0398, (01)04026704400515(17)260228(10)KMZ21C0883, (01)04026704400515(17)260228(10)KMZ21C1159, (01)04026704400515(17)260228(10)KMZ21C1160, (01)04026704400515(17)260228(10)KMZ21C1358, (01)04026704400515(17)260328(10)KMZ21D0420, (01)04026704400515(17)260528(10)KMZ21F0227, (01)04026704400515(17)260528(10)KMZ21F0265, (01)04026704400515(17)260528(10)KMZ21F0296, (01)04026704400515(17)260528(10)KMZ21F0335, (01)04026704400515(17)260528(10)KMZ21F0375, (01)04026704400515(17)260528(10)KMZ21F0430, (01)04026704400515(17)260528(10)KMZ21F0467, (01)04026704400515(17)260528(10)KMZ21F0468, (01)04026704400515(17)260528(10)KMZ21F0478, (01)04026704400515(17)260528(10)KMZ21F0501, (01)04026704400515(17)260528(10)KMZ21F0517, (01)04026704400515(17)260528(10)KMZ21F0534, (01)04026704400515(17)260528(10)KMZ21F0550, (01)04026704400515(17)260528(10)KMZ21F0551, (01)04026704400515(17)260528(10)KMZ21F0552, (01)04026704400515(17)260528(10)KMZ21F0557, (01)04026704400515(17)260528(10)KMZ21F0567, (01)04026704400515(17)260528(10)KMZ21F0568, (01)04026704400515(17)260528(10)KMZ21F0588, (01)04026704400515(17)260528(10)KMZ21F0594, (01)04026704400515(17)260528(10)KMZ21F0742, (01)04026704400515(17)260528(10)KMZ21F0777, (01)04026704400515(17)260628(10)KMZ21G0399, (01)04026704400515(17)260628(10)KMZ21G0403, (01)04026704400515(17)260628(10)KMZ21G0404, (01)04026704400515(17)260628(10)KMZ21G0751, (01)04026704400515(17)260628(10)KMZ21G0765, (01)04026704400515(17)260628(10)KMZ21G0770, (01)04026704400515(17)260628(10)KMZ21G0778, (01)04026704400515(17)260628(10)KMZ21G0786, (01)04026704400515(17)260628(10)KMZ21G0800, (01)04026704400515(17)260628(10)KMZ21G0806, (01)04026704400515(17)260628(10)KMZ21G0838, (01)04026704400515(17)260628(10)KMZ21G0911, (01)04026704400515(17)260628(10)KMZ21G0912, (01)04026704400515(17)260628(10)KMZ21G0917, (01)04026704400515(17)260628(10)KMZ21G0954, (01)04026704400515(17)260628(10)KMZ21G0955, (01)04026704400515(17)260628(10)KMZ21G0956, (01)04026704400515(17)260628(10)KMZ21G0964, (01)04026704400515(17)260628(10)KMZ21G0971, (01)04026704400515(17)260628(10)KMZ21G0982, (01)04026704400515(17)260628(10)KMZ21G0986, (01)04026704400515(17)260628(10)KMZ21G0992, (01)04026704400515(17)260628(10)KMZ21G0993, (01)04026704400515(17)260628(10)KMZ21G0995, (01)04026704400515(17)260628(10)KMZ21G1001, (01)04026704400515(17)260628(10)KMZ21G1008, (01)04026704400515(17)260628(10)KMZ21G1017, (01)04026704400515(17)260628(10)KMZ21G1037, (01)04026704400515(17)260628(10)KMZ21G1038, (01)04026704400515(17)260728(10)KMZ21H0115, (01)04026704400515(17)260728(10)KMZ21H0257, (01)04026704400515(17)260728(10)KMZ21H0299, (01)04026704400515(17)260728(10)KMZ21H0405, (01)04026704400515(17)260728(10)KMZ21H0429, (01)04026704400515(17)260728(10)KMZ21H0431, (01)04026704400515(17)260728(10)KMZ21H0433, (01)04026704400515(17)260728(10)KMZ21H0435, (01)04026704400515(17)260728(10)KMZ21H0504, (01)04026704400515(17)260728(10)KMZ21H0508, (01)04026704400