LaBri's Body Health Atomic, 60 capsules
Class I - DangerousWhat Should You Do?
- Check if you have this product: All lot codes, manufacturing codes and expiration dates.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Envy Me
- Reason for Recall:
- Marketed without an approved NDA/ANDA: Product contains undeclared Sibutramine.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
LaBri's Body Health Atomic, 60 capsules
Product Codes/Lot Numbers:
All lot codes, manufacturing codes and expiration dates.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: D-0867-2017