Automated Impella Controller (AIC), used for left heart support blood pump, labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, AU; Product Code: 0042-0000-AU. 2. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 3. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 4. Impella Controller, Packaged, EU, Loaner; Product Code: 0042-0000-EU-L. 5. Impella Controller, Packaged, IN; Product Code: 0042-0000-IN. 6. Impella Controller, Packaged, JP; Product Code: 0042-0000-JP. 7. Impella Controller, Japan, Loaner; Product Code: 0042-0000-JP-L. 8. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 9. Impella Controller, UK, Loaner; Product Code: 0042-0000-UK-L. 10. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 11. Impella Controller, Packaged, US, Loaner; Product Code: 0042-0000-US-L. 12. Impella Optical Controller, Packaged, AU; Product Code: 0042-0010-AU. 13. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 14. Impella Optical Controller, EU, Loaner; Product Code: 0042-0010-EU-L. 15. Impella Optical Controller, Packaged, IN; Product Code: 0042-0010-IN. 16. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 17. Impella Optical Controller, UK, Loaner; Product Code: 0042-0010-UK-L. 18. Impella Optical Controller, Packaged, US; Product Code: 0042-0010-US. 19. Impella Optical Controller, US, Loaner; Product Code: 0042-0010-US-L. 20. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU. 21. Optical AIC w/Impella Connect,Pack'd, CA; Product Code: 0042-0040-CA. 22. AIC w/Remote Link, Pkgd, Cln Trl, US; Product Code: 0042-0040-CT. 23. AIC w/Remote Link, Pkgd, Cln Trl, US; Product Code: 0042-0040-CT-L. 24. Optical AIC w/Impella Connect,Pack'd, EU; Product Code: 0042-0040-EU. 25. Optical AIC w/Impella Connect,EU,Loaner; Product Code: 0042-0040-EU-L. 26. Optical AIC w/Impella Connect, Pkgd, JP; Product Code: 0042-0040-JP. 27. Optical AIC w/Impella Connect, Pkgd, JP; Product Code: 0042-0040-JP-L. 28. Optical, AIC, Impella Connect, Pkgd, UK; Product Code: 0042-0040-UK. 29. Optical, AIC, Impella Connect, Pkgd, US; Product Code: 0042-0040-US. 30. Optical AIC Impella Connect, US, Loaner; Product Code: 0042-0040-US-L. 31. Dbl optical, AIC Impella Connect, Phg US; Product Code: 1000201. 32. AIC w/Impella Connect for ECP; Product Code: 1000432.

Class I - Dangerous
🏥 Medical Devices Recalled: October 1, 2025 Abiomed Infusion Pumps Nationwide

What Should You Do?

