Evergreen Farm Pear Juice Gochujang (Hot Peppeer Paste). Net Wt 70 ounce (4.4 LBS) glass jars. 1023 Yardville Allentown Road, Hamilton NJ 08620. Product of USA. UPC # 89076726177.
Class I - DangerousWhat Should You Do?
- Check if you have this product: N/A
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Evergreen Orchard Farm, LLC
- Reason for Recall:
- Juice and paste products display incorrect labeling, product descriptions and formulations.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Evergreen Farm Pear Juice Gochujang (Hot Peppeer Paste). Net Wt 70 ounce (4.4 LBS) glass jars. 1023 Yardville Allentown Road, Hamilton NJ 08620. Product of USA. UPC # 89076726177.
Product Codes/Lot Numbers:
N/A
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: F-1344-2012
Related Recalls
Class I - Dangerous
Undeclared caffeine
Halved Pears IN PEAR JUICE FROM CONCNETRATE; NET WT 15 OZ (425 g); Ingredients: Pears, Water, Pear Juice Concentrate; Distributed By: Wegmans Food Markets, Inc. Rochester, NY 14603
Pacific Coast Producers
Class I - Dangerous
Potential contamination with heavy metal (Lead).
Class I - Dangerous
Potential Listeria monocytogenes contamination.