515(17)260728(10)KMZ21H0526, (01)04026704400515(17)260728(10)KMZ21H0560, (01)04026704400515(17)260728(10)KMZ21H0561, (01)04026704400515(17)260828(10)KMZ21J0050, (01)04026704400515(17)260828(10)KMZ21J0077, (01)04026704400515(17)260828(10)KMZ21J0088, (01)04026704400515(17)260828(10)KMZ21J0089, (01)04026704400515(17)260828(10)KMZ21J0446, (01)04026704400515(17)260828(10)KMZ21J0447, (01)04026704400515(17)260828(10)KMZ21J0448, (01)04026704400515(17)260828(10)KMZ21J0484, (01)04026704400515(17)260828(10)KMZ21J0485, (01)04026704400515(17)260828(10)KMZ21J0534, (01)04026704400515(17)260828(10)KMZ21J0546, (01)04026704400515(17)260828(10)KMZ21J0554, (01)04026704400515(17)260828(10)KMZ21J0555, (01)04026704400515(17)260828(10)KMZ21J0570, (01)04026704400515(17)260828(10)KMZ21J0571, (01)04026704400515(17)260828(10)KMZ21J0579, (01)04026704400515(17)260828(10)KMZ21J0580, (01)04026704400515(17)260928(10)KMZ21K0045, (01)04026704400515(17)260928(10)KMZ21K0114, (01)04026704400515(17)260928(10)KMZ21K0163, (01)04026704400515(17)260928(10)KMZ21K0164, (01)04026704400515(17)260928(10)KMZ21K0266, (01)04026704400515(17)260928(10)KMZ21K0274, (01)04026704400515(17)260928(10)KMZ21K0295, (01)04026704400515(17)260928(10)KMZ21K0299, (01)04026704400515(17)260928(10)KMZ21K0308, (01)04026704400515(17)260928(10)KMZ21K0314, (01)04026704400515(17)260928(10)KMZ21K0318, (01)04026704400515(17)260928(10)KMZ21K0323, (01)04026704400515(17)260928(10)KMZ21K0452, (01)04026704400515(17)260928(10)KMZ21K0453, (01)04026704400515(17)260928(10)KMZ21K0509, (01)04026704400515(17)260928(10)KMZ21K0926, (01)04026704400515(17)260928(10)KMZ21K0953, (01)04026704400515(17)260928(10)KMZ21K0954, (01)04026704400515(17)261028(10)KMZ21L0047, (01)04026704400515(17)261028(10)KMZ21L0070, (01)04026704400515(17)261028(10)KMZ21L0074, (01)04026704400515(17)261028(10)KMZ21L0077, (01)04026704400515(17)261028(10)KMZ21L0079, (01)04026704400515(17)261028(10)KMZ21L0219, (01)04026704400515(17)261028(10)KMZ21L0220, (01)04026704400515(17)261028(10)KMZ21L0238, (01)04026704400515(17)261028(10)KMZ21L0239, (01)04026704400515(17)261028(10)KMZ21L0244, (01)04026704400515(17)261028(10)KMZ21L0295, (01)04026704400515(17)261028(10)KMZ21L0296, (01)04026704400515(17)261028(10)KMZ21L0307, (01)04026704400515(17)261028(10)KMZ21L0308, (01)04026704400515(17)261028(10)KMZ21L0316, (01)04026704400515(17)261028(10)KMZ21L0333, (01)04026704400515(17)261028(10)KMZ21L0334, (01)04026704400515(17)261028(10)KMZ21L0337, (01)04026704400515(17)261028(10)KMZ21L0344, (01)04026704400515(17)261028(10)KMZ21L0367, (01)04026704400515(17)261028(10)KMZ21L0368, (01)04026704400515(17)261028(10)KMZ21L0382, (01)04026704400515(17)261028(10)KMZ21L0389, (01)04026704400515(17)261028(10)KMZ21L0541, (01)04026704400515(17)261028(10)KMZ21L0798, (01)04026704400515(17)261028(10)KMZ21L0878, (01)04026704400515(17)261028(10)KMZ21L0880, (01)04026704400515(17)261028(10)KMZ21L0901, (01)04026704400515(17)261028(10)KMZ21L0924, (01)04026704400515(17)261128(10)KMZ21M0050, (01)04026704400515(17)261128(10)KMZ21M0690, (01)04026704400515(17)261128(10)KMZ21M0692, (01)04026704400515(17)261128(10)KMZ21M0719, (01)04026704400515(17)261128(10)KMZ21M0756, (01)04026704400515(17)261128(10)KMZ21M0781, (01)04026704400515(17)261128(10)KMZ21M0786, (01)04026704400515(17)261128(10)KMZ21M0790, (01)04026704400515(17)261128(10)KMZ21M0805, (01)04026704400515(17)261228(10)KMZ22A0635, (01)04026704400515(17)261228(10)KMZ22A0636, (01)04026704400515(17)261228(10)KMZ22A0650, (01)04026704400515(17)261228(10)KMZ22A0871, (01)04026704400515(17)270128(10)KMZ22B0130, (01)04026704400515(17)270128(10)KMZ22B0131, (01)04026704400515(17)270128(10)KMZ22B0192, (01)04026704400515(17)270128(10)KMZ22B0509, (01)04026704400515(17)270128(10)KMZ22B0524, (01)04026704400515(17)270128(10)KMZ22B0525, (01)04026704400515(17)270128(10)KMZ22B0556, (01)04026704400515(17)270128(10)KMZ22B0557, (01)04026704400515(17)270128(10)KMZ22B0784, (01)04026704400515(17)270128(10)KMZ22B0785, (01)04026704400515(17)270228(10)KMZ22C0395, (01)04026704400515(17)270228(10)KMZ22C0620, (01)04026704400515(17)270228(10)KMZ22C0802, (01)04026704400515(17)270228(10)KMZ22C0803.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- TELEFLEX LLC
- Reason for Recall:
- Incidents of device splitting or detaching during use
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Iso-Gard Filter S, with Expandi-Flex, Tethered Cap, REF: a) 19261T, OUS only; b) 19262T, OUS only; c) 19272T, OUS only; Microbial medical gas filter, single-use,
Product Codes/Lot Numbers:
a) 19261T, UDI: (01)04026704400492(17)250528(10)20FT51, (01)04026704400492(17)251028(10)KMH20L0409, (01)04026704400492(17)251228(10)KMH21A0086, (01)04026704400492(17)251228(10)KMH21A0190, (01)04026704400492(17)260128(10)KMH21B0163, (01)04026704400492(17)260128(10)KMH21B0343, (01)04026704400492(17)260128(10)KMH21B0355, (01)04026704400492(17)260228(10)KMH21C0177, (01)04026704400492(17)260228(10)KMH21C0387, (01)04026704400492(17)260528(10)KMH21F0156, (01)04026704400492(17)260528(10)KMH21F0337, (01)04026704400492(17)260628(10)KMH21G0426, (01)04026704400492(17)260828(10)KMH21J0020, (01)04026704400492(17)260828(10)KMH21J0245, (01)04026704400492(17)260928(10)KMH21K0142, (01)04026704400492(17)261128(10)KMH21M0182, (01)04026704400492(17)270128(10)KMH22B0151, (01)04026704400492(17)270128(10)KMH22B0248; b) 19262T, UDI: (01)04026704400508(17)250528(10)202025, (01)04026704400508(17)250528(10)202026, (01)04026704400508(17)250628(10)202030, (01)04026704400508(17)250828(10)KMZ20J0163, (01)04026704400508(17)250828(10)KMZ20J0171, (01)04026704400508(17)250828(10)KMZ20J0582, (01)04026704400508(17)250828(10)KMZ20J0586, (01)04026704400508(17)250828(10)KMZ20J0587, (01)04026704400508(17)250928(10)KMZ20K0681, (01)04026704400508(17)251028(10)KMZ20L0968, (01)04026704400508(17)251028(10)KMZ20L0977, (01)04026704400508(17)251028(10)KMZ20L1275, (01)