  1. Check if you have this product:
    1. Product Code: 0042-0000-AU; GTIN: 813502010886; All Serial Numbers. 2. Product Code: 0042-0000-CA; GTIN: 813502011272; All Serial Numbers. 3. Product Code: 0042-0000-EU; GTIN: 813502011289; All Serial Numbers. 4. Product Code: 0042-0000-EU-L; GTIN: 813502011289; All Serial Numbers. 5. Product Code: 0042-0000-IN; GTIN: 813502011418; All Serial Numbers. 6. Product Code: 0042-0000-JP; GTIN: 813502010428; All Serial Numbers. 7. Product Code: 0042-0000-JP-L; GTIN: 813502010428; All Serial Numbers. 8. Product Code: 0042-0000-UK; GTIN: 813502011296; All Serial Numbers. 9. Product Code: 0042-0000-UK-L; GTIN: 813502011296; All Serial Numbers. 10. Product Code: 0042-0000-US; GTIN: 813502010022; All Serial Numbers. 11. Product Code: 0042-0000-US-L; GTIN: 813502010022; All Serial Numbers. 12. Product Code: 0042-0010-AU; GTIN: 813502010930; All Serial Numbers. 13. Product Code: 0042-0010-EU; GTIN: 813502010961; All Serial Numbers. 14. Product Code: 0042-0010-EU-L; GTIN: 813502010961; All Serial Numbers. 15. Product Code: 0042-0010-IN; GTIN: 813502011425; All Serial Numbers. 16. Product Code: 0042-0010-UK; GTIN: 813502010978; All Serial Numbers. 17. Product Code: 0042-0010-UK-L; GTIN: 813502010978; All Serial Numbers. 18. Product Code: 0042-0010-US; GTIN: 813502010985; All Serial Numbers. 19. Product Code: 0042-0010-US-L; GTIN: 813502010985; All Serial Numbers. 20. Product Code: 0042-0040-AU; GTIN: 813502012972; All Serial Numbers. 21. Product Code: 0042-0040-CA; GTIN: 813502012958; All Serial Numbers. 22. Product Code: 0042-0040-CT; GTIN: 813502011722; All Serial Numbers. 23. Product Code: 0042-0040-CT-L; GTIN: 813502011722; All Serial Numbers. 24. Product Code: 0042-0040-EU; GTIN: 813502012217; All Serial Numbers. 25. Product Code: 0042-0040-EU-L; GTIN: 813502012217; All Serial Numbers. 26. Product Code: 0042-0040-JP; GTIN: 813502011814; All Serial Numbers. 27. Product Code: 0042-0040-JP-L; GTIN: 813502011814; All Serial Numbers. 28. Product Code: 0042-0040-UK; GTIN: 813502012484; All Serial Numbers. 29. Product Code: 0042-0040-US; GTIN: 813502011401; All Serial Numbers. 30. Product Code: 0042-0040-US-L; GTIN: 813502011401; All Serial Numbers. 31. Product Code: 1000201; GTIN: 813502010442; All Serial Numbers. 32. Product Code: 1000432; GTIN: 813502013030; All Serial Numbers.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Abiomed, Inc.
Reason for Recall:
Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Automated Impella Controller (AIC), used for left heart support blood pump, labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, AU; Product Code: 0042-0000-AU. 2. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 3. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 4. Impella Controller, Packaged, EU, Loaner; Product Code: 0042-0000-EU-L. 5. Impella Controller, Packaged, IN; Product Code: 0042-0000-IN. 6. Impella Controller, Packaged, JP; Product Code: 0042-0000-JP. 7. Impella Controller, Japan, Loaner; Product Code: 0042-0000-JP-L. 8. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 9. Impella Controller, UK, Loaner; Product Code: 0042-0000-UK-L. 10. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 11. Impella Controller, Packaged, US, Loaner; Product Code: 0042-0000-US-L. 12. Impella Optical Controller, Packaged, AU; Product Code: 0042-0010-AU. 13. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 14. Impella Optical Controller, EU, Loaner; Product Code: 0042-0010-EU-L. 15. Impella Optical Controller, Packaged, IN; Product Code: 0042-0010-IN. 16. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 17. Impella Optical Controller, UK, Loaner; Product Code: 0042-0010-UK-L. 18. Impella Optical Controller, Packaged, US; Product Code: 0042-0010-US. 19. Impella Optical Controller, US, Loaner; Product Code: 0042-0010-US-L. 20. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU. 21. Optical AIC w/Impella Connect,Pack'd, CA; Product Code: 0042-0040-CA. 22. AIC w/Remote Link, Pkgd, Cln Trl, US; Product Code: 0042-0040-CT. 23. AIC w/Remote Link, Pkgd, Cln Trl, US; Product Code: 0042-0040-CT-L. 24. Optical AIC w/Impella Connect,Pack'd, EU; Product Code: 0042-0040-EU. 25. Optical AIC w/Impella Connect,EU,Loaner; Product Code: 0042-0040-EU-L. 26. Optical AIC w/Impella Connect, Pkgd, JP; Product Code: 0042-0040-JP. 27. Optical AIC w/Impella Connect, Pkgd, JP; Product Code: 0042-0040-JP-L. 28. Optical, AIC, Impella Connect, Pkgd, UK; Product Code: 0042-0040-UK. 29. Optical, AIC, Impella Connect, Pkgd, US; Product Code: 0042-0040-US. 30. Optical AIC Impella Connect, US, Loaner; Product Code: 0042-0040-US-L. 31. Dbl optical, AIC Impella Connect, Phg US; Product Code: 1000201. 32. AIC w/Impella Connect for ECP; Product Code: 1000432.

