NUVASIVE SPECIALIZED ORTHOPEDICS, INC. Precice Bone Transport Antegrade Femur Trochanter 10 Degree Bend Component of the orthopedic Precice Bone Transport System of intramedullary nail, locking screws, reusable instruments, and hand-held External Remote Controller (ERC).

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Item Number: PA1160-10D400-7 Description: ANTE FEMUR TROCH 10DEG11.5MM X 400MM 7/10CM ST Lot/Serial Number: 9011815AAA 9021510AAA 9022225AAA 9072512AAA Item Number: PA1161-10D400-7 Description: ANTE FEMUR TROCH 10DEG13MM X 400MM, 7/10CM ST Lot/Serial Number: 9012506AAA 9061720AAA Item Number: PA1160-10SJ360-7 Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 360MM 7/10CM ST Lot/Serial Number: 9013007AAA 9030706AAA 9031514AAA 9042208AAA 9050811AAA 9072218AAA Item Number: PA1159-10SJ360-7 Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 360MM, 7/10CM STK Lot/Serial Number: 9020607AAA 9022815AAA 9050211AAA 9071604AAA Item Number: PA1160-90SJ340-7 Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 340MM, 7/9CM ST Lot/Serial Number: 9020608AAA 9052110AAA 9052906AAA 9072217AAA Item Number: PA1160-80SJ320-7 Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 320MM, 7/8CM ST Lot/Serial Number: 9021222AAA Item Number: PA1159-80SJ320-7 Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 320MM, 7/8CM STK Lot/Serial Number: 9021223AAA Item Number: PA1160-70SJ300-7 Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 300MM, 7/7CM ST Lot/Serial Number: 9021224AAA Item Number: PA1160-10D380-7 Description: ANTE FEMUR TROCH 10DEG11.5MM X 380MM 7/10CM ST Lot/Serial Number: 9021512AAA 9072513AAA 9072912AAA 9080106AAA Item Number: PA1159-90SJ340-7 Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 340MM, 7/9CM STK Lot/Serial Number: 9020609AAA 9052208AAA 9071603AAA Item Number: PA1161-10X380-7 Description: RETRO FEMUR STRAIGHT13MM X 380MM, 7/10CM ST Lot/Serial Number: 9021913AAA Item Number: PA1160-10X360-7 Description: RETRO FEMUR STRAIGHT11.5MM X 360MM, 7/10CM ST Lot/Serial Number: 9022227AAA Item Number: PA1161-10X360-7 Description: RETRO FEMUR STRAIGHT13MM X 360MM, 7/10CM ST Lot/Serial Number: 9022226AAA Item Number: PA1160-10SJ400-7 Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 400MM 7/10CM ST Lot/Serial Number: 9020809AAA Item Number: PA1159-10SJ400-7 Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 400MM, 7/10CM STK Lot/Serial Number: 9020810AAA Item Number: PA1160-10SJ380-7 Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 380MM 7/10CM ST Lot/Serial Number: 9022813AAA 9042506AAA Item Number: PA1160-10B380-7 Description: ANTE FEMUR PIRIFOMIS11.5MM X 380MM 7/10CM ST Lot/Serial Number: 9021513AAA Item Number: PA1160-10B360-7 Description: ANTE FEMUR PIRIFOMIS11.5MM X 360MM 7/10CM