Carl Zeiss INTRABEAM PRS 500, product #304534-0000-500. The firm name on the label is Carl Zeiss Meditec AG, Jena, Germany. The INTRABEAM System is indicated for radiation therapy treatments
Class I - DangerousWhat Should You Do?
- Check if you have this product: Intrabeam ser. #6402501246, Ion chamber ser. #SN001254; Intrabeam ser. #6402501241, Ion chamber ser. #SN001745; Intrabeam ser. #6402100932, Ion chamber ser. #SN001994; Intrabeam ser. #6402100958, Ion chamber ser. #SN002096; Intrabeam ser. #6402101042, Ion chamber ser. #SN002265; Intrabeam ser. #6402101029, Ion chamber ser. #SN002270; Intrabeam ser. #6402101034, Ion chamber ser. #SN002302; Intrabeam ser. #6402101105, Ion chamber ser. #SN002651; and Intrabeam ser. #6402101130, Ion chamber ser. #SN002684.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Carl Zeiss Meditec AG
- Reason for Recall:
- The device had an incorrect printed calibration value on the calibration certificate.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Carl Zeiss INTRABEAM PRS 500, product #304534-0000-500. The firm name on the label is Carl Zeiss Meditec AG, Jena, Germany. The INTRABEAM System is indicated for radiation therapy treatments
Product Codes/Lot Numbers:
Intrabeam ser. #6402501246, Ion chamber ser. #SN001254; Intrabeam ser. #6402501241, Ion chamber ser. #SN001745; Intrabeam ser. #6402100932, Ion chamber ser. #SN001994; Intrabeam ser. #6402100958, Ion chamber ser. #SN002096; Intrabeam ser. #6402101042, Ion chamber ser. #SN002265; Intrabeam ser. #6402101029, Ion chamber ser. #SN002270; Intrabeam ser. #6402101034, Ion chamber ser. #SN002302; Intrabeam ser. #6402101105, Ion chamber ser. #SN002651; and Intrabeam ser. #6402101130, Ion chamber ser. #SN002684.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0001-2018
Related Recalls
ZEISS miLOOP Lens Fragmentation Device REF 303071-9090-000. Used during cataract surgery.
Carl Zeiss Meditec AG
High friction of the slider can cause the device to stick, or not move as intended.
IOLMaster 700
Carl Zeiss Meditec AG
When using software 1.90.2.09 or 1.90.8.06 and using modality worklist functionality for patient data transfer, the selection of the patient in the patient list may not match the patient information displayed on the right side of the screen
Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.