Euphoric Premium Male Performance Enhancer capsules, packaged in a)1-count blister cards, UPC 6 96859 28646 9, b) 3-count bottle, UPC 6 96859 28647 6 and c) 12-count bottle, UPC 6 96859 28648 3.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All lots
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Epic Products, LLC
- Reason for Recall:
- Marketed Without An Approved NDA/ANDA: product contains undeclared sildenafil, tadalafil, and oxytetracycline, FDA approved drug products making Euphoric an unapproved drug.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Euphoric Premium Male Performance Enhancer capsules, packaged in a)1-count blister cards, UPC 6 96859 28646 9, b) 3-count bottle, UPC 6 96859 28647 6 and c) 12-count bottle, UPC 6 96859 28648 3.
Product Codes/Lot Numbers:
All lots
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: D-0288-2019