ST Lot/Serial Number: 9021514AAA Item Number: PA1159-10SJ380-7 Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 380MM, 7/10CM STK Lot/Serial Number: 9031310AAA 9071605AAA Item Number: PA1160-10D360-7 Description: ANTE FEMUR TROCH 10DEG11.5MM X 360MM 7/10CM ST Lot/Serial Number: 9022816AAA Item Number: PA1160-60SJ280-6 Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 280MM, 6/6CM ST Lot/Serial Number: 9021914AAA Item Number: PA1159-70SJ300-7 Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 300MM, 7/7CM STK Lot/Serial Number: 9021915AAA Item Number: PA1160-90X340-7 Description: RETRO FEMUR STRAIGHT11.5MM X 340MM, 7/9CM ST Lot/Serial Number: 9041617AAA Item Number: PA1160-10X380-7 Description: RETRO FEMUR STRAIGHT11.5MM X 380MM 7/10CM ST Lot/Serial Number: 9041618AAA Item Number: PA1159-10D380-7 Description: ANTE FEMUR TROCH 10DEG10MM X 380MM, 7/10CM STK Lot/Serial Number: 9042211AAA Item Number: PA1161-10B380-7 Description: ANTE FEMUR PIRIFOMIS13MM X 380MM, 7/10CM ST Lot/Serial Number: 9052305AAA 9061009AAA Item Number: PA1161-10B400-7 Description: ANTE FEMUR PIRIFOMIS13MM X 400MM, 7/10CM ST Lot/Serial Number: 9052306AAA Item Number: PA1160-10B400-7 Description: ANTE FEMUR PIRIFOMIS11.5MM X 400MM 7/10CM ST Lot/Serial Number: 9053110AAA Item Number: PA1160-10X400-7 Description: RETRO FEMUR STRAIGHT11.5MM X 400MM 7/10CM ST Lot/Serial Number: 9061208AAA Item Number: PA1161-10X400-7 Description: RETRO FEMUR STRAIGHT13MM X 400MM, 7/10CM ST Lot/Serial Number: 9061209AAA Item Number: PA1161-10B400-7 Description: ANTE FEMUR PIRIFOMIS13MM X 400MM, 7/10CM ST Lot/Serial Number: 9061008AAA Item Number: PA1159-80X320-7 Description: RETO FEMUR STRAIGHT10MM X 320MM, 7/8CM STK Lot/Serial Number: 9062416AAA Item Number: PA1161-10D380-7 Description: ANTE FEMUR TROCH 10DEG13MM X 380MM, 7/10CM ST Lot/Serial Number: 9061721AAA 9071825AAA Item Number: PA1159-10B400-7 Description: ANTE FEMUR PIRIFORMIS10MM X 400MM, 7/10CM STK Lot/Serial Number: 9072219AAA Item Number: PA1159-10D400-7 Description: ANTE FEMUR TROCH 10DEG10MM X 400MM, 7/10CM STK Lot/Serial Number: 9072220AAA 9072911AAA Item Number: PA1159-10B380-7 Description: ANTE FEMUR PIRIFORMIS10MM X 380MM, 7/10CM STK Lot/Serial Number: 9072221AAA
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Nuvasive Specialized Orthopedics Inc
Reason for Recall:
During explantation, standard removal tools can generate sufficient forces to cause distal plug to disassociate from the bone transport system.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. Precice Bone Transport Antegrade Femur Trochanter 10 Degree Bend Component of the orthopedic Precice Bone Transport System of intramedullary nail, locking screws, reusable instruments, and hand-held External Remote Controller (ERC).

