Aespire 7100, 7100 (Variant), and Aespire 7100 S/5 Protiva configurations. These anesthesia gas machines are intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.
Class I - DangerousWhat Should You Do?
- Check if you have this product: a. Aespire 7100 - GTIN: 00840682102261. Model Number: 1009-9000-000. All Serial Numbers. b. Aespire 7100 (Variant) - GTIN: 00840682102261. Model Numbers: 1009-9101-000 and 1009-9101-000. All Serial Numbers. c. Aespire 7100 S/5 Protiva - No GTIN. Model Number: 1009-9004-000. All Serial Numbers. Not model or lot specific. All devices of the indicated products configured with the AGCO option are potentially affected.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Datex-Ohmeda, Inc.
- Reason for Recall:
- Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Aespire 7100, 7100 (Variant), and Aespire 7100 S/5 Protiva configurations. These anesthesia gas machines are intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.
Product Codes/Lot Numbers:
a. Aespire 7100 - GTIN: 00840682102261. Model Number: 1009-9000-000. All Serial Numbers. b. Aespire 7100 (Variant) - GTIN: 00840682102261. Model Numbers: 1009-9101-000 and 1009-9101-000. All Serial Numbers. c. Aespire 7100 S/5 Protiva - No GTIN. Model Number: 1009-9004-000. All Serial Numbers. Not model or lot specific. All devices of the indicated products configured with the AGCO option are potentially affected.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0074-2025
Related Recalls
Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.