The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;
Globus Medical
Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.
Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 04U42-84 Software Version: All software versions Product Description: CELL-DYN Ruby
Abbott Laboratories
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
A specific lot of Latitude Humeral Trial Stems has the incorrect direct part marking. " Laser Marked: 9030103, Humeral Stem Trial 150mm Large / Right, Yellow color indicator (incorrect) " Device Actually: 9030092, Humeral Stem Trial 150mm Medium / Left, Red color indicator (correct)
Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
Users trying to restock a single bin location of an automated dispensing cabinet, during pocket exchange restocking workflow, may experience a software issue that results in 2 destock transactions and 2 stock transactions, which results in doubling the bin quantity, which may result in a delay in access to desired medications and/or supplies, which may lead to temporary body function impairment.
The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
Beckman Coulter DxI 9000 Access Immunoassay Analyzer
Beckman Coulter
When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.
Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-03 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No
Abbott Laboratories
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
Due to product outer packaging incorrectly labeled.
Nerve Monitoring System with certain software versions has potential for increased stimulus artifact, which may require troubleshooting, and may lead to procedure delay, extubation, and medical intervention.
software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope.
Incorrect blister labelling.
The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
Medline Convenience kits used for various procedures: 1) NAIL PACK, Model Number: SUT15830A;
MEDLINE INDUSTRIES, LP - Northfield
There is a potential for open/weak seals on purchased breather pouches used for packaging of select sterile convenience kits.
Brand Name: ID NOW Influenza A/B 2 24T Product Name: ID NOW Influenza A/B 2 24T Model/Catalog Number: 427-000 Software Version: N/A Product Description: Influenza A And Influenza B Multiplex Nucleic Acid Assay; Real Time Nucleic Acid Component: No
Abbott Diagnostics Scarborough
the impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use.
Lack of sterility assurance for closed suction catheter systems
t:slim X2 Insulin Pump with Interoperable Technology
Tandem Diabetes Care
A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous glucose monitor (CGM) sensor, which can lead to under-delivery or over-delivery of insulin based on inaccurate result leading to severe cases of hypoglycemia or hyperglycemia.
There is a potential for open/weak seals on purchased breather pouches used for packaging of select sterile convenience kits.