Product Codes/Lot Numbers:

Item Number: PA1160-10D400-7 Description: ANTE FEMUR TROCH 10DEG11.5MM X 400MM 7/10CM ST Lot/Serial Number: 9011815AAA 9021510AAA 9022225AAA 9072512AAA Item Number: PA1161-10D400-7 Description: ANTE FEMUR TROCH 10DEG13MM X 400MM, 7/10CM ST Lot/Serial Number: 9012506AAA 9061720AAA Item Number: PA1160-10SJ360-7 Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 360MM 7/10CM ST Lot/Serial Number: 9013007AAA 9030706AAA 9031514AAA 9042208AAA 9050811AAA 9072218AAA Item Number: PA1159-10SJ360-7 Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 360MM, 7/10CM STK Lot/Serial Number: 9020607AAA 9022815AAA 9050211AAA 9071604AAA Item Number: PA1160-90SJ340-7 Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 340MM, 7/9CM ST Lot/Serial Number: 9020608AAA 9052110AAA 9052906AAA 9072217AAA Item Number: PA1160-80SJ320-7 Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 320MM, 7/8CM ST Lot/Serial Number: 9021222AAA Item Number: PA1159-80SJ320-7 Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 320MM, 7/8CM STK Lot/Serial Number: 9021223AAA Item Number: PA1160-70SJ300-7 Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 300MM, 7/7CM ST Lot/Serial Number: 9021224AAA Item Number: PA1160-10D380-7 Description: ANTE FEMUR TROCH 10DEG11.5MM X 380MM 7/10CM ST Lot/Serial Number: 9021512AAA 9072513AAA 9072912AAA 9080106AAA Item Number: PA1159-90SJ340-7 Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 340MM, 7/9CM STK Lot/Serial Number: 9020609AAA 9052208AAA 9071603AAA Item Number: PA1161-10X380-7 Description: RETRO FEMUR STRAIGHT13MM X 380MM, 7/10CM ST Lot/Serial Number: 9021913AAA Item Number: PA1160-10X360-7 Description: RETRO FEMUR STRAIGHT11.5MM X 360MM, 7/10CM ST Lot/Serial Number: 9022227AAA Item Number: PA1161-10X360-7 Description: RETRO FEMUR STRAIGHT13MM X 360MM, 7/10CM ST Lot/Serial Number: 9022226AAA Item Number: PA1160-10SJ400-7 Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 400MM 7/10CM ST Lot/Serial Number: 9020809AAA Item Number: PA1159-10SJ400-7 Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 400MM, 7/10CM STK Lot/Serial Number: 9020810AAA Item Number: PA1160-10SJ380-7 Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 380MM 7/10CM ST Lot/Serial Number: 9022813AAA 9042506AAA Item Number: PA1160-10B380-7 Description: ANTE FEMUR PIRIFOMIS11.5MM X 380MM 7/10CM ST Lot/Serial Number: 9021513AAA Item Number: PA1160-10B360-7 Description: ANTE FEMUR PIRIFOMIS11.5MM X 360MM 7/10CM ST Lot/Serial Number: 9021514AAA Item Number: PA1159-10SJ380-7 Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 380MM, 7/10CM STK Lot/Serial Number: 9031310AAA 9071605AAA Item Number: PA1160-10D360-7 Description: ANTE FEMUR TROCH 10DEG11.5MM X 360MM 7/10CM ST Lot/Serial Number: 9022816AAA Item Number: PA1160-60SJ280-6 Description: ANTE TIBIA 10DEG, 3-HOLE11.5MM X 280MM, 6/6CM ST Lot/Serial Number: 9021914AAA Item Number: PA1159-70SJ300-7 Description: ANTE TIBIA 10DEG, 3-HOLE10MM X 300MM, 7/7CM STK Lot/Serial Number: 9021915AAA Item Number: PA1160-90X340-7 Description: RETRO FEMUR STRAIGHT11.5MM X 340MM, 7/9CM ST Lot/Serial Number: 9041617AAA Item Number: PA1160-10X380-7 Description: RETRO FEMUR STRAIGHT11.5MM X 380MM 7/10CM ST Lot/Serial Number: 9041618AAA Item Number: PA1159-10D380-7 Description: ANTE FEMUR TROCH 10DEG10MM X 380MM, 7/10CM STK Lot/Serial Number: 9042211AAA Item Number: PA1161-10B380-7 Description: ANTE FEMUR PIRIFOMIS13MM X 380MM, 7/10CM ST Lot/Serial Number: 9052305AAA 9061009AAA Item Number: PA1161-10B400-7 Description: ANTE FEMUR PIRIFOMIS13MM X 400MM, 7/10CM ST Lot/Serial Number: 9052306AAA Item Number: PA1160-10B400-7 Description: ANTE FEMUR PIRIFOMIS11.5MM X 400MM 7/10CM ST Lot/Serial Number: 9053110AAA Item Number: PA1160-10X400-7 Description: RETRO FEMUR STRAIGHT11.5MM X 400MM 7/10CM ST Lot/Serial Number: 9061208AAA Item Number: PA1161-10X400-7 Description: RETRO FEMUR STRAIGHT13MM X 400MM, 7/10CM ST Lot/Serial Number: 9061209AAA Item Number: PA1161-10B400-7 Description: ANTE FEMUR PIRIFOMIS13MM X 400MM, 7/10CM ST Lot/Serial Number: 9061008AAA Item Number: PA1159-80X320-7 Description: RETO FEMUR STRAIGHT10MM X 320MM, 7/8CM STK Lot/Serial Number: 9062416AAA Item Number: PA1161-10D380-7 Description: ANTE FEMUR TROCH 10DEG13MM X 380MM, 7/10CM ST Lot/Serial Number: 9061721AAA 9071825AAA Item Number: PA1159-10B400-7 Description: ANTE FEMUR PIRIFORMIS10MM X 400MM, 7/10CM STK Lot/Serial Number: 9072219AAA Item Number: PA1159-10D400-7 Description: ANTE FEMUR TROCH 10DEG10MM X 400MM, 7/10CM STK Lot/Serial Number: 9072220AAA 9072911AAA Item Number: PA1159-10B380-7 Description: ANTE FEMUR PIRIFORMIS10MM X 380MM, 7/10CM STK Lot/Serial Number: 9072221AAA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0069-2021