Product Codes/Lot Numbers:

1. Product Code: 0042-0000-AU; GTIN: 813502010886; All Serial Numbers. 2. Product Code: 0042-0000-CA; GTIN: 813502011272; All Serial Numbers. 3. Product Code: 0042-0000-EU; GTIN: 813502011289; All Serial Numbers. 4. Product Code: 0042-0000-EU-L; GTIN: 813502011289; All Serial Numbers. 5. Product Code: 0042-0000-IN; GTIN: 813502011418; All Serial Numbers. 6. Product Code: 0042-0000-JP; GTIN: 813502010428; All Serial Numbers. 7. Product Code: 0042-0000-JP-L; GTIN: 813502010428; All Serial Numbers. 8. Product Code: 0042-0000-UK; GTIN: 813502011296; All Serial Numbers. 9. Product Code: 0042-0000-UK-L; GTIN: 813502011296; All Serial Numbers. 10. Product Code: 0042-0000-US; GTIN: 813502010022; All Serial Numbers. 11. Product Code: 0042-0000-US-L; GTIN: 813502010022; All Serial Numbers. 12. Product Code: 0042-0010-AU; GTIN: 813502010930; All Serial Numbers. 13. Product Code: 0042-0010-EU; GTIN: 813502010961; All Serial Numbers. 14. Product Code: 0042-0010-EU-L; GTIN: 813502010961; All Serial Numbers. 15. Product Code: 0042-0010-IN; GTIN: 813502011425; All Serial Numbers. 16. Product Code: 0042-0010-UK; GTIN: 813502010978; All Serial Numbers. 17. Product Code: 0042-0010-UK-L; GTIN: 813502010978; All Serial Numbers. 18. Product Code: 0042-0010-US; GTIN: 813502010985; All Serial Numbers. 19. Product Code: 0042-0010-US-L; GTIN: 813502010985; All Serial Numbers. 20. Product Code: 0042-0040-AU; GTIN: 813502012972; All Serial Numbers. 21. Product Code: 0042-0040-CA; GTIN: 813502012958; All Serial Numbers. 22. Product Code: 0042-0040-CT; GTIN: 813502011722; All Serial Numbers. 23. Product Code: 0042-0040-CT-L; GTIN: 813502011722; All Serial Numbers. 24. Product Code: 0042-0040-EU; GTIN: 813502012217; All Serial Numbers. 25. Product Code: 0042-0040-EU-L; GTIN: 813502012217; All Serial Numbers. 26. Product Code: 0042-0040-JP; GTIN: 813502011814; All Serial Numbers. 27. Product Code: 0042-0040-JP-L; GTIN: 813502011814; All Serial Numbers. 28. Product Code: 0042-0040-UK; GTIN: 813502012484; All Serial Numbers. 29. Product Code: 0042-0040-US; GTIN: 813502011401; All Serial Numbers. 30. Product Code: 0042-0040-US-L; GTIN: 813502011401; All Serial Numbers. 31. Product Code: 1000201; GTIN: 813502010442; All Serial Numbers. 32. Product Code: 1000432; GTIN: 813502013030; All Serial Numbers.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0069-2026

