Ortho OS Solutions UltOS 3.5mm locking screw 12mm to 50mm Ortho Solutions LTD, Unit 5 West Station Business Park, Spital Road, Maldon, Essex CM96FF Rx Only a) Sterile b) Non-Sterile Orthopaedic Bone Screw
Class I - DangerousWhat Should You Do?
- Check if you have this product: Part Number (Lot Number) infomration is as follows: OS421712-NS (lot 7891), OS421714-NS (lot 7891) OS421716-NS (lot 7891, 7520), OS421718-NS (lot 7891, 7520) OS421720-NS (lot 7891), OS421722-NS (lot 7891, 7520) OS421724-NS (lot 7891, 7520), OS421726-NS (lot 7891) OS421728-NS (lot 7891), OS421730-NS (lot 7891) OS421735-NS (lot 7520, 7891) ,OS421740-NS (lot 7520, 7891) OS421745-NS (lot 7520, 7891) ,OS421750-NS (lot 7520, 7891) OS421716 (lot 7519) ,OS421718 (lot 7519)
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Ortho Solutions Inc
- Reason for Recall:
- Ortho Solutions Inc is initiating a recall on behalf of UltOS 3.5mm locking screw. The locking thread feature of the screw was not manufactured to specification. As a consequence there is a risk that a screw may not lock onto the plate at the time of primary insertion into the plate and the bone. In the event of this occurring the non functioning screw would need to be replaced with an alternativ
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Ortho OS Solutions UltOS 3.5mm locking screw 12mm to 50mm Ortho Solutions LTD, Unit 5 West Station Business Park, Spital Road, Maldon, Essex CM96FF Rx Only a) Sterile b) Non-Sterile Orthopaedic Bone Screw
Product Codes/Lot Numbers:
Part Number (Lot Number) infomration is as follows: OS421712-NS (lot 7891), OS421714-NS (lot 7891) OS421716-NS (lot 7891, 7520), OS421718-NS (lot 7891, 7520) OS421720-NS (lot 7891), OS421722-NS (lot 7891, 7520) OS421724-NS (lot 7891, 7520), OS421726-NS (lot 7891) OS421728-NS (lot 7891), OS421730-NS (lot 7891) OS421735-NS (lot 7520, 7891) ,OS421740-NS (lot 7520, 7891) OS421745-NS (lot 7520, 7891) ,OS421750-NS (lot 7520, 7891) OS421716 (lot 7519) ,OS421718 (lot 7519)
Distribution:
Distributed in: US, CA, CO, GA, IL, LA, NC, NY, TX, WA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0026-2016
Related Recalls
Following the identification of corrosion residue found inside the cannulated devices in Forefoot and Midfoot trays Orhto Solutions Group has taken the decision to recall ALL x106 System 26 Forefoot and Midfoot Implants and Instruments tray from the US market.
There is a material mislabeling of the AFN End Cap.
FPS 35mm x 3.5 mm non-locking screw
Ortho Solutions
Mislabeling of a single batch of FPS 3.5 x 35 mm Locking Screw as being Non-Locking Screws.