Related Recalls

Precice Max REF: PM10.0-80B265 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 265mm PM10.0-80B280 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 280mm PM10.0-80B305 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 305mm PM10.0-80B335 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 335mm PM10.0-80B365 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 365mm PM10.0-80TJ265 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 10.0mm x 265mm PM10.0-80TJ280 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 10.0mm x 280mm PM10.0-80TJ305 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 10.0mm x 305mm PM11.5-80B265 PRECICE MAX Antegrade Femur Piriformis Straight 11.5mm x 265mm PM11.5-80B280 PRECICE MAX Antegrade Femur Piriformis Straight 11.5mm x 280mm PM11.5-80B305 PRECICE MAX Antegrade Femur Piriformis Straight 11.5mm x 305mm PM11.5-80B335 PRECICE MAX Antegrade Femur Piriformis Straight 11.5mm x 335mm PM11.5-80B365 PRECICE MAX Antegrade Femur Piriformis Straight 11.5mm x 365mm PM11.5-80TJ265 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 11.5mm x 265mm PM11.5-80TJ280 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 11.5mm x 280mm PM11.5-80TJ305 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 11.5mm x 305mm PM13.0-80B280 PRECICE MAX Antegrade Femur Piriformis Straight 13.0mm x 280mm PM13.0-80B305 PRECICE MAX Antegrade Femur Piriformis Straight 13.0mm x 305mm PM13.0-80B365 PRECICE MAX Antegrade Femur Piriformis Straight 13.0mm x 365mm The Precice Max Nail is intended to stabilize fractures, nonunions, and malunions of the femur, tibia, and humerus. In addition, the Precice Nail is intended to provide magnetically controlled lengthening of long bones and/or transporting of bone segments in the femur, tibia, and humerus.

NuVasive Specialized Orthopedics

Class I - Dangerous

Due to potential issues with design, manufacturing and documentation issues related to the validation of the affected products/devices.

Sep 30, 2024 Other Medical Devices View Details →

During accelerated shelf-life extension studies it was noted that the final packaging was experiencing a potential of failure of the sterile packaging barrier which could result in a hazardous situation where the device is no longer sterile.

Nov 1, 2022 Infusion Pumps Nationwide View Details →

MAGEC 1 System

NuVasive Specialized Orthopedics

Class I - Dangerous

Safety Alert to all surgeons informing them of post-implantation fractures of an internal metallic component for device manufactured prior to March 26, 2015.

Jun 25, 2019 Other Medical Devices View Details →