Grilled Vegetables PLU# 23657100000. Expiration date: 8/6/2020 Ingredients: GRILLED VEGETABLES [RED ONION, GREEN BELL PEPPER, YELLOW SQUASH, ZUCCHINI, ASPARAGUS, EGGPLANT, RED BELL PEPPER, YELLOW BELL PEPPER, OIL BLEND (CANOLA OIL, EXTRA VIRGIN OLIVE OIL), SALT, BLACK PEPPER]. Onions are cooked
Class I - DangerousWhat Should You Do?
- Check if you have this product: PLU#23657100000
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Giant Eagle
- Reason for Recall:
- Potential contamination with Salmonella Newport in onions and prepared foods containing onions.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Grilled Vegetables PLU# 23657100000. Expiration date: 8/6/2020 Ingredients: GRILLED VEGETABLES [RED ONION, GREEN BELL PEPPER, YELLOW SQUASH, ZUCCHINI, ASPARAGUS, EGGPLANT, RED BELL PEPPER, YELLOW BELL PEPPER, OIL BLEND (CANOLA OIL, EXTRA VIRGIN OLIVE OIL), SALT, BLACK PEPPER]. Onions are cooked
Product Codes/Lot Numbers:
PLU#23657100000
Distribution:
Distributed in: IN, OH, MD, WV, PA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: F-1383-2020
Related Recalls
Deep-brand PREMIUM Select Mixed Vegetables 4 way KEEP FROZEN a) Net Wt. 3.85 LB. (1.75 kg); b) Net Wt. 2 LB (907 g)
CHETAK NEW YORK
Product tested positive Salmonella .
Potential contamination with Listeria monocytogenes.
Potential contamination with Listeria monocytogenes.