ENVOY 500 AST Reagent Kit, reference Model/Catalog 55255, for Envoy 500. Intended Use of the Product: Envoy 500 AST Reagent is for the quantitative in vitro diagnostic determination of aspartate aminotransferase (AST) in human serum and plasma using the Envoy 500 Chemistry System. Aspartate aminotransferase (AST) measurements are used in the diagnosis and treatment of certain types of liver and heart diseases.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot 4014, Expiration date 07/2015; Lot 4051, Expiration date 09/2015; Lot 4074, Expiration date 11/2015;
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
ELITech Clinical Systems SAS
Reason for Recall:
Some users of ENVOY 500 AST Reagent Kit, reference 55255, for Envoy 500 systems are observing that Quality Control (QC) fails before the product expiry is reached.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

ENVOY 500 AST Reagent Kit, reference Model/Catalog 55255, for Envoy 500. Intended Use of the Product: Envoy 500 AST Reagent is for the quantitative in vitro diagnostic determination of aspartate aminotransferase (AST) in human serum and plasma using the Envoy 500 Chemistry System. Aspartate aminotransferase (AST) measurements are used in the diagnosis and treatment of certain types of liver and heart diseases.

Product Codes/Lot Numbers:

Lot 4014, Expiration date 07/2015; Lot 4051, Expiration date 09/2015; Lot 4074, Expiration date 11/2015;

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0006-2015

Related Recalls

ELITech Clinical Systems SAS initiated this recall because some users of ENVOY 500 ISE CALIBRATOR Kit (Part number: 55117), for Envoy 500 systems are observing trouble while calibrating with some vials. Users contacted the firm when they received the messages "ISE OUT OF REPRODUCIBILITY" or "ISE SLOPE OUT OF RANGE" (i.e., failed calibration). This has resulted in delay in obtaining patient results until the calibrator lot is replaced.

Apr 18, 2017 Other Medical Devices View Details →

ELITech Clinical Systems reagents based on Trinder reaction used on the Selectra ProS and ProM systems are recalled due to potential drug interference. No complaints were received.

Sep 22, 2015 Other Medical Devices Nationwide View Details →

ELITech clinical Systems reagents based on Trinder reaction used on the Eon 100 instrument system are recalled due to potential drug interference. No complaints were received.

Sep 22, 2015 Other Medical Devices Nationwide View Details →