Medtronic Navigation O-arm O2 Surgical Imaging System Product Catalog Number: BI-700-02000 Product Usage: The O-arm Surgical Imaging System is a multi-dimensional surgical imaging platform that is designed for use in spine, orthopaedic, and trauma-related surgeries. It provides real-time, intra-operative imaging of a patients anatomy with high quality images and a large field-of-view in both two and three dimensions
Class I - DangerousWhat Should You Do?
- Check if you have this product: Software revisions 4.0.0, 4.0.1 and 4.0.2. System Serial Numbers: C0930 C0981 C0986 C1113 C1069 C1050 C1208 C1043 C1144 C1163 C1145 C1162 C1002 C1074
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Navigation, Inc.-Littleton
- Reason for Recall:
- O-arm O2 Surgical Imaging System Spatial calibration may be erroneous in Stealth Station navigated images
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Medtronic Navigation O-arm O2 Surgical Imaging System Product Catalog Number: BI-700-02000 Product Usage: The O-arm Surgical Imaging System is a multi-dimensional surgical imaging platform that is designed for use in spine, orthopaedic, and trauma-related surgeries. It provides real-time, intra-operative imaging of a patients anatomy with high quality images and a large field-of-view in both two and three dimensions
Product Codes/Lot Numbers:
Software revisions 4.0.0, 4.0.1 and 4.0.2. System Serial Numbers: C0930 C0981 C0986 C1113 C1069 C1050 C1208 C1043 C1144 C1163 C1145 C1162 C1002 C1074
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0070-2017
Related Recalls
O-arm O2 Imaging System-Mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients. Product Number: BI70002000
Medtronic Navigation, Inc.-Littleton
Potential for an electrical component of the O-arm" O2 Imaging System to intermittently shut off,failure can cause the potential for surgical delay, additional imaging, cancellation and rescheduling of surgery, or completion of surgery without the benefit of image guidance.
Ground cable installed incorrectly.
O-arm" O2 Imaging System, Product Number/CFN: BI-700-02000
Medtronic Navigation, Inc.-Littleton
Due to a manufacturing issue affecting particular O-arm" O2 Imaging Systems, Medtronic has determined that the gantry tractor motor drive belt may loosen over an extended number of uses at a higher frequency in these systems. If the belt loosens, it may result in rotation of 3D images about the gantry isocenter. While the 3D image is anatomically accurate within the image itself and may be used to confirm therapy, its electronic registered location may be rotated relative to the actual physical position of the patient.