04026704400508(17)251128(10)KMZ20M0544, (01)04026704400508(17)251128(10)KMZ20M0661, (01)04026704400508(17)251228(10)KMZ21A0175, (01)04026704400508(17)251228(10)KMZ21A0186, (01)04026704400508(17)251228(10)KMZ21A0187, (01)04026704400508(17)251228(10)KMZ21A0208, (01)04026704400508(17)260128(10)KMZ21B0483, (01)04026704400508(17)260128(10)KMZ21B0485, (01)04026704400508(17)260128(10)KMZ21B0495, (01)04026704400508(17)260128(10)KMZ21B0811, (01)04026704400508(17)260128(10)KMZ21B0918, (01)04026704400508(17)260128(10)KMZ21B1172, (01)04026704400508(17)260228(10)KMZ21C0884, (01)04026704400508(17)260228(10)KMZ21C0910, (01)04026704400508(17)260228(10)KMZ21C0929, (01)04026704400508(17)260228(10)KMZ21C0931, (01)04026704400508(17)260328(10)KMZ21D0419, (01)04026704400508(17)260528(10)KMZ21F0223, (01)04026704400508(17)260528(10)KMZ21F0260, (01)04026704400508(17)260528(10)KMZ21F0528, (01)04026704400508(17)260628(10)KMZ21G0728, (01)04026704400508(17)260628(10)KMZ21G0735, (01)04026704400508(17)260628(10)KMZ21G0739, (01)04026704400508(17)260728(10)KMZ21H0055, (01)04026704400508(17)260828(10)KMZ21J0006, (01)04026704400508(17)260828(10)KMZ21J0007, (01)04026704400508(17)260828(10)KMZ21J0008, (01)04026704400508(17)260928(10)KMZ21K0510, (01)04026704400508(17)260928(10)KMZ21K0524, (01)04026704400508(17)260928(10)KMZ21K0539, (01)04026704400508(17)260928(10)KMZ21K0572, (01)04026704400508(17)260928(10)KMZ21K0578, (01)04026704400508(17)260928(10)KMZ21K0592, (01)04026704400508(17)260928(10)KMZ21K0952, (01)04026704400508(17)261028(10)KMZ21L0218, (01)04026704400508(17)261128(10)KMZ21M0674, (01)04026704400508(17)270128(10)KMZ22B0125, (01)04026704400508(17)270128(10)KMZ22B0345, (01)04026704400508(17)270128(10)KMZ22B0809, (01)04026704400508(17)270128(10)KMZ22B0828, (01)04026704400508(17)270128(10)KMZ22B0838, (01)04026704400508(17)270128(10)KMZ22B0869, (01)04026704400508(17)270228(10)KMZ22C0659; c) 19272T, UDI: (01)04026704400515(17)250528(10)202025, (01)04026704400515(17)250528(10)202026, (01)04026704400515(17)250628(10)202027, (01)04026704400515(17)250628(10)202028, (01)04026704400515(17)250628(10)202029, (01)04026704400515(17)250628(10)202030, (01)04026704400515(17)250628(10)202031, (01)04026704400515(17)250628(10)KMZ20G0005, (01)04026704400515(17)250628(10)KMZ20G0026, (01)04026704400515(17)250628(10)KMZ20G0105, (01)04026704400515(17)250628(10)KMZ20G0112, (01)04026704400515(17)250728(10)KMZ20H0081, (01)04026704400515(17)250728(10)KMZ20H0096, (01)04026704400515(17)250728(10)KMZ20H0158, (01)04026704400515(17)250728(10)KMZ20H0379, (01)04026704400515(17)250728(10)KMZ20H0530, (01)04026704400515(17)250728(10)KMZ20H0540, (01)04026704400515(17)250728(10)KMZ20H0564, (01)04026704400515(17)250728(10)KMZ20H0812, (01)04026704400515(17)250728(10)KMZ20H0841, (01)04026704400515(17)250728(10)KMZ20H1021, (01)04026704400515(17)250728(10)KMZ20H1024, (01)04026704400515(17)250728(10)KMZ20H1031, (01)04026704400515(17)250828(10)KMZ20