4500MD Triple Quad Mass Spectrometer, Part Number 5032522
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 00842365102049; Serial Numbers: EJ230051806PL, EJ230061807PL, EJ230071808, EJ230101810, EJ230111811, EJ230121811, EJ230141811, EJ230151812, EJ230161812, EJ230171901, EJ230181901, EJ230191901, EJ230211905, EJ230221908, EJ230231908, EJ230251909, EJ230271910, EJ230281910, EJ230311911, EJ230331912, EJ230362003, EJ230372003, EJ230392004, EJ230412009, EJ230422011, EJ230442012, EJ230452012, EJ230472101, EJ230482103, EJ230512105, EJ240012110, EJ240022111, EJ240052111, EJ240082207, EJ240092209, EJ240142307, EJ240132305, EJ240062201
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- AB Sciex, LLC
- Reason for Recall:
- In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
4500MD Triple Quad Mass Spectrometer, Part Number 5032522
Product Codes/Lot Numbers:
UDI-DI: 00842365102049; Serial Numbers: EJ230051806PL, EJ230061807PL, EJ230071808, EJ230101810, EJ230111811, EJ230121811, EJ230141811, EJ230151812, EJ230161812, EJ230171901, EJ230181901, EJ230191901, EJ230211905, EJ230221908, EJ230231908, EJ230251909, EJ230271910, EJ230281910, EJ230311911, EJ230331912, EJ230362003, EJ230372003, EJ230392004, EJ230412009, EJ230422011, EJ230442012, EJ230452012, EJ230472101, EJ230482103, EJ230512105, EJ240012110, EJ240022111, EJ240052111, EJ240082207, EJ240092209, EJ240142307, EJ240132305, EJ240062201
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0038-2024
Related Recalls
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.