Ondal AC2000 (PDI) Television Mounting System

Class II - Moderate

What Should You Do?

  1. Check if you have this product:
    (215 distributed to 5 hospitals)
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Report problems: Report any issues to the FDA's Safety Reporting Portal.

Recall Details

Company:
Unknown Manufacturer
Reason for Recall:
The arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.
Classification:
Class II - Moderate

Products that might cause a temporary health problem, or pose a slight threat of a serious nature.

Status:
ongoing
Hazard:
The arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.
Remedy:
Secure the unit to prevent movement and contact TRUMPF Medical Systems for information on receiving a repair kit that includes the missing washers.

Product Images

Ondal AC2000 (PDI) Television Mounting System

Product Information

Full Description:

The mounting systems were sold for use in hospital Intensive Care Units. The units are off-white and have multiple joints between the ceiling mount and the TV mount at the end of the arm system.

Product Codes/Lot Numbers:

(215 distributed to 5 hospitals)

Official Source

Always verify recall information with the official CPSC source:

View on CPSC.gov

CPSC Recall Number: 12742

Related Recalls

Class I - Dangerous

The recalled air rifles can fire unexpectedly when releasing the bolt before the air rifle is fully cocked, posing a risk of serious injury or death from projectile hazard.

Oct 15, 2026 Clothing & Accessories Nationwide View Details β†’