J0092, (01)04026704400515(17)250828(10)KMZ20J0102, (01)04026704400515(17)250828(10)KMZ20J0227, (01)04026704400515(17)250828(10)KMZ20J0441, (01)04026704400515(17)250828(10)KMZ20J0535, (01)04026704400515(17)250828(10)KMZ20J0551, (01)04026704400515(17)250828(10)KMZ20J0566, (01)04026704400515(17)250828(10)KMZ20J0640, (01)04026704400515(17)250828(10)KMZ20J0654, (01)04026704400515(17)250828(10)KMZ20J0663, (01)04026704400515(17)250828(10)KMZ20J0893, (01)04026704400515(17)250828(10)KMZ20J0958, (01)04026704400515(17)250928(10)KMZ20K0034, (01)04026704400515(17)250928(10)KMZ20K0196, (01)04026704400515(17)250928(10)KMZ20K0236, (01)04026704400515(17)250928(10)KMZ20K0430, (01)04026704400515(17)250928(10)KMZ20K0659, (01)04026704400515(17)250928(10)KMZ20K0662, (01)04026704400515(17)250928(10)KMZ20K0838, (01)04026704400515(17)250928(10)KMZ20K0946, (01)04026704400515(17)250928(10)KMZ20K1260, (01)04026704400515(17)250928(10)KMZ20K1289, (01)04026704400515(17)250928(10)KMZ20K1294, (01)04026704400515(17)250928(10)KMZ20K1310, (01)04026704400515(17)250928(10)KMZ20K1339, (01)04026704400515(17)250928(10)KMZ20K1340, (01)04026704400515(17)251028(10)KMZ20L0003, (01)04026704400515(17)251028(10)KMZ20L0030, (01)04026704400515(17)251028(10)KMZ20L0031, (01)04026704400515(17)251028(10)KMZ20L0117, (01)04026704400515(17)251028(10)KMZ20L0214, (01)04026704400515(17)251028(10)KMZ20L0351, (01)04026704400515(17)251028(10)KMZ20L0359, (01)04026704400515(17)251028(10)KMZ20L0499, (01)04026704400515(17)251028(10)KMZ20L0537, (01)04026704400515(17)251028(10)KMZ20L0706, (01)04026704400515(17)251028(10)KMZ20L0724, (01)04026704400515(17)251028(10)KMZ20L0885, (01)04026704400515(17)251028(10)KMZ20L0886, (01)04026704400515(17)251028(10)KMZ20L0933, (01)04026704400515(17)251028(10)KMZ20L0945, (01)04026704400515(17)251028(10)KMZ20L1056, (01)04026704400515(17)251028(10)KMZ20L1091, (01)04026704400515(17)251028(10)KMZ20L1103, (01)04026704400515(17)251028(10)KMZ20L1104, (01)04026704400515(17)251128(10)KMZ20M0387, (01)04026704400515(17)251128(10)KMZ20M0777, (01)04026704400515(17)251128(10)KMZ20M0778, (01)04026704400515(17)251128(10)KMZ20M0787, (01)04026704400515(17)251128(10)KMZ20M0822, (01)04026704400515(17)251128(10)KMZ20M0842, (01)04026704400515(17)251128(10)KMZ20M0938, (01)04026704400515(17)251128(10)KMZ20M0967, (01)04026704400515(17)251128(10)KMZ20M0990, (01)04026704400515(17)251128(10)KMZ20M1249, (01)04026704400515(17)251128(10)KMZ20M1250, (01)04026704400515(17)251128(10)KMZ20M1257, (01)04026704400515(17)251128(10)KMZ20M1267, (01)04026704400515(17)251228(10)KMZ21A0012, (01)04026704400515(17)251228(10)KMZ21A0024, (01)04026704400515(17)251228(10)KMZ21A0117, (01)04026704400515(17)251228(10)KMZ21A0126, (01)04026704400515(17)251228(10)KMZ21A0155, (01)04026704400515(17)251228(10)KMZ21A0162, (01)04026704400515(17)251228(10)KMZ21A0164, (01)04026704400515(17)251228(10)KMZ21A0203, (01)04026704400515(17)251228(10)KMZ21A0218, (01)04026704