DABRA RA-308 Excimer Laser, SN: ******* - Product Usage: The device is intended for use in ablating a channel in occlusive peripheral vascular disease.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Number DABRA Excimer Laser RA-308; Serial Numbers: RA00027, RA00039, RA00046, RA00049, RA00050, RA00054, RA00055, RA00056, RA00057, RA00058, RA00063, RA00064, RA00065, RA00067, RA00069, RA00070, RA00073, RA00074, RA00081, RA00082, RA00083, RA00084, RA00085, RA00086, RA00088, RA00089, RA00097, RA00098, RA00103, RA00114, RA00115, RA00117, RA00118, RA00120, RA00121, RA00124, RA00125, RA00127, RA00135, RA00136, RA00137, RA00138, RA00143, RA00147, RA00148, RA00153, RA00154, RA00160, RA00164, RA00165, RA00167, RA00170, RA00175, and RA00177.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Ra Medical Systems, Inc.
- Reason for Recall:
- Due to the taller wheel mounting on the laser, the laser may fall over potentially causing user or patient injury.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
DABRA RA-308 Excimer Laser, SN: ******* - Product Usage: The device is intended for use in ablating a channel in occlusive peripheral vascular disease.
Product Codes/Lot Numbers:
Model Number DABRA Excimer Laser RA-308; Serial Numbers: RA00027, RA00039, RA00046, RA00049, RA00050, RA00054, RA00055, RA00056, RA00057, RA00058, RA00063, RA00064, RA00065, RA00067, RA00069, RA00070, RA00073, RA00074, RA00081, RA00082, RA00083, RA00084, RA00085, RA00086, RA00088, RA00089, RA00097, RA00098, RA00103, RA00114, RA00115, RA00117, RA00118, RA00120, RA00121, RA00124, RA00125, RA00127, RA00135, RA00136, RA00137, RA00138, RA00143, RA00147, RA00148, RA00153, RA00154, RA00160, RA00164, RA00165, RA00167, RA00170, RA00175, and RA00177.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0050-2021
Related Recalls
Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.
A software issue was identified which could result in user or patient injury, or may adversely impact laser performance. It is possible that an unintended release of laser light (radiation) may occur, injuring the user or patient.
The firm has become aware there is a potential problem with its DABRA catheters resulting in higher catheter calibration failure rate due to the product shelf life is currently less than the labeled shelf life.