dilTIAZem HCl 125 mg per 125 mL (1 mg per mL) in 0.7% Sodium Chloride Injection, Rx only, SterRx, 141 Idaho Ave., Plattsburgh, NY, 12903, NDC 70324-976-01.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lots S21164/BTV, 3-Dec-21; S21191/BUU, 18-DEC-21; S21197/BUZ, 31-Dec-21; S21220/BVV, 13-Jan-22; S21271/BX0, 27-Jan-22; S21272/BXP, 28-Jan-22; S21299/BYP, 18-Fev-22; S21300/BYQ, 19-Feb-22
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- SterRx, LLC
- Reason for Recall:
- Lack of Assurance of Sterility
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
dilTIAZem HCl 125 mg per 125 mL (1 mg per mL) in 0.7% Sodium Chloride Injection, Rx only, SterRx, 141 Idaho Ave., Plattsburgh, NY, 12903, NDC 70324-976-01.
Product Codes/Lot Numbers:
Lots S21164/BTV, 3-Dec-21; S21191/BUU, 18-DEC-21; S21197/BUZ, 31-Dec-21; S21220/BVV, 13-Jan-22; S21271/BX0, 27-Jan-22; S21272/BXP, 28-Jan-22; S21299/BYP, 18-Fev-22; S21300/BYQ, 19-Feb-22
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: D-0302-2022
Related Recalls
Lack of Assurance of Sterility
Lack of Assurance of Sterility
Lack of Assurance of Sterility