Xoran MiniCAT CT Scanner
Class I - DangerousWhat Should You Do?
- Check if you have this product: Minicat
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Xoran Technologies, Inc.
- Reason for Recall:
- A description of the defect in the product or the manner in which the product fails to comply wvith an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evalu
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Xoran MiniCAT CT Scanner
Product Codes/Lot Numbers:
Minicat
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0037-2014
Related Recalls
Xoran MiniCAT X-ray imaging device
Xoran Technologies
Out of tolerance kVp, Unexposed (black) frames, and Out of tolerance Dose.
Xoran Minicat00158
Xoran Technologies
The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evaluation of the hazards reasonably related to defect or the failure to comply with the Federal standard; An evaluation of the health ha
Xoran MiniCAT00202
Xoran Technologies
The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xorans specifications.