400515(17)251228(10)KMZ21A0226, (01)04026704400515(17)251228(10)KMZ21A0368, (01)04026704400515(17)251228(10)KMZ21A0618, (01)04026704400515(17)251228(10)KMZ21A0635, (01)04026704400515(17)251228(10)KMZ21A0693, (01)04026704400515(17)251228(10)KMZ21A0706, (01)04026704400515(17)251228(10)KMZ21A0714, (01)04026704400515(17)251228(10)KMZ21A0725, (01)04026704400515(17)251228(10)KMZ21A0734, (01)04026704400515(17)251228(10)KMZ21A0950, (01)04026704400515(17)260128(10)KMZ21B0101, (01)04026704400515(17)260128(10)KMZ21B0163, (01)04026704400515(17)260128(10)KMZ21B0181, (01)04026704400515(17)260128(10)KMZ21B0183, (01)04026704400515(17)260128(10)KMZ21B0185, (01)04026704400515(17)260128(10)KMZ21B0189, (01)04026704400515(17)260128(10)KMZ21B0227, (01)04026704400515(17)260128(10)KMZ21B0234, (01)04026704400515(17)260128(10)KMZ21B0305, (01)04026704400515(17)260128(10)KMZ21B0320, (01)04026704400515(17)260128(10)KMZ21B0417, (01)04026704400515(17)260128(10)KMZ21B0421, (01)04026704400515(17)260128(10)KMZ21B0435, (01)04026704400515(17)260128(10)KMZ21B0465, (01)04026704400515(17)260128(10)KMZ21B0471, (01)04026704400515(17)260128(10)KMZ21B0479, (01)04026704400515(17)260128(10)KMZ21B0533, (01)04026704400515(17)260128(10)KMZ21B0689, (01)04026704400515(17)260128(10)KMZ21B1171, (01)04026704400515(17)260228(10)KMZ21C0270, (01)04026704400515(17)260228(10)KMZ21C0388, (01)04026704400515(17)260228(10)KMZ21C0398, (01)04026704400515(17)260228(10)KMZ21C0883, (01)04026704400515(17)260228(10)KMZ21C1159, (01)04026704400515(17)260228(10)KMZ21C1160, (01)04026704400515(17)260228(10)KMZ21C1358, (01)04026704400515(17)260328(10)KMZ21D0420, (01)04026704400515(17)260528(10)KMZ21F0227, (01)04026704400515(17)260528(10)KMZ21F0265, (01)04026704400515(17)260528(10)KMZ21F0296, (01)04026704400515(17)260528(10)KMZ21F0335, (01)04026704400515(17)260528(10)KMZ21F0375, (01)04026704400515(17)260528(10)KMZ21F0430, (01)04026704400515(17)260528(10)KMZ21F0467, (01)04026704400515(17)260528(10)KMZ21F0468, (01)04026704400515(17)260528(10)KMZ21F0478, (01)04026704400515(17)260528(10)KMZ21F0501, (01)04026704400515(17)260528(10)KMZ21F0517, (01)04026704400515(17)260528(10)KMZ21F0534, (01)04026704400515(17)260528(10)KMZ21F0550, (01)04026704400515(17)260528(10)KMZ21F0551, (01)04026704400515(17)260528(10)KMZ21F0552, (01)04026704400515(17)260528(10)KMZ21F0557, (01)04026704400515(17)260528(10)KMZ21F0567, (01)04026704400515(17)260528(10)KMZ21F0568, (01)04026704400515(17)260528(10)KMZ21F0588, (01)04026704400515(17)260528(10)KMZ21F0594, (01)04026704400515(17)260528(10)KMZ21F0742, (01)04026704400515(17)260528(10)KMZ21F0777, (01)04026704400515(17)260628(10)KMZ21G0399, (01)04026704400515(17)260628(10)KMZ21G0403, (01)04026704400515(17)260628(10)KMZ21G0404, (01)04026704400515(17)260628(10)KMZ21G0751, (01)04026704400515(17)260628(10)KMZ21G0765, (01)04026704400515(17)260628(10)KMZ21G0770, (01)04026704400515(17)260628(10)KMZ21G0778, (01)04026704400515(17)260628(10