Related Recalls

Automated Impella Controller (AIC) labeled as the following with corresponding Product Codes: 1. AIC w/Impella Connect for ECP, Product Code: 1000432. 2. Dbl optical, AIC Impella Connect, Phg US, Product Code: 1000201. 3. Impella Controller, Packaged, AU, Product Code: 0042-0000-AU. 4. Impella Controller, Packaged, CA, Product Code: 0042-0000-CA. 5. Impella Controller, Packaged, EU, Product Code: 0042-0000-EU. 6. Impella Controller, Packaged, IN, Product Code: 0042-0000-IN. 7. Impella Controller, Packaged, JP, Product Code: 0042-0000-JP. 8. Impella Controller, Packaged, UK, Product Code: 0042-0000-UK. 9. Impella Controller, Packaged, US, Product Code: 0042-0000-US. 10. Impella Optical Controller, Packaged, AU, Product Code: 0042-0010-AU. 11. Impella Optical Controller, Packaged, EU, Product Code: 0042-0010-EU. 12. Impella Optical Controller, Packaged, IN, Product Code: 0042-0010-IN. 13. Impella Optical Controller, Packaged, UK, Product Code: 0042-0010-UK. 14. Impella Optical Controller, Packaged, US, Product Code: 0042-0010-US. 15. Optical AIC w/Impella Connect, Pkgd, JP, Product Codes: 0042-0040-JP. 16. Optical AIC w/Impella Connect, Pkgd, JP, Product Codes: 0042-0040-JP-L. 17. Optical AIC w/Impella Connect, Pack'd, CA, Product Code: 0042-0040-CA. 18. Optical AIC w/Impella Connect, Pack'd, EU, Product Code: 0042-0040-EU. 19. Optical, AIC, Impella Connect, Pkgd, AU, Product Code: 0042-0040-AU. 20. Optical, AIC, Impella Connect, Pkgd, UK, Product Code: 0042-0040-UK. 21. Optical, AIC, Impella Connect, Pkgd, US, Product Code: 0042-0040-US.

Abiomed

Class I - Dangerous

Potential for Automated Impella Controller (AIC) purge pressure issues due to purge retainer failures.

Sep 16, 2025 Other Medical Devices Nationwide View Details →

Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.

Aug 20, 2025 Infusion Pumps View Details →

Automated Impella Controller (AIC), used in left heart support blood pump, including the following: 1. Impella Controller, Packaged, AU; Product Code: 0042-0000-AU. 2. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 3. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 4. Impella Controller, Packaged, EU, Loaner; Product Code: 0042-0000-EU-L. 5. Impella Controller, Packaged, IN; Product Code: 0042-0000-IN. 6. Impella Controller, Packaged, JP; Product Code: 0042-0000-JP. 7. Impella Controller, Japan, Loaner; Product Code: 0042-0000-JP-L. 8. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 9. Impella Controller, UK, Loaner; Product Code: 0042-0000-UK-L. 10. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 11. Impella Controller, Packaged, US, Loaner; Product Code: 0042-0000-US-L. 12. Impella Optical Controller, Packaged, AU; Product Code: 0042-0010-AU. 13. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 14. Impella Optical Controller, EU, Loaner; Product Code: 0042-0010-EU-L. 15. Impella Optical Controller, Packaged, IN; Product Code: 0042-0010-IN. 16. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 17. Impella Optical Controller, UK, Loaner; Product Code: 0042-0010-UK-L. 18. Impella Optical Controller, Packaged, US; Product Code: 0042-0010-US. 19. Impella Optical Controller, US, Loaner; Product Code: 0042-0010-US-L. 20. Optical, AIC, Impella Connect, Packaged, AU; Product Code: 0042-0040-AU. 21. Optical AIC w/Impella Connect, Packaged, CA; Product Code: 0042-0040-CA. 22. AIC w/Remote Link, Packaged, Cln Trl, US; Product Code: 0042-0040-CT-L. 23. Optical AIC w/Impella Connect, Packaged, EU; Product Code: 0042-0040-EU. 24. Optical AIC w/Impella Connect, EU, Loaner; Product Code: 0042-0040-EU-L. 25. Optical AIC w/Impella Connect, Pkgd, JP; Product Code: 0042-0040-JP. 26. Optical AIC w/Impella Connect, Pkgd, JP; Product Code: 0042-0040-JP-L. 27. Optical, AIC, Impella Connect, Pkgd, UK; Product Code: 0042-0040-UK. 28. Optical, AIC, Impella Connect, Pkgd, US; Product Code: 0042-0040-US. 29. Optical AIC Impella Connect, US, Loaner; Product Code: 0042-0040-US-L. 30. Dbl optical, AIC Impella Connect, Phg US; Product Code: 1000201. 31. AIC w/Impella Connect for ECP; Product Code: 1000432.

Abiomed

Class I - Dangerous

A potential issue with the Automated Impella Controller (AIC) not detecting an Impella pump when it is connected

Jun 23, 2025 Infusion Pumps Nationwide View Details →