Sysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot Numbers: A6051 A6052 A6053 A6054 A6055 A6056 A6057 A6058 A6059 A6060 A6061 A6062 A6063 A6064 A6065 A6066 A6067 A6068 A6069 A6070 A6071 A6072 A6073 A6074 A6075 A6076 A6077 A6078 A6079 A6080 A6081 A6082 A6083 A6084 A6085 A6086 A6087 A6088 A6089 A6090
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Sysmex America, Inc.
Reason for Recall:
There is potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%, #) results obtained using the lots listed above may exhibit lower than actual PLT-F and IPF values. The issue may be recognized by a large discrepancy between the impedance platelet (PLT-I) result and the PLT-F result. In most cases, the falsely decreased PLT-F results display a PLT Abn Scattergram message with an asterisk beside the result or an Action Message: Difference between PLT and PLT- F. Check the results. These are indicators of unreliable data. In rare instances, platelet flags may be absent. The PLT-I results are unaffected.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
completed

Product Information

Full Description:

Sysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers

Product Codes/Lot Numbers:

Lot Numbers: A6051 A6052 A6053 A6054 A6055 A6056 A6057 A6058 A6059 A6060 A6061 A6062 A6063 A6064 A6065 A6066 A6067 A6068 A6069 A6070 A6071 A6072 A6073 A6074 A6075 A6076 A6077 A6078 A6079 A6080 A6081 A6082 A6083 A6084 A6085 A6086 A6087 A6088 A6089 A6090

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0045-2018

Related Recalls

Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.

Jan 6, 2025 Other Medical Devices Nationwide View Details →