)KMZ21G0786, (01)04026704400515(17)260628(10)KMZ21G0800, (01)04026704400515(17)260628(10)KMZ21G0806, (01)04026704400515(17)260628(10)KMZ21G0838, (01)04026704400515(17)260628(10)KMZ21G0911, (01)04026704400515(17)260628(10)KMZ21G0912, (01)04026704400515(17)260628(10)KMZ21G0917, (01)04026704400515(17)260628(10)KMZ21G0954, (01)04026704400515(17)260628(10)KMZ21G0955, (01)04026704400515(17)260628(10)KMZ21G0956, (01)04026704400515(17)260628(10)KMZ21G0964, (01)04026704400515(17)260628(10)KMZ21G0971, (01)04026704400515(17)260628(10)KMZ21G0982, (01)04026704400515(17)260628(10)KMZ21G0986, (01)04026704400515(17)260628(10)KMZ21G0992, (01)04026704400515(17)260628(10)KMZ21G0993, (01)04026704400515(17)260628(10)KMZ21G0995, (01)04026704400515(17)260628(10)KMZ21G1001, (01)04026704400515(17)260628(10)KMZ21G1008, (01)04026704400515(17)260628(10)KMZ21G1017, (01)04026704400515(17)260628(10)KMZ21G1037, (01)04026704400515(17)260628(10)KMZ21G1038, (01)04026704400515(17)260728(10)KMZ21H0115, (01)04026704400515(17)260728(10)KMZ21H0257, (01)04026704400515(17)260728(10)KMZ21H0299, (01)04026704400515(17)260728(10)KMZ21H0405, (01)04026704400515(17)260728(10)KMZ21H0429, (01)04026704400515(17)260728(10)KMZ21H0431, (01)04026704400515(17)260728(10)KMZ21H0433, (01)04026704400515(17)260728(10)KMZ21H0435, (01)04026704400515(17)260728(10)KMZ21H0504, (01)04026704400515(17)260728(10)KMZ21H0508, (01)04026704400515(17)260728(10)KMZ21H0526, (01)04026704400515(17)260728(10)KMZ21H0560, (01)04026704400515(17)260728(10)KMZ21H0561, (01)04026704400515(17)260828(10)KMZ21J0050, (01)04026704400515(17)260828(10)KMZ21J0077, (01)04026704400515(17)260828(10)KMZ21J0088, (01)04026704400515(17)260828(10)KMZ21J0089, (01)04026704400515(17)260828(10)KMZ21J0446, (01)04026704400515(17)260828(10)KMZ21J0447, (01)04026704400515(17)260828(10)KMZ21J0448, (01)04026704400515(17)260828(10)KMZ21J0484, (01)04026704400515(17)260828(10)KMZ21J0485, (01)04026704400515(17)260828(10)KMZ21J0534, (01)04026704400515(17)260828(10)KMZ21J0546, (01)04026704400515(17)260828(10)KMZ21J0554, (01)04026704400515(17)260828(10)KMZ21J0555, (01)04026704400515(17)260828(10)KMZ21J0570, (01)04026704400515(17)260828(10)KMZ21J0571, (01)04026704400515(17)260828(10)KMZ21J0579, (01)04026704400515(17)260828(10)KMZ21J0580, (01)04026704400515(17)260928(10)KMZ21K0045, (01)04026704400515(17)260928(10)KMZ21K0114, (01)04026704400515(17)260928(10)KMZ21K0163, (01)04026704400515(17)260928(10)KMZ21K0164, (01)04026704400515(17)260928(10)KMZ21K0266, (01)04026704400515(17)260928(10)KMZ21K0274, (01)04026704400515(17)260928(10)KMZ21K0295, (01)04026704400515(17)260928(10)KMZ21K0299, (01)04026704400515(17)260928(10)KMZ21K0308, (01)04026704400515(17)260928(10)KMZ21K0314, (01)04026704400515(17)260928(10)KMZ21K0318, (01)04026704400515(17)260928(10)KMZ21K0323, (01)04026704400515(17)260928(10)KMZ21K0452, (01)04026704400515(17)260928(10)KMZ21K0453, (01)04026704400515(17)260928(10)KMZ21K0509, (01)04026