EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model No. BF-H190; UDI: 4953170335174; All Serial No.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Olympus Corporation of the Americas
Reason for Recall:
Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190

Product Codes/Lot Numbers:

Model No. BF-H190; UDI: 4953170335174; All Serial No.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0053-2026

Related Recalls

Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.

Oct 14, 2025 Diagnostic Equipment Nationwide View Details →

Device kits contained an incorrectly sized 4mm x 4cm urinary tract balloon catheter (BURS0404) instead of the correct 8mm x 15cm nephrostomy tract balloon catheter (BPCN0815). All other components in the kits were correctly included and packaged in the kit.

Oct 17, 2025 Surgical Instruments View Details →