THERMOCOOL SMARTTOUCH Bi-Directional Navigation Catheter (D-132704-S, D-132705-S)
Class I - DangerousWhat Should You Do?
- Check if you have this product: 17445775M 17694994M 17697951M 17685286M 17697907M 17697991M 17691864M 17697910M 17697994M
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Biosense Webster, Inc.
- Reason for Recall:
- Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
THERMOCOOL SMARTTOUCH Bi-Directional Navigation Catheter (D-132704-S, D-132705-S)
Product Codes/Lot Numbers:
17445775M 17694994M 17697951M 17685286M 17697907M 17697991M 17691864M 17697910M 17697994M
Distribution:
Distributed in: US
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0056-2018
Related Recalls
Due to an observed trend of neurovascular events
Due to a manufacturing issue, device under process validation phase were inadvertently mixed into the main manufacturing process.
Biosense Webster Carto Vizigo Bi-Directional Guiding Sheath, 8.5F(Small), REF D-1385-01-S, RX CE 2797, Sterile
Biosense Webster
Incorrect labeling; specifically, the product name, lot number and product code on the carton label for Bi-Directional Guiding Sheaths does not match the inner label pouch