704400515(17)260928(10)KMZ21K0926, (01)04026704400515(17)260928(10)KMZ21K0953, (01)04026704400515(17)260928(10)KMZ21K0954, (01)04026704400515(17)261028(10)KMZ21L0047, (01)04026704400515(17)261028(10)KMZ21L0070, (01)04026704400515(17)261028(10)KMZ21L0074, (01)04026704400515(17)261028(10)KMZ21L0077, (01)04026704400515(17)261028(10)KMZ21L0079, (01)04026704400515(17)261028(10)KMZ21L0219, (01)04026704400515(17)261028(10)KMZ21L0220, (01)04026704400515(17)261028(10)KMZ21L0238, (01)04026704400515(17)261028(10)KMZ21L0239, (01)04026704400515(17)261028(10)KMZ21L0244, (01)04026704400515(17)261028(10)KMZ21L0295, (01)04026704400515(17)261028(10)KMZ21L0296, (01)04026704400515(17)261028(10)KMZ21L0307, (01)04026704400515(17)261028(10)KMZ21L0308, (01)04026704400515(17)261028(10)KMZ21L0316, (01)04026704400515(17)261028(10)KMZ21L0333, (01)04026704400515(17)261028(10)KMZ21L0334, (01)04026704400515(17)261028(10)KMZ21L0337, (01)04026704400515(17)261028(10)KMZ21L0344, (01)04026704400515(17)261028(10)KMZ21L0367, (01)04026704400515(17)261028(10)KMZ21L0368, (01)04026704400515(17)261028(10)KMZ21L0382, (01)04026704400515(17)261028(10)KMZ21L0389, (01)04026704400515(17)261028(10)KMZ21L0541, (01)04026704400515(17)261028(10)KMZ21L0798, (01)04026704400515(17)261028(10)KMZ21L0878, (01)04026704400515(17)261028(10)KMZ21L0880, (01)04026704400515(17)261028(10)KMZ21L0901, (01)04026704400515(17)261028(10)KMZ21L0924, (01)04026704400515(17)261128(10)KMZ21M0050, (01)04026704400515(17)261128(10)KMZ21M0690, (01)04026704400515(17)261128(10)KMZ21M0692, (01)04026704400515(17)261128(10)KMZ21M0719, (01)04026704400515(17)261128(10)KMZ21M0756, (01)04026704400515(17)261128(10)KMZ21M0781, (01)04026704400515(17)261128(10)KMZ21M0786, (01)04026704400515(17)261128(10)KMZ21M0790, (01)04026704400515(17)261128(10)KMZ21M0805, (01)04026704400515(17)261228(10)KMZ22A0635, (01)04026704400515(17)261228(10)KMZ22A0636, (01)04026704400515(17)261228(10)KMZ22A0650, (01)04026704400515(17)261228(10)KMZ22A0871, (01)04026704400515(17)270128(10)KMZ22B0130, (01)04026704400515(17)270128(10)KMZ22B0131, (01)04026704400515(17)270128(10)KMZ22B0192, (01)04026704400515(17)270128(10)KMZ22B0509, (01)04026704400515(17)270128(10)KMZ22B0524, (01)04026704400515(17)270128(10)KMZ22B0525, (01)04026704400515(17)270128(10)KMZ22B0556, (01)04026704400515(17)270128(10)KMZ22B0557, (01)04026704400515(17)270128(10)KMZ22B0784, (01)04026704400515(17)270128(10)KMZ22B0785, (01)04026704400515(17)270228(10)KMZ22C0395, (01)04026704400515(17)270228(10)KMZ22C0620, (01)04026704400515(17)270228(10)KMZ22C0802, (01)04026704400515(17)270228(10)KMZ22C0803.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0075-2023
Related Recalls
There are incomplete pre-cleaning instructions within the Instructions for Use (IFU).
There are incomplete pre-cleaning instructions within the Instructions for Use (IFU).
There are incomplete pre-cleaning instructions within the Instructions for Use (IFU).