Hammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: 00810633022047 Lot Numbers: MHC220119, MHC220120 Expiry Date: 1 June 2027
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

โš ๏ธ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Synthes (USA) Products LLC
Reason for Recall:
Drill Pin (drill bit) is not compatible with the Drill Template included in the kit due to a packaging error, result in a surgical delay and may cause soft tissue damage if a larger incision is required to accommodate an alternative form of fixation. Additionally, there is a potential for adverse tissue reaction if the surgeon attempts to use the guide regardless of tight fit and plastic debris is created and ends up in the wound
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Hammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001

Product Codes/Lot Numbers:

UDI-DI: 00810633022047 Lot Numbers: MHC220119, MHC220120 Expiry Date: 1 June 2027

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0008-2023

Related Recalls

Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.

Sep 26, 2025 Other Medical Devices Nationwide View Details โ†’

Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.

Sep 26, 2025 Other Medical Devices Nationwide View Details โ†’